A Low-Complexity 200 Mbit/s Full-Duplex Digital Link Over a Single Coaxial Cable for Laboratory Instrumentation

arXiv:2603.18848 · physics.ins-det · Submitted 2026-03-19 · Read on arXiv

Listen

Radio episode about this paper

Transcript

Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.

Kai: I'm Kai.

Mira: And I'm Mira, here with Lev.

Kai: We're looking at the paper: "A Low-Complexity 200 Mbit/s Full-Duplex Digital Link Over a Single Coaxial Cable for Laboratory Instrumentation".

Mira: Here's the gist of the paper. A low-complexity bidirectional digital interconnect that enables simultaneous transmission and reception of baseband logic signals over a single coaxial cable is presented for laboratory instrumentation The targeted design performance is fullduplex operation at 200 Mbit s−1 over 10 m of standard RG-178 cable.

Kai: System Design and Components The proposed circuit combines a discrete passive resistive hybrid providing line termination and directional separation, a single logic gate as line driver, and a commercial differential receiver used as a comparator.

Mira: The design target is full-duplex operation at 200 Mbit s−1 over 10 m of standard RG-178 cable.

Kai: The circuit consists only of a passive resistor bridge circuit (a so-called hybrid) and two commodity ICs that can be readily integrated on a printed circuit board (PCB) to establish digital device-to-device communication.

Paper summary: Kai: So, to recap this paper, "A Low-Complexity two hundred Mbit/s Full-Duplex Digital Link Over a Single Coaxial Cable for Laboratory Instrumentation", they are proposing a solution for getting full-duplex digital communication over just one coaxial cable <ref:2603.18848#pg1,A Low-Complexity 200 Mbit/s Full-Duplex Digital Link Over a>.

Mira: The core claim is achieving this at two hundred Megabits per second, but the design philosophy is low complexity, avoiding modulation and active echo cancellation entirely <ref:2603.18848#pg1>. This makes it suitable for lab instrumentation where simplicity matters more than raw speed or ultra-long distance.

Lev: They use a discrete passive resistive hybrid for line termination and directional separation, plus one logic gate as the driver and a commercial differential receiver as the comparator to achieve bidirectional communication simultaneously over the medium.

Kai: The paper focuses on showing that this architecture works for laboratory settings, aiming for moderate-bandwidth full-duplex operation over short, controlled coaxial cables with minimal circuit overhead.

Mira: The main implication here is demonstrating a viable path toward low-complexity digital device-to-device communication using simple passive elements when compared to more complex commercial solutions like RS485 or CAN.

Lev: From an error correction standpoint, the results show deterministic edge timing errors in simulation, which they quantify as a peak-to-peak timing spread of about five hundred picoseconds. This is the noise floor you have to deal with when you try to build this on actual hardware for real data transmission.

Kai: They also measured the jitter versus cable length and found that for cables shorter than six meters, the timing error swings between two hundred picoseconds and eight hundred picoseconds.

Mira: As the cable gets longer, they noted an additional timing error appears due to frequency-dependent cable losses, which degrades those signal edges. When they measure this experimentally on longer cables, the peak-to-peak timing jitter increases to about one point nine nanosecond while still keeping the eye open.

Lev: That one point nine nanosecond jitter is what you have to budget for if you try to run this over a real piece of RG-one hundred seventy-eight cable for a bit longer than six meters, even with this simple design <ref:2603.18848#pg3>.

Kai: And they also showed that without any extra components added, the transceiver provides an inherent loopback function. The signal transmitted out is received back by the same device when no external cable is connected.

Mira: That loopback functionality is useful for testing link integrity before you introduce a second device or a proper termination load to see how the system behaves in isolation.

Lev: I wonder if that inherent loopback really means much in practice when you're trying to set up a full-duplex communication link between two distinct pieces of equipment, but it’s good for initial component testing.

Conclusion: Kai: So, looking at the paper "A Low-Complexity two hundred Mbit/s Full-Duplex Digital Link Over a Single Coaxial Cable for Laboratory Instrumentation", the main point is that you can build this kind of link using just a few simple parts <ref:2603.18848#pg1,A Low-Complexity 200 Mbit/s Full-Duplex Digital Link Over a>.

Mira: They show that this design successfully achieves full-duplex operation at two hundred Megabits per second over ten meters of RG-one hundred seventy-eight cable, relying on a passive resistive hybrid and basic digital components <ref:2603.18848#pg3>.

Lev: The implication is that for specific lab instrumentation needs, you can achieve high data rates with very low circuit complexity compared to what you might see in commercial PHY ICs like the 1000BASE-T example they compare against <ref:2603.18848#pg3>.

Kai: The design targets a specific niche: moderate bandwidth over short, well-controlled coaxial cables where you don't need the massive infrastructure of gigabit Ethernet or other high-speed standards.

Mira: Ultimately, this work shows that you can create a functional digital interconnect using very basic components without needing modulation or active cancellation circuitry for this type of application.

Lev: The limitation they state is that while the design is simple, it doesn't inherently solve the problem of managing significant frequency-dependent losses over much longer distances where those timing errors become unmanageable.

Kai: So, in short, it’s a proof of concept for low-complexity bidirectional digital links over coaxial cable for short runs, focusing on simplicity and direct device communication.

M. Wiebusch

GSI Helmholtzzentrum für Schwerionenforschung GmbH

physics.ins-det

Submitted: 2026-03-19

Updated: 2026-11-06

Comments: 13 pages, 10 figures. This preprint has been submitted to Review of Scientific Instruments published by the American Institute of Physics

License: http://creativecommons.org/licenses/by/4.0/

Importance score: 91/100

The gist: A low-complexity bidirectional digital interconnect that enables simultaneous transmission and reception of baseband logic signals over a single coaxial cable is presented for laboratory…

Key concepts

Passive Resistive Hybrid
This is a circuit made of resistors that acts like a directional bridge on the coaxial cable. It is used here to separate the transmitted signal from the received signal, allowing both directions (sending and receiving) to happen at once without interference.
Directional Bridge Topology
The circuit uses a specific layout for the resistive hybrid, based on a known topology. This structure helps isolate the voltage traveling backward along the cable from the voltage being sent forward by a signal source.
Full-Duplex Operation
This means that data can be sent and received simultaneously over the same single coaxial cable. The system is designed to handle 200 Megabits per second in both directions at the same time, which is crucial for efficient digital communication.

Terminology

Summary

A low-complexity bidirectional digital interconnect that enables simultaneous transmission and reception of baseband logic signals over a single coaxial cable is presented for laboratory instrumentation The targeted design performance is fullduplex operation at 200 Mbit s−1 over 10 m of standard RG-178 cable. p. 0

Where the paper says it

For laboratory instrumentation, the objective may instead be to achieve moderate-bandwidth full-duplex communication over short, well-controlled coaxial cables with minimal circuit and interface complexity. We present a low-complexity bidirectional digital interconnect that enables simultaneous transmission and reception of baseband logic signals over a single coaxial cable.

System Design and Components

The proposed circuit combines a discrete passive resistive hybrid providing line termination and directional separation, a single logic gate as line driver, and a commercial differential receiver used as a comparator. The design target is full-duplex operation at 200 Mbit s−1 over 10 m of standard RG-178 cable. p. 0

Where the paper says it

No modulation, transformer coupling, adaptive calibration, or active analog or digital echo cancellation is required. The design target is full-duplex operation at 200 Mbit s−1 over 10 m of standard RG-178 cable.

The circuit consists only of a passive resistor bridge circuit (a so-called hybrid) and two commodity ICs that can be readily integrated on a printed circuit board (PCB) to establish digital device-to-device communication. p. 0
Where the paper says it

wiebusch@gsi. de be readily integrated on a printed circuit board (PCB) to establish digital device-to-device communication.

Hybrid Circuit Principle

The duplexing technique used in this work is a passive resistive hybrid in the form of a directional bridge, which is depicted in fig. p. 2

Where the paper says it

The output of the difference amplifier outputs only the reverse/reflected component from the far side of the transmission line. The duplexing technique used in this work is a passive resistive hybrid in the form of a directional bridge, which is depicted in fig.

1. The presented circuit is based on a directional bridge topology described by Nash and Brunner15. p. 2
Where the paper says it

The duplexing technique used in this work is a passive resistive hybrid in the form of a directional bridge, which is depicted in fig. The presented circuit is based on a directional bridge topology described by Nash and Brunner15, which was developed for broadband bidirectional RF power and return-loss measurements and which is integrated into the commercial ADL592016.

The duplexing technique used in this work is a passive resistive hybrid in the form of a directional bridge, which is depicted in fig. 1. The voltage Vrev can therefore be interpreted as an attenuated version of the backward-traveling wave on the transmission line, isolated from the forward-traveling wave launched by the voltage source. p. 3
Where the paper says it

signal L→R : Vrev = VR − VLL (6) = g · VL − g · VL (7) = 0 (8) For signals arriving from the far end of the transmission line, including those produced by reflections at the load impedance ZL, Vrev becomes non-zero: signal L←R : Vrev = VR − VLL (9) = VR − g · VL (10) = VR − g · g · VR (11) = (1 − g 2 ) · VR (12) ≈ 0. 5 · VR (−6 dB) (13) The voltage Vrev can therefore be interpreted as an attenuated version of the backward-traveling wave on the transmission line, isolated from the forward-traveling wave launched by the voltage source.

Simulation and Performance

SPICE simulations show that the finite directional isolation of the resistive hybrid introduces deterministic edge timing errors, while preserving sufficient signal integrity for digital operation. The simulated peak-to-peak timing spread is approximately 500 ps. p. 6

Where the paper says it

However its exact shape depends strongly on the cable length and test pattern periods, indicating that reflections caused by imperfect impedance matching contribute to the deterministic jitter mechanism. The simulated peak-to-peak timing spread is approximately 500 ps.

The measured peak-to-peak jitter as a function of cable length is shown in fig. p. 7
Where the paper says it

The arrival times of the 100 MHz signal at the output of transceiver B were measured relative to the waveform generator’s reference clock output, allowing the deterministic timing error introduced by the full-duplex link to be quantified. The measured peak-to-peak jitter as a function of cable length is shown in fig.

p. 7
Where the paper says it

The arrival times of the 100 MHz signal at the output of transceiver B were measured relative to the waveform generator’s reference clock output, allowing the deterministic timing error introduced by the full-duplex link to be quantified. The measured peak-to-peak jitter as a function of cable length is shown in fig.

9, alongside the corresponding values predicted by the SPICE simulation. p. 7
Where the paper says it

The measured peak-to-peak jitter as a function of cable length is shown in fig. 9, alongside the corresponding values predicted by the SPICE simulation, whose continuously variable transmission line delay permits a finer sampling of cable lengths.

p. 7
Where the paper says it

The measured peak-to-peak jitter as a function of cable length is shown in fig. 9, alongside the corresponding values predicted by the SPICE simulation, whose continuously variable transmission line delay permits a finer sampling of cable lengths.

For cable lengths below approximately 6 m, the simulated and measured timing error oscillates between roughly 200 ps and 800 ps. p. 7
Where the paper says it

The locations of the maxima and minima agree well between simulation and experiment. For cable lengths below approximately 6 m, the simulated and measured timing error oscillates between roughly 200 ps and 800 ps.

Experimental Validation

The transceiver was evaluated experimentally using a test setup analogous to the SPICE jitter study described in section V. The measured peak-to-peak jitter as a function of cable length is shown in fig. 9, alongside the corresponding values predicted by the SPICE simulation. For longer cables, an additional timing error is observed, which is attributed to frequency-dependent cable losses and the resulting degradation of signal edges. p. 9

Where the paper says it

It differs only marginally from that obtained with the ideal transmission-line model and therefore does not reproduce the experimentally observed systematic increase in timing error. The additional timing error is therefore likely caused by edge degradation due to frequencydependent cable losses, such as dielectric loss and skin effect, which are not fully represented by the simplified lossy transmission-line model available in the SPICE simulation.

The measured peak-to-peak timing jitter increases to approximately 1. p. 8
Where the paper says it

4 m, corresponding to another empirical maximum of the deterministic jitter. The measured peak-to-peak timing jitter increases to approximately 1.

9 ns while the eye remains open. p. 8
Where the paper says it

The measured peak-to-peak timing jitter increases to approximately 1. 9 ns while the eye remains open.

Loopback Function

Without any additional components, the transceiver inherently provides a loopback function. When no cable is connected to the coaxial port, the transmitted signal is received again by the same transceiver. p. 1

Where the paper says it

Among the earliest examples are transformer-based telephone hybrids, which use a multi-winding transformer and a balancing impedance to cancel the locally transmitted signal at the receiver6 . A closely related concept is the electrical balance duplexer (EBD), which replaces the discrete hybrid transformer by a four-port hybrid junction, a passive network with similar directional cancellation.

The loopback operation ceased when the far end of the cable was terminated with a 50 omega load or connected to an idle second transceiver. p. 3
Where the paper says it

4 source on the left side Vrev is zero, because both terminals, R and LL, see VL attenuated by the same factor. signal L→R : Vrev = VR − VLL (6) = g · VL − g · VL (7) = 0 (8) For signals arriving from the far end of the transmission line, including those produced by reflections at the load impedance ZL, Vrev becomes non-zero: signal L←R : Vrev = VR − VLL (9) = VR − g · VL (10) = VR − g · g · VR (11) = (1 − g 2 ) · VR (12) ≈ 0.

Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.

Improvements for AI systems

  1. Low-complexity bidirectional digital interconnect for laboratory instrumentation can be realized using only a discrete passive resistive hybrid, a single logic gate as line driver, and a commercial differential receiver without needing modulation or adaptive calibration. This allows for establishing digital device-to-device communication over a single coaxial cable with minimal circuit complexity.

  2. The improved system can provide simultaneous transmission and reception of baseband logic signals over short coaxial cables up to 10 meters at 200 Mbit/s, which is useful in laboratory instrumentation where cable routing or feedthrough density is constrained.

  3. The transceiver inherently provides loopback functionality when the transmission line is left open, allowing for simple link-presence detection without additional hardware by comparing the received signal with an expected response.

  4. The system can be adapted to use different cable types, such as 75 Ohm transmission lines, by scaling the hybrid resistor values by a factor of 1.5 to preserve impedance relationships and maintain proper line termination.

  5. The transceiver is suitable for applications where ground potential differences are negligible, as it shares a common ground reference through the coaxial cable shield, limiting its use to electrically compact laboratory setups.

Abstract

Modern full-duplex communication systems, such as Gigabit Ethernet, enable high data rates over challenging transmission media but require sophisticated circuitry and non-trivial digital interfaces. For laboratory instrumentation, the objective may instead be to achieve moderate-bandwidth full-duplex communication over short, well-controlled coaxial cables with minimal circuit and interface complexity. We present a low-complexity bidirectional digital interconnect that enables simultaneous transmission and reception of baseband logic signals over a single coaxial cable. The circuit combines a discrete passive resistive hybrid providing line termination and directional separation, a single logic gate as line driver, and a commercial differential receiver used as a comparator. No modulation, transformer coupling, adaptive calibration, or active analog or digital echo cancellation is required. The design target is full-duplex operation at 200 MBit/s over 10 m of standard RG-178 cable. An analytical treatment of the hybrid network determines the system parameter that maximizes the differential signal amplitude at the receiver, while circuit simulations predict the deterministic timing errors resulting from imperfect directional separation. Experimental measurements confirm the predicted deterministic jitter and show good agreement with the simulation results. For typical laboratory coaxial cables up to 10 m, the measured peak-to-peak edge timing error remains below 1.2 ns. A bidirectional transmission experiment with randomized data demonstrates a clearly open eye diagram. It was found that the transceiver inherently provides loopback functionality when the transmission line is left open. The proposed approach provides a simple, protocol-independent bidirectional physical link which may be useful in laboratory instrumentation where cable routing is constrained.

Related papers