Characterization of CRYO ASIC for charge readout in the nEXO experiment

arXiv:2512.11116 · physics.ins-det, nucl-ex · Submitted 2025-12-11 · Read on arXiv

Listen

Radio episode about this paper

Transcript

Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.

Kai: I'm Kai.

Mira: And I'm Mira, here with Lev.

Kai: We're looking at the paper: "Characterization of CRYO ASIC for charge readout in the nEXO experiment".

Mira: Here's the gist of the paper. nEXO is a proposed next-generation experiment searching for neutrinoless double beta decay of 136Xe using a tonne-scale liquid xenon (LXe) time projection chamber (TPC) The CRYO ASIC achieves the desired performance at LXe temperatures, showing relative response variations below 0.2% for nearly all channels during a 14-hour interval These results establish CRYO ASIC as a viable low-noise in-LXe charge readout solution for nEXO ASIC Design and Functionality The CRYO ASIC is a System-on-Chip (SoC) waveform digitizer and serializer implemented in TSMC’s 130 nm CMOS process, designed specifically for noble liquid TPC experiments The ASIC integrates both analog and digital functions into a single chip, providing 64 input channels each sampled at 2 MSPS with 12-bit resolution across a 1.6 V range Data are read at a rate of 896 Mbps The design minimizes the need for external components by employing on-chip supply regulation by distributing low-dropout (LDO) voltage regulators across the chip CRYO ASIC has been designed using known best-practices for extended operation in cryogenic temperatures

5: and enables the user to correct for temperature-dependent effects by directly programming the voltage levels in core sections through internal registers Cryogenic Testing and Stability The initial performance of CRYO ASIC are demonstrated in the gas-nitrogen environmental chamber with both Cold Board and the FEMB boards The optimized register settings for the LXe temperature is then applied to all tests presented in this work The pulse response of the FEMB board is verified by enabling the internal pulser on ASICs, as illustrated in the block diagram in Fig.

Kai: 1, using configurations with four different peaking times The channel gains are measured once per hour using both the internal and the external pulser calibration systems The average amplitudes of both systems calibrated on one representative channel (channel 27) are shown in the lower left panel of Fig.

Mira: 5, where the two gain-variation curves track each other well which indicates the ability of self-calibration by the internal pulses In-LXe Noise Performance Significant boiling is observed after the CRYO ASIC is powered on, once the LXe conditions stabilize at a pressure of 0.095 MPa and a temperature of 170.45 K, as shown in the left panel of Fig.

Kai: 6 The mitigation of LXe boiling is confirmed by both the camera monitoring and the CRYO ASIC noise measurements Under these conditions, the ASIC achieves an average noise performance of 156 e− ENC at a gain of 28.6 mV/fC, a peaking time of 1.

Paper summary: Kai: So, this paper is about testing a chip called CRYO ASIC for reading out signals in the nEXO experiment.

Mira: It's basically checking if this specific piece of hardware can actually handle the liquid xenon conditions it needs to work in.

Kai: The authors are looking at how stable this chip is when you run it at those very low temperatures and pressures.

Mira: They are showing that the response variations stay below point zero two percent for nearly all channels over a fourteen hour period in liquid xenon.

Kai: That level of stability, if they can maintain it under real detector conditions, means we have a much more reliable way to measure the tiny signals they're looking for.

Mira: But we need to remember the big picture here, this chip is just one part of a tonne-scale experiment searching for neutrinoless double beta decay.

Kai: Right, so it’s not about finding the decay itself right now, but making sure we have a trustworthy tool to measure what happens if it does occur.

Mira: The main implication is that this chip proves it’s viable as a low-noise readout solution for the nEXO experiment.

Kai: And they're setting up the next steps to see how this component fits into the whole detector design and overall system performance.

Conclusion: Kai: So we're wrapping up this look at the "Characterization of CRYO ASIC for charge readout in the nEXO experiment."

Mira: Basically, they're showing that this specific chip can handle the liquid xenon environment well enough to be a reliable readout component.

Kai: The authors are focused on proving that this hardware works under those specific low temperatures and pressures they need.

Mira: They demonstrate that the response variations stay below zero point two percent for almost every channel over a fourteen hour test period in liquid xenon.

Kai: That level of stability means we have a much more dependable way to measure those tiny signals they're hunting for in the experiment.

Mira: But we gotta remember this chip is just one part of the whole thing, which is searching for neutrinoless double beta decay across a tonne-scale detector.

Kai: Right, so the main point isn't finding that decay right now, but making sure we have a trustworthy tool ready to measure it when it happens.

Mira: The big takeaway here is that this chip proves itself viable as a low-noise readout solution for the nEXO experiment.

Kai: And they’re already planning what comes next, figuring out how this component fits into the overall detector design and system performance.

Mira: That leads us right into whether this chip can actually handle the scale and complexity required for a tonne-scale search.

University of Hawaii at Manoa · SLAC National Accelerator Laboratory · Stanford University · Drexel University · University of California San Diego · University of Alabama · È Amherst Center for Fundamental Interactions and Physics Department, University of Massachusetts Amherst Center for Fundamental Interactions and Physics Department, University of Massachusetts Amherst Center for Fundamental Interactions and Physics Department, University of Massachusetts Amherst Center for Fundamental Interactions and Physics Department, University of Massachusetts Amherst Center for Fundamental Interactions and Physics Department, University of Massachusetts Amherst Center for Fundamental Interactions and Physics Department · Pacific Northwest National Laboratory · Université de Sherbrooke · National Research Center “Kurchatov Institute” · Lawrence Livermore National Laboratory · ž McGill University · ƒ SNOLAB Lively, ON P3Y 1N2 Canada · Institute of High Energy Physics, Chinese Academy of Sciences Beijing, 100049 China · Institute of Microelectronics, Chinese Academy of Sciences Beijing, 100029 China · Department of Physics and Astronomy University of Kentucky Lexington KY 40506 USA · Brookhaven National Laboratory Upton NY 11973 USA · Rensselaer Polytechnic Institute Troy NY 12180 USA · ž TRIUMF Vancouver BC V6T 2A3 Canada · Department of Physics and Astronomy University of Windsor Windsor ON N9B 3P4 Canada · Oak Ridge National Laboratory Oak Ridge TN 37831 USA · Colorado School of Mines Golden CO 80401 USA

physics.ins-det, nucl-ex

Submitted: 2025-12-11

Updated: 2026-11-10

Code: https://github.com/slaclab/surf

Project page: https://slaclab.github.io/rogue

License: http://creativecommons.org/licenses/by/4.0/

Importance score: 99/100

The gist: nEXO is a proposed next-generation experiment searching for neutrinoless double beta decay of 136Xe using a tonne-scale liquid xenon (LXe) time projection chamber (TPC) The CRYO ASIC achieves the…

Key concepts

CRYO ASIC
This is the System-on-Chip device designed to digitize and serialize charge signals from the liquid xenon detector. It integrates analog and digital functions onto a single chip, using 64 channels sampled at 2 MSPS with 12-bit resolution, while minimizing external components through on-chip power regulation.
LXe Boiling Mitigation
When the CRYO ASIC is powered on in liquid xenon (LXe) conditions, significant boiling occurs. The experiment confirmed that monitoring and noise measurements allowed for successful mitigation of this boiling effect once the LXe stabilized at 170.45 K.
ENC
Electronic Noise Charge (ENC) is a key metric used to assess the noise performance of the readout system. The ASIC achieved an average ENC of 156 e- at a gain of 28.6 mV/fC, indicating its low noise capability for detecting rare events in the experiment.

Terminology

Summary

nEXO is a proposed next-generation experiment searching for neutrinoless double beta decay of 136Xe using a tonne-scale liquid xenon (LXe) time projection chamber (TPC) The CRYO ASIC achieves the desired performance at LXe temperatures, showing relative response variations below 0. p. 2

Where the paper says it

Abstract: nEXO is a proposed next-generation experiment searching for the neutrinoless double beta decay of 136Xe using a tonne-scale liquid xenon (LXe) time projection chamber (TPC). To image the ionization signals from events in the LXe, the detector will employ metallized fusedsilica charge collection tiles instrumented with cryogenic application-specific integrated circuits (ASICs), referred to as CRYO ASIC, which are designed to operate directly in LXe to minimize input capacitance and pick-up noise.

2% for nearly all channels during a 14-hour interval These results establish CRYO ASIC as a viable low-noise in-LXe charge readout solution for nEXO

ASIC Design and Functionality

The CRYO ASIC is a System-on-Chip (SoC) waveform digitizer and serializer implemented in TSMC’s 130 nm CMOS process, designed specifically for noble liquid TPC experiments The ASIC integrates both analog and digital functions into a single chip, providing 64 input channels each sampled at 2 MSPS with 12-bit resolution across a 1.6 V range Data are read at a rate of 896 Mbps The design minimizes the need for external components by employing on-chip supply regulation by distributing low-dropout (LDO) voltage regulators across the chip CRYO ASIC has been designed using known best-practices for extended operation in cryogenic temperatures [5] and enables the user to correct for temperature-dependent effects by directly programming the voltage levels in core sections through internal registers

Cryogenic Testing and Stability

The initial performance of CRYO ASIC are demonstrated in the gas-nitrogen environmental chamber with both Cold Board and the FEMB boards The optimized register settings for the LXe temperature is then applied to all tests presented in this work The pulse response of the FEMB board is verified by enabling the internal pulser on ASICs, as illustrated in the block diagram in Fig. 1, using configurations with four different peaking times The channel gains are measured once per hour using both the internal and the external pulser calibration systems The average amplitudes of both systems calibrated on one representative channel (channel 27) are shown in the lower left panel of Fig. p. 9

Where the paper says it

The channel gains are measured once per hour using both the internal and the external pulser calibration systems. The external pulse generator and associated cabling adds significant noise to the input when connected, and several grounding configurations were tested to minimize pickup noise to acceptable levels (∼ 20 ADC RMS).

5, where the two gain-variation curves track each other well which indicates the ability of self-calibration by the internal pulses

In-LXe Noise Performance

Significant boiling is observed after the CRYO ASIC is powered on, once the LXe conditions stabilize at a pressure of 0.095 MPa and a temperature of 170. p. 4

Where the paper says it

To comply with the stringent lowradioactivity requirements, the design minimizes the need for external components by employing on-chip supply regulation by distributing low-dropout (LDO) voltage regulators across the chip. CRYO ASIC has been designed using known best-practices for extended operation in cryogenic temperatures [5], and enables the user to correct for temperature-dependent effects by directly programming the voltage levels in core sections through internal registers.

45 K, as shown in the left panel of Fig. p. 6
Where the paper says it

Data is transmitted using a custom high-speed cable from SAMTEC. 2 Cryogenic test stands A liquid-nitrogen-cooled environmental chamber [12], shown in the left panel of Fig.

6 The mitigation of LXe boiling is confirmed by both the camera monitoring and the CRYO ASIC noise measurements Under these conditions, the ASIC achieves an average noise performance of 156 e− ENC at a gain of 28. p. 3
Where the paper says it

2 Study of boiling conditions during LXe operation 8 3. 3 Noise measurements with CRYO ASIC in LXe 9 4 Summary and outlook 10 5 Acknowledgments 11 1 Introduction The nEXO experiment is a proposed next-generation search for neutrinoless double beta decay (0𝜈𝛽𝛽), aiming to achieve a half-life sensitivity beyond 1028 years using an isotopically-enriched liquid xenon (LXe) time projection chamber (TPC) [1].

6 mV/fC, a peaking time of 1. p. 4
Where the paper says it

The ASIC integrates both analog and digital functions into a single chip, providing 64 input channels each sampled at 2 MSPS with 12-bit resolution across a 1. The gain and peaking time of the front end are configurable between four different values each to optimize between resolution, dynamic range, and noise.

Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.

Improvements for AI systems

  1. System can achieve low-noise in-LXe charge readout by employing CRYO ASIC which is designed to operate directly in LXe to minimize input capacitance and pick-up noise. This allows for a noise level of 156 e minus ENC at a gain of 28.6 mV/fC and a peaking time of 1.2 us with an input capacitance of 22 pF.

  2. System can mitigate LXe boiling to enable stable in-LXe operation by operating the system above ~0.1 MPa, which is confirmed to be sufficient for nEXO operating conditions around 0.15 MPa, as the results indicate that the CRYO ASIC will not induce boiling under the planned detector operating conditions.

  3. System can perform periodic in-situ calibration of channel response by utilizing an internal pulser system which can be used to track charge response, and gains calibrated by both internal and external pulsers are shown to satisfy the "<0.2% variation requirement for nEXO."

Abstract

nEXO is a proposed next-generation experiment searching for the neutrinoless double beta decay of 136 Xe using a tonne-scale liquid xenon (LXe) time projection chamber (TPC). To image the ionization signals from events in the liquid xenon, the detector will employ metallized fused-silica charge collection tiles instrumented with cryogenic application-specific integrated circuits (ASICs), referred to as CRYO ASIC, which are designed to operate directly in LXe to minimize input capacitance and pick-up noise. Here we present the performance of the CRYO ASIC mounted on an auxiliary printed circuit board and evaluated both in a cryogenic environmental chamber and in a dedicated LXe test stand. We demonstrate that the ASICs achieve the desired performance at liquid xenon temperatures, showing a gain stability better than 0.2% over 24-hour operation and reliable in-situ calibration using an on-chip pulser. In the LXe test stand, we show that boiling caused by the chip heat dissipation can be mitigated by operating the system above 0.1 MPa. The in-LXe noise measured agrees with simulation, which indicates it the 150 e- design requirement can be satisfied. These results establish CRYO ASIC as a viable low-noise in-LXe charge readout solution for nEXO.

Sources

Related papers