OASIS-ERIS: Open Acquisition System for IEPE Sensors - Electrically Refined and Inter-device Synchronization

summary

The gist

OASIS-ERIS introduces an Open Acquisition System for IEPE Sensors that enhances laboratory usability through inter-device synchronization and increases achievable sampling frequency to 100 kHz…

In short

OASIS-ERIS is an open-source system for IEPE sensors that improves lab usability through synchronization and increases sampling frequency to 100 kHz, matching commercial performance. It uses custom hardware and software stacks to provide a low-cost alternative for experimental modal analysis.

Key concepts

Inter-device Synchronization
This feature physically connects sample clocks between multiple acquisition boards using BNC connectors. A 'Sync Source' broadcasts a hardware clock signal via OneWire, allowing all connected 'Sync Sinks' to acquire data at the exact same time, crucial for accurate modal analysis.
IEPE Sensors
IEPE stands for Integrated Excitation Potentiostat electrodes. These are specialized sensors used in experimental modal analysis to measure mechanical vibrations. OASIS-ERIS is designed specifically to interface with these sensors efficiently.
OASIS-API
This Python API serves as the software bridge, handling all serial communication with the OASIS boards. It manages robust sampling routines and organizes data into structured HDF5 files, making it easy to integrate into existing analysis frameworks.
Frequency Response Function (FRF) Estimation
FRF estimation is a technique used in experimental modal analysis to determine the dynamic characteristics of a structure by measuring how it responds to various inputs. OASIS-ERIS aims to match commercial systems for this estimation when signal levels are high.

Terminology used across episodes

This episode discusses

The paper

OASIS-ERIS: Open Acquisition System for IEPE Sensors - Electrically Refined and Inter-device Synchronization · Read on arXiv

Technical University of Munich

OASIS-ERIS, the Open Acquisition System for IEPE Sensors - Electrically Refined and Inter-device Synchronization, is the successor to the previously released OASIS-UROS. It is based on the Analog Devices AD7606C-18 analog-to-digital converter, offering 18-bit resolution on eight simultaneously sampled channels, and the Espressif ESP32-S3 microcontroller. The laboratory usability of OASIS-ERIS is significantly improved compared to OASIS-UROS due to the added inter-device synchronization. By physically connecting the sample clocks across devices, a truly synchronized data acquisition can be achieved with an arbitrary number of acquisition boards. Additionally, a migration of the firmware to ESP-IDF allows a major increase of the achievable sampling frequency to 100 kHz. This paper documents the hardware and software stack of OASIS-ERIS, and provides all materials necessary for reproduction. Further, the performance of OASIS-ERIS is tested against a commercial measurement system in an Experimental Modal Analysis campaign. The results show that OASIS-ERIS closely matches the performance of the commercial system, making it a viable replacement for academia and projects with tightly constrained budgets.

DOI: 10.1016/j.ohx.2026.e00855

Transcript

Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.

Kai: I'm Kai.

Mira: And I'm Mira, here with Lev.

Kai: We're looking at the paper: "OASIS-ERIS: Open Acquisition System for IEPE Sensors - Electrically Refined and Inter-device Synchronization".

Mira: Here's the gist of the paper. OASIS-ERIS introduces an Open Acquisition System for IEPE Sensors that enhances laboratory usability through inter-device synchronization and increases achievable sampling frequency to 100 kHz because it closely matches commercial system performance in Experimental Modal Analysis while offering a low-cost, open-source alternative.

Kai: Hardware Description The OASIS-ERIS board utilizes an Analog Devices AD7606C-18 analog-to-digital converter providing 18-bit resolution on eight simultaneously sampled channels and is controlled by the Espressif ESP32-S3 microcontroller.

Mira: Key hardware features include an Improved Boost Converter based on the Texas Instruments LM5157 for a 24 V IEPE supply and a Simplified IEPE Front-End using Texas Instruments LM334M constant current sources.

Kai: The board incorporates internal USB Connector embedding into Linux ecosystems, an Improved WiFi Interface with a refined on-board antenna, and a Local Sample Cache allowing for fully autonomous device operation.

Paper summary: Kai: So, we're looking at this paper, "OASIS-ERIS: Open Acquisition System for IEPE Sensors - Electrically Refined and Inter-device Synchronization." It's about taking an existing setup and making it better with synchronization between different sensors.

Mira: Exactly. The main idea here is that they developed this new system to improve how we use IEPE sensors in the lab by getting them all to work together in a truly synchronized way. They claim this open acquisition system, OASIS-ERIS, can increase the sampling frequency up to one hundred kHz because they switched the firmware implementation <ref:2607.22061#pg1>.

Lev: From an error correction standpoint, that jump to one hundred kHz is significant when you think about real hardware overhead <ref:2607.22061#pg1>. We need to see how much that increased speed actually impacts the stability of our modal parameters, especially when we're dealing with structural dynamics.

Kai: Right, so it’s not just about speed; they built this board using the Analog Devices AD7606C-eighteen analog-to-digital converter with eighteen bits of resolution on eight channels, controlled by an ESP32 S3 microcontroller <ref:2607.22061#pg1,the Analog Devices AD7606C-18 analog-to-digital converter>. They also swapped out some parts for a better boost converter and a simplified front-end using Texas Instruments LM334M constant current sources.

Mira: And the real focus of this paper seems to be that inter-device synchronization feature. They physically connect the sample clocks across devices using BNC connectors so they can achieve truly synchronized data acquisition with an arbitrary number of boards.

Lev: That physical connection mechanism sounds like it would introduce some complexity in a real deployment, but if it works, it’s a way to get high-fidelity results from multiple sensors simultaneously without relying on software timing alone.

Kai: The validation tests they did showed that this synchronization test resulted in a maximum sample offset of one sample or ten microseconds for normal sampling. Plus, when they compared it against a commercial measurement system in Experimental Modal Analysis, OASIS-ERIS closely matched the performance for frequency response function estimation if the signal level is sufficiently high.

Mira: So, they're saying that even with this open-source approach and the new synchronization method, you can get results that are comparable to what you'd see on a commercial system when your signal is strong enough.

Lev: But I wonder about the assumptions behind matching performance; does it hold up if we push the signal level down, or if we have more than just two boards synchronized?

Kai: They also detailed the software stack evolution, moving from Arduino ESP32 to a direct C++ implementation using ESP-IDF for faster hardware access and those higher sample rates. They have several components there, including a Python API called OASIS-API that handles serial communication and structured HDF5 data storage.

Paper summary: Mira: And they have the OASIS-TUI, which is this guided terminal user interface optimized for remote operation via SSH, and then the OASIS-GUI for classical lab use. It seems they've covered the whole spectrum from remote control to traditional lab setup.

Lev: I’m interested in that Python API part because that’s where you actually have to write the code to handle all those structured data saves, like HDF5, and see how much overhead that adds on top of the hardware processing speed they're promising.

Kai: And for data storage, stored acquisition files are saved on an exFAT SD card using commands like ACQ:SAMPLE and meta files are also saved there. Stored measurement transfer over CDC serial is disabled; you have to get the data out through USB MSC instead.

Mira: That’s a specific detail about the file system interaction, which is important because it dictates how easily you can retrieve those long acquisition records for analysis later on.

Lev: So, what’s the real limitation here from an engineering standpoint? I want to know where this setup struggles when you try to run it on actual hardware versus in a simulated environment.

Kai: The paper points out that the bill of materials cost is about one hundred forty to one hundred sixty euros, which makes it very accessible for projects with tight budgets compared to proprietary systems. They also provided design files in the Zenodo repository including KiCad project and manufacturing files, which speaks to their commitment to openness.

Mira: The paper emphasizes the open-source aspects of this work, mentioning Openness, Freedom, Flexibility and Customization as key benefits for anyone who wants to audit the source code and modify it.

Lev: Auditing is one thing; running it on real hardware with that specific synchronization setup is another. I need to know how much of the complexity lands on the user interface side versus needing custom drivers for their specific IEPE sensor setup.

Kai: They did confirm that they can trigger acquisition by any acquire command where a master board broadcasts the acquisition duration, sampling rate, and filename base over GPIO13 one-wire to listening boards.

Mira: That distributed one-Wire Acquisition sounds like it solves the problem of getting multiple boards to start and stop at exactly the same time, which is a major usability win for complex modal testing <ref:2607.22061#pg1>.

Lev: If the master board broadcasts parameters over one wire, that’s a relatively simple communication protocol, which is good for robustness on hardware. But how reliable is that one-wire broadcast across different physical setups?

Paper summary: Kai: The experimental validation summary confirms that inter-device synchronization with a maximum offset of one sample or ten microseconds was possible, and for triggered sampling, all boards are always perfectly synchronized.

Mira: That perfect synchronization result is what really validates their core claim about the inter-device feature; they showed that even with this method working, the cross-correlation analysis showed that the maximum sample offset across all boards and tests was one for normal sampling at a sample rate of one hundred kHz <ref:2607.22061#pg1>.

Lev: One sample offset is tight for high-speed acquisition; that tells me how much timing jitter we are dealing with when we try to run this on a real system, and it’s a good number to target in our hardware design.

Kai: When comparing the modal identification results against the commercial system, they found barely any difference in identified eigenfrequencies between OASIS-ERIS and the commercial system. They also noted that the exact match of the identified mode shapes indicates that inter-device synchronization is successful, requiring responses from all sensors connected to different OASIS-ERIS boards.

Mira: So, it’s not just matching the frequency; it’s matching the actual shape of the structural response when you link all those boards together, which is a deeper level of validation than just a frequency plot.

Lev: That means if we use this system, we can trust that the modal parameters derived from our data are consistent across all sensors involved in the experiment, provided they’re all linked correctly.

Kai: The paper concludes by presenting OASIS-ERIS as a viable replacement for projects with tightly constrained budgets because of its open-source license and its ability to closely match commercial performance. They highlight the benefits of using this system for day-to-day lab use.

Mira: The implication is that we don't have to commit to expensive proprietary systems just to get high sampling rates and synchronized data acquisition for our structural dynamic applications.

Lev: It’s a solid piece of open hardware, but the caveat remains that it depends heavily on the user setting up the physical synchronization correctly, which is a manual assembly task they mentioned requiring BNC and SMB connectors.

Kai: So, to wrap up this paper on "OASIS-ERIS: Open Acquisition System for IEPE Sensors - Electrically Refined and Inter-device Synchronization," it’s an open system that aims to bridge the gap between commercial performance and accessible, open hardware for structural analysis.

Mira: It shows that with the right synchronization mechanism, you can achieve high sampling rates like one hundred kHz while keeping the data acquisition process highly flexible and synchronized across multiple devices <ref:2607.22061#pg1>.

Lev: The real challenge we see is making sure that the physical synchronization works reliably in a real-world environment without excessive setup time for each experiment.

Conclusion: Kai: It’s about taking something existing and making it open source so people can actually use it in the lab without paying a fortune for proprietary gear.

Lev: Yeah, the authors are focused on getting high performance—one hundred kilohertz sampling—while keeping the hardware accessible because they use an open license for the code.

Mira: But what does that actually mean for us? It means we can finally get a system that matches commercial results in things like experimental modal analysis without needing to buy a custom, expensive setup from some big vendor.

Kai: It’s about closing the gap between what you can build yourself and what the commercial systems do when the signal is strong enough.

Lev: The real thing is they managed to get this synchronization working physically across different devices using those BNC connectors, which is a big step because software timing alone doesn't always cut it at that speed.

Mira: So, it’s not just about the hardware being cheap; it’s about the method of synchronization allowing all those sensors to actually talk to each other and give you consistent data.

Kai: Exactly. It means for anyone doing structural dynamics research, you can have a system that works reliably without being locked into one specific brand's ecosystem.

Lev: It’s a solid piece of hardware with one hundred sixty euros worth of materials, which is low cost, but the setup still requires manual assembly with those connectors.

Mira: So the takeaway is that this work shows you can get high sampling rates and reliable cross-device synchronization without needing a massive budget or proprietary lock-in.

Kai: It’s a step toward making sophisticated acquisition systems available for regular lab use instead of just being locked away in specialized research facilities.

Lev: And the next thing we need to look at is how much real overhead that physical synchronization adds when you try to run it on actual hardware versus just simulating the timing.

More episodes

← Home