Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities

arXiv:2606.25804 · physics.ins-det, physics.med-ph · Submitted 2026-06-24 · Read on arXiv

Listen

Radio episode about this paper

Transcript

Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.

Kai: I'm Kai.

Mira: And I'm Mira, here with Lev.

Kai: We're looking at the paper: "Positron Emission Tomography with quantum-entangled Compton events".

Mira: Here's the gist of the paper. Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities investigates exploiting the quantum entanglement of annihilation quanta to enhance PET imaging by measuring polarization correlations through Compton scattering events in a novel single-layer detector setup.

Kai: The results show that a spatial resolution of 2.5±0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.6 ±0.3 mm to 4.9 ±0.3 mm, depending on data selection criteria <ref:2606.25804#pg8> The results show that a spatial resolution of 2.5±0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.6 ±0.3 mm to 4.9 ±0.3 mm, depending on data selection criteria <ref:2606.25804#pg8>.

Mira: Experimental Setup and Data Acquisition The experimental setup comprised two detector modules based on segmented scintillators coupled to silicon photomultiplier (SiPM) arrays and front-end readout electronics <ref:2606.25804#pg4> Each module had a dedicated cooling device that served a dual purpose of venting the ASICs in the front-end boards and of maintaining stability of the SiPMs’ temperature <ref:2606.25804#pg4>.

Kai: The modules were mounted on an aluminum ring supported by mechanical construction, with a sliding platform to position the investigated object in the field of view <ref:2606.25804#pg4> Each module contained 256 GAGG:Ce scintillator crystals with 3.2 mm pitch, assembled from four 8 × 8 detector matrices <ref:2606.25804#pg4>.

Paper summary: Kai: So, to recap, this paper is about investigating whether we can use the quantum entanglement of annihilation quanta by exploiting their polarization correlations to enhance PET imaging one <ref:2606.25804#pg1>. The core idea is that these correlations are a consequence of quantum mechanical properties following the decay of the electron plus positron system two <ref:2606.25804#pg1>.

Mira: They built a demonstrator specifically to measure these polarization correlations using single-layer scintillator polarimeters that function as both the Compton scatterer and the absorber three <ref:2606.25804#pg1>. They claim this setup is feasible, having been previously demonstrated in a lab and optimized under other conditions eleven nineteen <ref:2606.25804#pg3>.

Lev: What they are claiming is that they can get PET images of sources with clinically relevant activities by using these polarization-correlated Compton events instead of just conventional single-pixel event selection three <ref:2606.25804#pg1>.

Kai: Their main result is the comparison between the spatial resolution achieved with single-pixel events, which was two point five plus or minus zero point one millimeters, and the resolution obtained when using those polarization-correlated Compton events, which ranged from three point six plus or minus zero three millimeters to four point nine plus or minus zero three millimeters depending on data selection criteria two <ref:2606.25804#pg1>.

Mira: They also found that the polarization-correlated Compton events exhibit up to twenty percent higher average signal to random background ratio compared to the single-pixel events two <ref:2606.25804#pg1>. That’s a significant statistical advantage in terms of image quality metrics.

Lev: The mechanism they are focusing on is how the uncertainty in assigning the location of interaction, which affects the assignment of LOR, results in this spread in resolution eight <ref:2606.25804#pg3>. They say that when that distance between those two pixels is smaller, that uncertainty also gets lower and the resulting resolution gets better eight <ref:2606.25804#pg3>.

Kai: So what this means for us right now is they’ve managed to get actual PET images of sources up to three hundred seventy-eight megabecquerels activity using these polarization-correlated events alongside conventional single-pixel event selection ten <ref:2606.25804#pg1>.

Mira: The implication is that if we can manage the trade-off between resolution and signal quality, exploiting this quantum entanglement could potentially give us a ten percent sensitivity increase while keeping the image quality high two <ref:2606.25804#pg1>.

Lev: I have to point out that they also state a limitation: the reduced spatial resolution of those inter-crystal Compton event images is currently posing a barrier to achieving the full benefit of this quantum entanglement enhanced PET imaging ten <ref:2606.25804#pg1>. That's where we have to be realistic about what’s achievable right now.

Kai: So, it shows that while they built something novel and demonstrated feasibility, the next step is overcoming that resolution hurdle so we can really see how much better this method can perform in practice one <ref:2606.25804#pg1>.

Conclusion: Kai: Looking at the title of this paper, "Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities", it really tells us the scope of what they aimed to achieve one <ref:2606.25804#pg1,Positron Emission Tomography with quantum-entangled Compton events: first imaging results at>. They weren't just doing a basic PET measurement; they were specifically aiming for something enhanced by quantum entanglement.

Mira: It shows that the whole point is using those polarization correlations of annihilation quanta, which are a consequence of quantum mechanics in the electron plus positron system two <ref:2606.25804#pg1,polarization correlations of annihilation quanta>. The authors are showing that this correlation can be exploited to improve image quality by measuring Compton scattering events three <ref:2606.25804#pg1>.

Lev: From an engineering perspective, they’ve built a demonstrator that shows the concept works. It proves the concept is physically achievable using their single-layer detector design three <ref:2606.25804#pg1>. They've done the hard part of building and testing it for real activity levels, which is something we need to consider when thinking about scaling this up ten <ref:2606.25804#pg1>.

Kai: So what does this mean for a clinician or a researcher? It means they have a tool that can potentially offer better sensitivity than current standard PET systems two <ref:2606.25804#pg1>. The results suggest that an improvement in sensitivity up to ten percent might be attainable while keeping the image quality at least the same, or even slightly improving it two <ref:2606.25804#pg1>.

Mira: But we need to remember the caveat they put forward about resolution: the reduced spatial resolution of those inter-crystal Compton event images is a current barrier ten <ref:2606.25804#pg1>. So, it’s not a full replacement for existing PET technology just yet.

Lev: The implication is that this work validates the physical possibility of using these quantum entanglement effects in PET imaging, but it also sets out the path forward by showing what's possible and where the limitations lie one <ref:2606.25804#pg1>.

Kai: So, in short, they proved you can measure these correlations with their novel detector setup and got first images of sources up to three hundred seventy-eight megabecquerels activity using these polarization-correlated events ten <ref:2606.25804#pg1>.

Mira: It's a demonstration that exploiting the quantum entanglement of annihilation quanta inscribed in the correlation of their polarizations has potential for enhancing PET imaging two <ref:2606.25804#pg1,annihilation quanta inscribed in the correlation of their polarizations>. This is a very interesting direction for future research in this field.

University of Zagreb Faculty of Science Department of Physics University Hospital Centre Zagreb School of Physics University of Sydney Instituto de Fisica Corpuscular CSIC-UV

physics.ins-det, physics.med-ph

Submitted: 2026-06-24

Updated: 2026-11-03

Comments: 20 pages, 11 figures

License: http://creativecommons.org/licenses/by/4.0/

Importance score: 76/100

The gist: Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities investigates exploiting the quantum entanglement of annihilation quanta to…

Key concepts

Quantum Entanglement of Annihilation Quanta
This refers to the quantum correlation between the particles produced when a positron annihilates with an electron. The paper investigates using this inherent entanglement, specifically its polarization correlations, as a feature in PET imaging detectors to potentially improve image quality and sensitivity beyond standard methods.
Compton Scattering Events
When gamma photons interact with detector material, they can undergo Compton scattering. In this study, researchers specifically selected events where both gamma photons scatter within the same detector module. Analyzing the kinematics of these dual scatters allows them to measure polarization correlations.
Spatial Resolution (SPE vs PCE)
Spatial resolution measures how precisely a system can locate a point in space. The study compared two methods: single-pixel events, which yielded 2.5±0.1 mm resolution, and polarization-correlated events, which ranged from 3.6±0.3 mm to 4.9±0.3 mm.

Terminology

Summary

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities investigates exploiting the quantum entanglement of annihilation quanta to enhance PET imaging by measuring polarization correlations through Compton scattering events in a novel single-layer detector setup. p. 0

Where the paper says it

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek 1 , Ana Marija Kožuljević1,2, Tomislav Bokulić 1 , Darko Grošev 3 , Zdenka Kuncic 4 , Siddharth Parashari 5 , and Luka Pavelić 2 1University of Zagreb Faculty of Science, Department of Physics, Zagreb, 10000, Croatia 2 Institute for Medical Research and Occupational Health, Division of Radiation Protection, Zagreb, 10000, Croatia 3University Hospital Centre Zagreb, Department of Nuclear Medicine, Zagreb, 10000, Croatia 4School of Physics, University of Sydney, Sydney, NSW 2006, Australia 5 Instituto de Física Corpuscular (IFIC), CSIC-UV, C/ Catedrático José Beltrán, 2, Paterna, Spain June 25, 2026 Abstract In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters.

The results show that a spatial resolution of 2. p. 0

Where the paper says it

We present a detailed study of the imaging of two 68Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality. The results show that a spatial resolution of 2.

5±0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3. p. 0
Where the paper says it

The results show that a spatial resolution of 2. 1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.

p. 0
Where the paper says it

The results show that a spatial resolution of 2. 1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.

p. 0
Where the paper says it

The results show that a spatial resolution of 2. 1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.

p. 0
Where the paper says it

The results show that a spatial resolution of 2. 1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.

6 ±0.3 mm to 4.9 ±0.3 mm, depending on data selection criteria <ref:2606. p. 0
Where the paper says it

1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3. 3 mm, depending on data selection criteria.

p. 0
Where the paper says it

1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3. 3 mm, depending on data selection criteria.

p. 0
Where the paper says it

1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3. 3 mm, depending on data selection criteria.

p. 0
Where the paper says it

1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3. 3 mm, depending on data selection criteria.

25804#pg8> The results show that a spatial resolution of 2. p. 0
Where the paper says it

We present a detailed study of the imaging of two 68Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality. The results show that a spatial resolution of 2.

5±0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.6 ±0.3 mm to 4.9 ±0.3 mm, depending on data selection criteria <ref:2606.25804#pg8>.

Experimental Setup and Data Acquisition

The experimental setup comprised two detector modules based on segmented scintillators coupled to silicon photomultiplier (SiPM) arrays and front-end readout electronics <ref:2606.25804#pg4> Each module had a dedicated cooling device that served a dual purpose of venting the ASICs in the front-end boards and of maintaining stability of the SiPMs’ temperature <ref:2606. p. 2

Where the paper says it

Methods Experimental setup The experimental setup comprised two detector modules based on segmented scintillators coupled to silicon photomultiplier (SiPM) arrays and front-end readout electronics (Figure 1 (a)). Each module had a dedicated cooling device that served a dual purpose of venting the ASICs in the front-end boards and of maintaining stability of the SiPMs’ temperature.

25804#pg4>. The modules were mounted on an aluminum ring supported by mechanical construction, with a sliding platform to position the investigated object in the field of view <ref:2606. p. 2
Where the paper says it

Each module had a dedicated cooling device that served a dual purpose of venting the ASICs in the front-end boards and of maintaining stability of the SiPMs’ temperature. The modules were mounted on an aluminum ring supported by mechanical construction, with a sliding platform to position the investigated object in the field of view (Figure 1 (b)).

25804#pg4> Each module contained 256 GAGG:Ce scintillator crystals with 3. p. 2
Where the paper says it

The modules were mounted on an aluminum ring supported by mechanical construction, with a sliding platform to position the investigated object in the field of view (Figure 1 (b)). Each module contained 256 GAGG:Ce scintillator crystals with 3.

2 mm pitch, assembled from four 8 × 8 detector matrices <ref:2606. p. 2
Where the paper says it

Each module contained 256 GAGG:Ce scintillator crystals with 3. 2 mm pitch, assembled from four 8 × 8 detector matrices.

25804#pg4>. The length of the crystals was 20 mm, and their cross sections were 3. p. 2
Where the paper says it

2 mm pitch, assembled from four 8 × 8 detector matrices. The length of the crystals was 20 mm, and their cross sections were 3.

0 mm x 3.0 mm in one and 2. p. 2
Where the paper says it

The length of the crystals was 20 mm, and their cross sections were 3. 0 mm in one and 2.

9 mm x 2.9 mm in the other module <ref:2606. p. 0
Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg4>.

Event Selection Criteria

From the full data set, events were selected where both gamma photons undergo Compton scattering within the respective module by following these requirements:

  1. exactly two pixels fire (Epx > 120 keV) in a module <ref:2606. p. 3

    Where the paper says it

    For the 620 mm (430 mm) diameter, the data were acquired at 16 (12) different angular positions, each rotated by 11. From the full data set, we selected events where both gamma photons undergo Compton scattering within the respective module, by the following requirements: (a) exactly two pixels fire (Epx > 120 keV) in a module, (b) the energy of the 511 keV gamma is shared between these two pixels, requiring their sum to be within ±3σ of the 511 keV peak maximum, (c) energy depositions correspond to Compton scattering kinematics, as described by Parashari et al.

    25804#pg5>

  2. the energy of the 511 keV gamma is shared between these two pixels, requiring their sum to be within ±3σ of the 511 keV peak maximum <ref:2606. p. 3

    Where the paper says it

    For the 620 mm (430 mm) diameter, the data were acquired at 16 (12) different angular positions, each rotated by 11. From the full data set, we selected events where both gamma photons undergo Compton scattering within the respective module, by the following requirements: (a) exactly two pixels fire (Epx > 120 keV) in a module, (b) the energy of the 511 keV gamma is shared between these two pixels, requiring their sum to be within ±3σ of the 511 keV peak maximum, (c) energy depositions correspond to Compton scattering kinematics, as described by Parashari et al.

    25804#pg5>

  3. energy depositions correspond to Compton scattering kinematics, as described by Parashari et al. p. 3

    Where the paper says it

    For the 620 mm (430 mm) diameter, the data were acquired at 16 (12) different angular positions, each rotated by 11. From the full data set, we selected events where both gamma photons undergo Compton scattering within the respective module, by the following requirements: (a) exactly two pixels fire (Epx > 120 keV) in a module, (b) the energy of the 511 keV gamma is shared between these two pixels, requiring their sum to be within ±3σ of the 511 keV peak maximum, (c) energy depositions correspond to Compton scattering kinematics, as described by Parashari et al.

    p. 3
    Where the paper says it

    For the 620 mm (430 mm) diameter, the data were acquired at 16 (12) different angular positions, each rotated by 11. From the full data set, we selected events where both gamma photons undergo Compton scattering within the respective module, by the following requirements: (a) exactly two pixels fire (Epx > 120 keV) in a module, (b) the energy of the 511 keV gamma is shared between these two pixels, requiring their sum to be within ±3σ of the 511 keV peak maximum, (c) energy depositions correspond to Compton scattering kinematics, as described by Parashari et al.

    2022 <ref:2606. p. 0
    Where the paper says it

    Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

    25804#pg5>.

Further event selection was done based on event topology, i. p. 3

Where the paper says it

In events satisfying these criteria, it is not possible to distinguish with certainty which pixel corresponds to the recoil electron and which to the Compton-scattered photon absorption. According to GEANT4[1] simulations of the setup[11], on average, a slight preference (≈ 51% cases) is given to the forward scattering scenario; therefore, the Compton scattering angles were reconstructed assuming the pixel with the lower energy (Epx1 ) corresponded to the recoil electron, as: θ= acos mec 2 Epx1+Epx2 − mec 2 Epx2 + 1 ; ϕ= atan ∆y ∆x (3) Further event selection was done based on event topology, i.

e., the distance of the fired pixels in the respective module d1,2 = p∆x squared + ∆y squared <ref:2606. p. 4
Where the paper says it

the fired pixels in the respective module, d1,2 = p ∆x 2 + ∆y 2, where (x,y) are coordinates of the pixel centers in the module reference frame. The detectors have non-uniform azimuthal acceptance (ϕ), since the absorption probability of the scattered gamma photon depends on the distance traveled.

25804#pg6> The raw distribution N(ϕ1 − ϕ2) distribution was acceptance-corrected according to Ncor(ϕ1 − ϕ2) = N(ϕ1 − ϕ2) / Amix(ϕ1 − ϕ2) <ref:2606. p. 4
Where the paper says it

Therefore, the ϕ angles covered by a combination of pixels with a large inter-pixel distance d, are less likely to be detected. To take this into account, the raw distribution N(ϕ1 − ϕ2) distribution, was acceptance-corrected according to: Ncor(ϕ1 − ϕ2) = N(ϕ1 − ϕ2) Amix(ϕ1 − ϕ2) (4) where Amix(ϕ1 − ϕ2) is the ϕ1 − ϕ2 distribution obtained by the event mixing technique[14].

25804#pg6>.

Performance Metrics and Sensitivity Estimation

The polarization-correlated Compton events exhibit up to 20% higher average signal to random background ratio compared to the single-pixel events <ref:2606.25804#pg8> Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events <ref:2606. p. 0

Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg8>.

Imaging Results and Quality Assessment

The first images of sources with up to 378 MBq activities were successfully reconstructed using solely polarizationcorrelated events with spatial resolution below 5 mm in the central field of view <ref:2606.25804#pg10> The spatial resolutions obtained, extracted from the profile, were 2. p. 6

Where the paper says it

The spatial resolution is the average of the full width at half maximum of the Gaussian fits to the peaks. The obtained spatial resolutions, extracted from the profile are 2.

5 ± 0.1 mm and 4.0 ± 0.1 mm for the SPE and PCE, respectively <ref:2606. p. 6
Where the paper says it

The acquisition time was 3 hours per position, with 16 (12) positions covering the full circle at 620 mm (430 mm) diameter. For imaging at 430 mm ring diameter, the axial image of the rod sources reconstructed with SPE-only and PCE-only are shown in Figure 4 (a) and (b), respectively.

25804#pg8>. The results show that S/B ratio is maintained when using additional events, provided their spatial resolution is below 4 mm <ref:2606. p. 6
Where the paper says it

The spatial resolutions of these images are listed in the left-most column of the Table 1, and the results demonstrate that they only marginally differ from the SPE image resolution, also shown in Figure 7 (a). In Figure 7 (b), we show the ratio of the SBR for PCE and for SPE extracted from the coincidence time plot (Figure 3, showing that selected PCE always contain a lower random fraction than the single-pixel events.

25804#pg9>.

Conclusion

We have successfully constructed and operated a novel detector setup to demonstrate and assess the feasibility of using quantum-entanglement of the annihilation quanta in PET imaging by exploiting their pronounced polarization correlations <ref:2606.25804#pg10> While the reduced spatial resolution of the inter-crystal Compton event images, currently poses a barrier to achieve the full benefit of the quantum-entanglement enhanced PET imaging, the results still suggest that an improvement in sensitivity up to 10% may be reached, while keeping or even slightly improving the image quality <ref:2606. p. 2

Where the paper says it

This work reports the performance of demonstrational setup, based on the single-layer concept for implementing polarization measurement in PET, and for the first time, presents the corresponding PET images of source distributions with clinically relevant activities, obtained by utilizing polarization-correlated Compton events. Further, we compare these to images obtained by conventional single-pixel event selection in terms of spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise, and estimate a sensitivity gain by exploiting the polarization-correlated inter-crystal scatter events.

25804#pg10>.

References

[1] Agostinelli S. et al. Geant4—a simulation toolkit <ref:2606. p. 0

Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg8>

[2] D. Bohm and Y. Aharonov. Discussion of experimental proof for the paradox of Einstein, Rosen, and Podolsky <ref:2606. p. 0

Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg2>

[3] A. Di Francesco, R. Bugalho, L. Oliveira, L. Pacher, A. Rivetti, M. Rolo, J. C. Silva, R. Silva, and J. Varela

[4] P. Dirac: The quantum theory of the electron <ref:2606. p. 0

Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg2>

[5] Francisco Enriquez-Mier-y Terán, Luping Zhou, Steve Meikle, and Andre Kyme

[6] H.M. Hudson and R.S. Larkin: Accelerated image reconstruction using ordered subsets of projection data <ref:2606. p. 5

Where the paper says it

The lines of response (LOR) were formed using the coordinates of the centers of pixel faces, in which the first interaction was assumed to happen. The Ordered Subset Expectation Maximization algorithm [6] was used for the image reconstruction.

25804#pg2>

[7] L. R. Kasday, J.D. Ullman, and C. S. Wu: Angular correlation of comptonscattered annihilation photons and hidden variables

[8] Kim D. et al.: Background reduction in PET by double compton scattering of quantum entangled annihilation photon

[9] Jeffrey A. Kolthammer, Kuan-Hao Su, Aarati Grover, Manoj Narayanan, David W. Jordan, and Raymond F. Muzic: Performance evaluation of the ingenuity tf pet/ct scanner with a focus on high count-rate conditions

[10] Kožuljević A. M. et al.: Investigation of the spatial resolution of PET imaging system measuring polarization-correlated Compton events

[11] Kožuljević A.M. et al.: Study of multi-pixel scintillator detector configurations for measuring polarized gamma radiation

[12] Kuncic Z. et al.: Polarization enhanced X-ray imaging for biomedicine <ref:2606. p. 2

Where the paper says it

this, we implemented the concept of a single-layer detector in which a single segmented scintillator matrix is used as the gamma-ray polarimeter, acting as both the Compton scatterer and the absorber. The feasibility of the concept to measure polarization correlation of annihilation quanta was previously demonstrated under laboratory conditions[14], and further optimized[11, 19].

25804#pg2>

[13] H. Langhoff: Die linearpolarisation der vernichtungsstrahlung von positronen <ref:2606. p. 0

Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg2>

[14] Makek M. et al.: Single-layer Compton detectors for measurement of polarization correlations of annihilation quanta

[15] McNamara A. et al.: Towards optimal imaging with PET: An in silico feasibility study

[16] Steven Meikle et al.: Quantitative PET in the 2020s: a roadmap <ref:2606. p. 0

Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg2>

[17] P. Moskal, D. Kumar, S. Sharma, E. Y. Beyene, N. Chug, A. Coussat, C. Curceanu, E. Czerwinski, M. Das, K. Dulski, M Gorgol: Non-maximal entanglement of photons from positron-electron annihilation demonstrated using a novel plastic PET scanner <ref:2606. p. 1

Where the paper says it

Keywords: Positron Emission Tomography; quantum entanglement; photon polarization; NEMA NU-4 phantom; spatial resolution; image quality Introduction Positron emission tomography (PET), an irreplaceable diagnostic tool for medical imaging, is based on the coincident detection of two gamma photons originating from positron annihilation in patient’s body. The gamma photons are typically detected by scintillation detectors capable of measuring photon energy, time of arrival and hit position.

25804#pg2>

[18] Tinsu Pan, Stephen A. Einstein, S. Cheenu Kappadath, Kelley S. Grogg, Claudia Lois Gomez, Adam M. Alessio, William C. Hunter, Georges El Fakhri, Paul E. Kinahan, and Osama R. Mawlawi: Performance evaluation of the 5-ring ge discovery mi pet/ct system using the national electrical manufacturers association nu 2-2012 standard <ref:2606.25804#pg2>

[19] Parashari S. et al.: Optimisation of detector modules for measuring gammaray polarization in Positron Emission Tomography <ref:2606. p. 0

Where the paper says it

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek 1 , Ana Marija Kožuljević1,2, Tomislav Bokulić 1 , Darko Grošev 3 , Zdenka Kuncic 4 , Siddharth Parashari 5 , and Luka Pavelić 2 1University of Zagreb Faculty of Science, Department of Physics, Zagreb, 10000, Croatia 2 Institute for Medical Research and Occupational Health, Division of Radiation Protection, Zagreb, 10000, Croatia 3University Hospital Centre Zagreb, Department of Nuclear Medicine, Zagreb, 10000, Croatia 4School of Physics, University of Sydney, Sydney, NSW 2006, Australia 5 Instituto de Física Corpuscular (IFIC), CSIC-UV, C/ Catedrático José Beltrán, 2, Paterna, Spain June 25, 2026 Abstract In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters.

25804#pg2>

[20] Eric Petersen, Andy LaBella, Yixin Li, Zipai Wang, and Amir H Goldan: Resolving inter-crystal scatter in a light-sharing depth-encoding pet detector <ref:2606.25804#pg2>

[21] M. H. L. Pryce and J. C. Ward: Angular correlation effects with annihilation radiation <ref:2606. p. 0

Where the paper says it

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek 1 , Ana Marija Kožuljević1,2, Tomislav Bokulić 1 , Darko Grošev 3 , Zdenka Kuncic 4 , Siddharth Parashari 5 , and Luka Pavelić 2 1University of Zagreb Faculty of Science, Department of Physics, Zagreb, 10000, Croatia 2 Institute for Medical Research and Occupational Health, Division of Radiation Protection, Zagreb, 10000, Croatia 3University Hospital Centre Zagreb, Department of Nuclear Medicine, Zagreb, 10000, Croatia 4School of Physics, University of Sydney, Sydney, NSW 2006, Australia 5 Instituto de Física Corpuscular (IFIC), CSIC-UV, C/ Catedrático José Beltrán, 2, Paterna, Spain June 25, 2026 Abstract In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters.

25804#pg2>

[22] Ivo Rausch, Jacobo Cal-González, Daniela Dapra, Hans J. Gallowitsch, Peter Lind, Thomas Beyer, and Geraldine Minear: Performance evaluation of the biograph mct flow pet/ct system according to the nema nu2-2012 standard <ref:2606.25804#pg2>

[23] G. Romanchek, G. Shoop, K. Gholami, E. Enlow, and S. Abbaszadeh: Quantum Entanglement Filtering: A PET feasibility study in imaging dual-positron and prompt gamma emission via Monte Carlo simulation <ref:2606. p. 1

Where the paper says it

Keywords: Positron Emission Tomography; quantum entanglement; photon polarization; NEMA NU-4 phantom; spatial resolution; image quality Introduction Positron emission tomography (PET), an irreplaceable diagnostic tool for medical imaging, is based on the coincident detection of two gamma photons originating from positron annihilation in patient’s body. The gamma photons are typically detected by scintillation detectors capable of measuring photon energy, time of arrival and hit position.

25804#pg2>

[24] Dennis R Schaart: Physics and technology of time-of-flight PET detectors <ref:2606.25804#pg2>

[25] Hartland S. Snyder, Simon Pasternack, and J. Hornbostel: Angular correlation of scattered annihilation radiation <ref:2606. p. 0

Where the paper says it

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek 1 , Ana Marija Kožuljević1,2, Tomislav Bokulić 1 , Darko Grošev 3 , Zdenka Kuncic 4 , Siddharth Parashari 5 , and Luka Pavelić 2 1University of Zagreb Faculty of Science, Department of Physics, Zagreb, 10000, Croatia 2 Institute for Medical Research and Occupational Health, Division of Radiation Protection, Zagreb, 10000, Croatia 3University Hospital Centre Zagreb, Department of Nuclear Medicine, Zagreb, 10000, Croatia 4School of Physics, University of Sydney, Sydney, NSW 2006, Australia 5 Instituto de Física Corpuscular (IFIC), CSIC-UV, C/ Catedrático José Beltrán, 2, Paterna, Spain June 25, 2026 Abstract In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters.

25804#pg2>

[26] Giulia Terragni, Vanessa Nadig, Elena Tribbia, Stefano di Gangi, Ekaterini Toumparidou, Thomas Meyer, Johann Marton, Volkmar Schulz, Stefan Gundacker, Marco Pizzichemi, and Etiennette Auffray: Exploring the performance of a doi-capable tof-pet module using different sipms <ref:2606.25804#pg2>

[27] Toghyani M. et al.: Polarisation-based coincidence event discrimination: An in silico study towards a feasible scheme for Compton-PET <ref:2606. p. 0

Where the paper says it

To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters. We present a detailed study of the imaging of two 68Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality.

25804#pg2>

[28] Watts D.P. et al.: Photon quantum entanglement in the MeV regime and its application in PET imaging <ref:2606. p. 1

Where the paper says it

Keywords: Positron Emission Tomography; quantum entanglement; photon polarization; NEMA NU-4 phantom; spatial resolution; image quality Introduction Positron emission tomography (PET), an irreplaceable diagnostic tool for medical imaging, is based on the coincident detection of two gamma photons originating from positron annihilation in patient’s body. The gamma photons are typically detected by scintillation detectors capable of measuring photon energy, time of arrival and hit position.

25804#pg2>

[29] Wettenhovi V. et al.: OMEGA—open-source emission tomography software <ref:2606. p. 5

Where the paper says it

The signal-to-random background ratio was then defined as: SBR = (T − R)/R. Image reconstruction and assessment Image reconstruction was performed using the open-source software OMEGA[29] in MATLAB.

25804#pg2>

[30] C. S. Wu and I. Shaknov: The angular correlation of scattered annihilation radiation <ref:2606. p. 0

Where the paper says it

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek 1 , Ana Marija Kožuljević1,2, Tomislav Bokulić 1 , Darko Grošev 3 , Zdenka Kuncic 4 , Siddharth Parashari 5 , and Luka Pavelić 2 1University of Zagreb Faculty of Science, Department of Physics, Zagreb, 10000, Croatia 2 Institute for Medical Research and Occupational Health, Division of Radiation Protection, Zagreb, 10000, Croatia 3University Hospital Centre Zagreb, Department of Nuclear Medicine, Zagreb, 10000, Croatia 4School of Physics, University of Sydney, Sydney, NSW 2006, Australia 5 Instituto de Física Corpuscular (IFIC), CSIC-UV, C/ Catedrático José Beltrán, 2, Paterna, Spain June 25, 2026 Abstract In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters.

25804#pg2>

--- Page 1 ---

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek, Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Zdenka Kuncic, Siddharth Parashari, and Luka Pavelić University of Zagreb Faculty of Science <ref:2606.25804#pg2>

June 25, 2026

Abstract In Positron Emission Tomography a potential yet unutilized enhancement may come from exploiting the quantum entanglement of the annihilation quanta inscribed in the correlation of their polarizations <ref:2606.25804#pg2> To investigate this we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering based on single-layer scintillator polarimeters <ref:2606. p. 0

Where the paper says it

Positron Emission Tomography with quantum-entangled Compton events: first imaging results at clinically relevant activities Mihael Makek 1 , Ana Marija Kožuljević1,2, Tomislav Bokulić 1 , Darko Grošev 3 , Zdenka Kuncic 4 , Siddharth Parashari 5 , and Luka Pavelić 2 1University of Zagreb Faculty of Science, Department of Physics, Zagreb, 10000, Croatia 2 Institute for Medical Research and Occupational Health, Division of Radiation Protection, Zagreb, 10000, Croatia 3University Hospital Centre Zagreb, Department of Nuclear Medicine, Zagreb, 10000, Croatia 4School of Physics, University of Sydney, Sydney, NSW 2006, Australia 5 Instituto de Física Corpuscular (IFIC), CSIC-UV, C/ Catedrático José Beltrán, 2, Paterna, Spain June 25, 2026 Abstract In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters.

25804#pg2> We present a detailed study of the imaging of two 68Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality <ref:2606. p. 0
Where the paper says it

To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters. We present a detailed study of the imaging of two 68Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality.

25804#pg2> The results show that a spatial resolution of 2. p. 0
Where the paper says it

We present a detailed study of the imaging of two 68Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality. The results show that a spatial resolution of 2.

5±0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.6 ±0.3 mm to 4.9 ±0.3 mm, depending on data selection criteria <ref:2606.25804#pg2> We also found that the polarization-correlated Compton events exhibit up to 20% higher average signal to random background ratio compared to the single-pixel events <ref:2606. p. 0
Where the paper says it

3 mm, depending on data selection criteria. We also found that the polarization-correlated Compton events exhibit up to 20% higher average signal to random background ratio compared to the single-pixel events.

25804#pg2> We also present the first imaging of the NEMA NU-4 phantom filled with a 68Ga solution of 378 MBq initial activity, successfully combining polarization-correlated events with conventional single-pixel event selection <ref:2606. p. 0
Where the paper says it

We also found that the polarization-correlated Compton events exhibit up to 20% higher average signal to random background ratio compared to the single-pixel events. We also present the first imaging of the NEMA NU-4 phantom filled with a 68Ga solution of 378 MBq initial activity, successfully combining polarization-correlated events with conventional single-pixel event selection.

25804#pg2> Based on the extracted spatial resolution, signal-to-background, signal-to noise, contrast, and contrast-to noise ratio we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events while preserving a high image quality <ref:2606. p. 0
Where the paper says it

Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarizationcorrelated events, while preserving a high image quality. 1 arXiv:2606.

25804#pg2>

Pruning redundant reasoning tokens

The results show that a spatial resolution of 2.5±0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.6 ±0.3 mm to 4.9 ±0.3 mm, depending on data selection criteria <ref:2606.25804#pg2>.

How the training signal is built

The DAQ process was automated in such a way that the acquisition at one position was followed by a precise rotation to the next position From the full data set, we selected events where both gamma photons undergo Compton scattering within the respective module by following these requirements:

  1. exactly two pixels fire (Epx > 120 keV) in a module

  2. the energy of the 511 keV gamma is shared between these two pixels, requiring their sum to be within ±3σ of the 511 keV peak maximum

  3. energy depositions correspond to Compton scattering kinematics, as described by Parashari et al. 2022.

Where the gains come from

The main reason for that is the uncertainty in the Compton interaction order, which results in uncertainty of assigning LOR to correct pixel position <ref:2606.25804#pg8> Consequently, when the d1,2 range is smaller, this uncertainty is also lower, and the resulting resolution is better <ref:2606.25804#pg8>. The benefit of adding the polarization-correlated events is a lower contamination of the sample with randoms than the single-pixel events <ref:2606. p. 0

Where the paper says it

The results show that a spatial resolution of 2. 1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.

25804#pg9>. Since the resolution improvement and statistical gain using PCE are reversely proportional, due to the current LOR assignment method, the trade-off is necessary <ref:2606. p. 6
Where the paper says it

The slight decrease in resolution notable in the PCE image with 620 mm ring diameter is expected due to the more pronounced non-collinearity effect at larger diameters. The poorer spatial resolution obtained with PCE, is assumed to stem from the uncertainty in the assignment of the first fired pixel in Compton events[10].

25804#pg17>.

We have successfully constructed and operated a novel detector setup to demonstrate and assess the feasibility of using quantum-entanglement of the annihilation quanta in PET imaging by exploiting their pronounced polarization correlations First images of sources with up to 378 MBq activities were successfully reconstructed using solely polarizationcorrelated events with spatial resolution below 5 mm in the central field of view. While the reduced spatial resolution of the inter-crystal Compton event images, currently poses a barrier to achieve the full benefit of the quantum-entanglement enhanced PET imaging, the results still suggest that an improvement in sensitivity up to 10% may be reached, while keeping or even slightly improving the image quality. p. 2

Where the paper says it

This work reports the performance of demonstrational setup, based on the single-layer concept for implementing polarization measurement in PET, and for the first time, presents the corresponding PET images of source distributions with clinically relevant activities, obtained by utilizing polarization-correlated Compton events. Further, we compare these to images obtained by conventional single-pixel event selection in terms of spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise, and estimate a sensitivity gain by exploiting the polarization-correlated inter-crystal scatter events.

Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.

Improvements for AI systems

  1. Develop an AI system that uses polarization-correlated Compton events to enhance PET image reconstruction by incorporating a modulation factor mu into its likelihood function, as described by Ncor(phi1 - phi2) = M[1 − µcos[2(phi1 − phi2)] (5). This system can reconstruct images with spatial resolutions up to 4.9 mm, potentially offering up to 10% sensitivity increase over single-pixel events.

  2. Create an AI model trained on the comparison between Single-Pixel Events (SPE) and Polarization-Correlated Events (PCE) that learns optimal event selection criteria based on LOR gain and inter-crystal scatter distance d1,2 range. This improved system can automatically select event subsets to maximize the signal to background ratio (SBR) while maintaining image quality metrics like Contrast and CNR.

  3. Implement a machine learning algorithm, such as a deep neural network, for automated LOR assignment based on event topology that accounts for the uncertainty in assigning the first fired pixel in Compton events. This capability allows reconstruction methods to overcome the uncertainty in the assignment of LOR to correct pixel position, potentially leading to better spatial resolution than current methods.

  4. Design an AI-driven image quality assessment module that utilizes reconstructed images from merged SPE and PCE samples to quantify improvements in Signal to Noise (SNR) and Contrast-to-Noise Ratio (CNR) as a function of the PCE contribution. This module can provide real-time feedback on the benefit of using polarization information for specific clinical activities.

Abstract

In Positron Emission Tomography, a potential, yet unutilized enhancement, may come from exploiting the quantum entanglement of the annihilation quanta, inscribed in the correlation of their polarizations. To investigate this, we built a PET demonstrator capable of measuring polarization correlations of annihilation quanta by their Compton scattering, based on single-layer scintillator polarimeters. We present a detailed study of the imaging of two 68 Ge line sources, 45 MBq each, to extract the spatial resolution and assess image quality. The results show that a spatial resolution of 2.5 plus or minus 0.1 mm is obtained using single-pixel events, while resolutions obtained with polarization-correlated Compton events range from 3.6 plus or minus 0.3 mm to 4.9 plus or minus 0.3 mm, depending on data selection criteria. We also found that the polarization-correlated Compton events exhibit up to 20% higher average signal to random background ratio compared to the single-pixel events. We also present the first imaging of the NEMA NU-4 phantom filled with a 68 Ga solution of 378 MBq initial activity, successfully combining polarization-correlated events with conventional single-pixel event selection. Based on the extracted spatial resolution, signal-to-background, signal-to-noise, contrast, and contrast-to-noise ratio, we estimate that up to 10% sensitivity increase may be attained by exploiting the polarization-correlated events, while preserving a high image quality.