Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom
Listen
Radio episode about this paper
Transcript
Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.
Kai: We're looking at the paper: "Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom".
Mira: Here's the gist of the paper. The goal of this study was to utilize an anthropomorphic phantom to: (1) evaluate image noise against the prescribed NI across different organs, (2) determine if DLIR reduces image noise compared to ASiR-V, and (3) use this relationship to determine the maximum volume CT dose index (CTDIvol) reduction at equivalent noise levels Study Purpose The primary objective of the research was to employ an anthropomorphic phantom to achieve three specific aims: (1) evaluate image noise against the prescribed NI across different organs, (2) determine if DLIR reduces image noise compared to ASiR-V, and (3) use this relationship to determine the maximum potential dose reduction at equivalent noise levels The study involved acquiring repeat images on a GE HealthCare Revolution Apex CT scanner across various NI values in an anthropomorphic adult-sized phantom from Kyoto Kagaku using both ASiR-V and DLIR reconstructions Organs of interest were contoured using TotalSegmentator in 3D Slicer The noise metric (NM) was defined as the standard deviation of the subtraction image across the two repetitions divided by the square root of 2 to account for noise propagation The NM-to-NI ratio was defined as the computed NM divided by the prescribed NI Key Findings on Noise and Dose The relationship between NI and CTDIvol was found to be proportional to a power law of exponent-2.23 The NM was dependent on NI using a power law with exponent under 1 with smaller exponents for DLIR The maximum potential absolute dose reduction from using DLIR was found in the liver at 1.3 mGy (approximately 22% of the corresponding ASiR-V CTDIvol) with an NI of 7.9 Across all organs, the relative dose reduction increased, and the noise reduction increased across NI Organ-Specific Dose Reductions The resulting impacts on dose as measured using the CTDIvol can be seen in Figure 7 The dose reduction possible when using DLIR reconstruction to achieve the equivalent NM as the ASiR-V reconstruction is maximized in the liver with a reduction of 1.3 mGy and minimized in the spinal cord with a reduction of 0.2 mGy The optimal NI for the gallbladder is 12.7 and 15.0 for the spinal cord Relationship Between Reconstruction Techniques The relationship between the noise reduction ratio and the prescribed NI is shown in Figure 8 All relationships were positive, meaning that increasing the prescribed NI increases the noise reduction capabilities of the DLIR reconstruction compared to the ASiR-V reconstruction The steepest slope was seen in the liver at 1.02 while the shallowest was seen in the gallbladder at 0.69 indicating their susceptibility to noise reduction as the prescribed NI increases Limitations and Future Directions There were several limitations to this study, including scanning only the inferior portion of the chest, entire abdomen, and superior portion of the pelvis in the phantom The study also only investigated a single slice thickness and a pitch of 1.375 Future work should investigate how these results fit into the broader scope of the diagnostic imaging practice and more in-depth analysis into the noise profiles of each organ using metrics like the noise power spectrum This study explored the effects of varying acquired CT dose levels, in terms of the NI, on the reconstructed image noise, in terms of the NM, in an anthropomorphic phantom with realistic organ volumes and respective CT numbers using the 50% ASiR-V and DLIR reconstruction algorithms Several observations were observed: (1) sharper edges were seen in the noise image at the lowest NI setting in both the ASiR-V and DLIR images, (2) CTDIvol related to NI by a power law with an exponent of-2.23, (3) the ratio of the maximum to minimum mA falls monotonically from 1.53 at a prescribed NI of 5 to 1.18 at a prescribed NI of 30, (4) the relationship between the computed NM and the prescribed NI is not linear, (5) the DLIR reconstruction provided additional noise suppression at the equivalent NI compared to ASiR-V, (6) dose reduction when using DLIR compared to ASiR-V is organ-specific, and (7) the NM-to-NI ratio was found to decrease as the prescribed NI increased for both the ASiR-V and DLIR images References 1.
Kai: Kopka L, Funke M, Breiter N, Hermann KP, Vosshenrich R, Grabbe E.
An anatomically adapted variation of the tube current in CT. [Kai: Studies on radiation dosage reduction and image quality].
Title and authors: Kai: We’ve seen that they are using an anthropomorphic adult-sized phantom from Kyoto Kagaku to evaluate image noise against a prescribed NI across different organs, which sets up their investigation into how these reconstruction algorithms perform in practice.
Mira: And the core of the paper is determining if DLIR reduces image noise when compared to ASiR-V, and then using that relationship to figure out the maximum possible volume CT dose index reduction at equivalent noise levels.
Lev: That leads us straight into their purpose: they want to find a direct link between how much noise is present and the prescribed NI setting, and then see how DLIR interacts with that relationship across different tissues.
Kai: It’s about mapping out this relationship across organs, which is interesting because we know dose reduction isn't uniform everywhere.
Mira: They also mention they contoured the organs of interest using TotalSegmentator in three dee Slicer, which means they are being specific about where they are measuring these noise differences—liver, spinal cord, gallbladder, and so on <ref:2611.07357#pg1>.
The paper's summary: Kai: So, to summarize what the study found is that there’s a clear relationship between the CT dose index volume and the prescribed NI setting; specifically, it follows a power law with an exponent of negative two point two three.
Mira: And they also found that image noise, which they define as the standard deviation of the subtraction image divided by the square root of two to account for noise propagation, is dependent on that prescribed NI using a power law with an exponent under one for DLIR, which is smaller than what’s seen for ASiR-V.
Lev: That’s interesting because if DLIR has a smaller exponent, it suggests that the AI reconstruction method might handle the noise suppression scaling differently depending on the specific organ you are looking at.
Kai: They also found that the maximum potential absolute dose reduction from using DLIR was found to be one point three mGy in the liver when they were at an NI of seven point nine which is roughly twenty-two percent of what ASiR-V achieved at that same NI.
Mira: They also noted that across all organs, both the relative dose reduction and the noise reduction increase as the prescribed NI goes up, but the way these two things scale with NI is not linear.
The paper's improvements: Kai: The paper suggests a few ways this work could be improved. For example, they point out that while DLIR provides additional noise suppression at the equivalent NI compared to ASiR-V, the benefit of using DLIR actually increases as the prescribed NI for ASiR-V reconstruction goes up.
Mira: And they also highlight that there are specific optimal NI values for certain organs; for instance, they found that twelve point seven and fifteen point zero are the optimal NI settings for the gallbladder and spinal cord, respectively.
Lev: From a hardware standpoint, those specific NI suggestions are really useful because it tells us exactly where to dial in our reconstruction parameters to get the best noise performance for a particular anatomy.
Kai: They also suggest that if you want to get even deeper into this, you should look at how the NM-to-NI ratio changes with the prescribed NI, and they find that this ratio actually decreases as the prescribed NI increases for both ASiR-V and DLIR images.
Mira: That relationship between the computed noise metric and the prescribed NI isn't linear, which means we can't just assume a simple straight line to predict performance everywhere.
Conclusion: Kai: So, to wrap up on this study of "Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom," the main implication is that noise reduction isn't uniform across the body when adjusting NI settings.
Mira: The study confirms that DLIR does provide extra noise suppression compared to ASiR-V at the same NI, but it’s highly dependent on the organ, with the liver showing a bigger dose reduction benefit than the spinal cord in this specific test.
Lev: For us building real hardware, those organ-specific dose reductions mean we need tailored approaches for setting our reconstruction parameters based on what's being scanned.
Kai: It’s important to remember that while they found these relationships, they did have limitations; specifically, the study only scanned the inferior portion of the chest, the entire abdomen, and a superior portion of the pelvis in this phantom.
Mira: And they also had to keep things simple by only investigating a single slice thickness and a pitch of one point three hundred seventy-five during their testing.
Lev: That’s where we can take this work next; we need to see how these relationships hold up when we move into more complex scenarios with variable slice thickness or different acquisition parameters.
Kai: That sounds like the direction for future research, moving from the phantom setup to a broader clinical context, which is where the real challenge lies.
Department of Radiology, Boston Children’s Hospital · Harvard Medical School · §GE HealthCare
physics.med-ph
Submitted: 2026-11-05
Updated: 2026-11-05
Code: https://github.com/LucasMc/CT-Noise
License: http://creativecommons.org/licenses/by/4.0/
Importance score: 82/100
The gist: The goal of this study was to utilize an anthropomorphic phantom to: (1) evaluate image noise against the prescribed NI across different organs…
Key concepts
- Image Noise (NM)
- This metric measures the deviation or randomness in an image subtraction process across two repeated scans. It quantifies how much random variation exists in the reconstructed image pixels. A lower NM indicates a cleaner, less noisy image.
- Prescribed NI
- The Prescribed Noise Index (NI) is a target level of noise that researchers aim to achieve or control during CT scanning. The study examined how different reconstruction methods perform when aiming for various NI targets across different body parts.
- DLIR vs. ASiR-V
- These are two different image reconstruction algorithms tested on the same CT data. DLIR (Deep Learning Image Reconstruction) was compared to ASiR-V (Adaptive Statistical Iterative Reconstruction). The goal was to see if DLIR provides superior noise reduction capabilities for achieving a target NI.
- CTDIvol
- CTDIvol is a measure of the radiation dose delivered by the CT scanner, specifically volume-averaged. The study investigated how using DLIR reconstruction at equivalent noise levels impacts this dose index, aiming to find maximum possible dose reductions.
Terminology
Summary
The goal of this study was to utilize an anthropomorphic phantom to: (1) evaluate image noise against the prescribed NI across different organs, (2) determine if DLIR reduces image noise compared to ASiR-V, and (3) use this relationship to determine the maximum volume CT dose index (CTDIvol) reduction at equivalent noise levels
Study Purpose
The primary objective of the research was to employ an anthropomorphic phantom to achieve three specific aims: (1) evaluate image noise against the prescribed NI across different organs, (2) determine if DLIR reduces image noise compared to ASiR-V, and (3) use this relationship to determine the maximum potential dose reduction at equivalent noise levels
Methodology and Metrics
The study involved acquiring repeat images on a GE HealthCare Revolution Apex CT scanner across various NI values in an anthropomorphic adult-sized phantom from Kyoto Kagaku using both ASiR-V and DLIR reconstructions Organs of interest were contoured using TotalSegmentator in 3D Slicer The noise metric (NM) was defined as the standard deviation of the subtraction image across the two repetitions divided by the square root of 2 to account for noise propagation The NM-to-NI ratio was defined as the computed NM divided by the prescribed NI
Key Findings on Noise and Dose
The relationship between NI and CTDIvol was found to be proportional to a power law of exponent-2.23 The NM was dependent on NI using a power law with exponent under 1 with smaller exponents for DLIR The maximum potential absolute dose reduction from using DLIR was found in the liver at 1. p. 1 The NM was dependent on NI using a power law with exponent
under 1 with smaller exponents for DLIR. The maximum potential absolute dose reduction from
using DLIR was found in the liver at 1. The maximum potential absolute dose reduction from
using DLIR was found in the liver at 1. 8, 95% confidence interval] mGy
(approximately 22% of the corresponding ASiR-V CTDIvol) with an NI of 7.Where the paper says it
Where the paper says it
Organ-Specific Dose Reductions
The resulting impacts on dose as measured using the CTDIvol can be seen in Figure 7 The dose reduction possible when using DLIR reconstruction to achieve the equivalent NM as the ASiR-V reconstruction is maximized in the liver with a reduction of 1.3 mGy and minimized in the spinal cord with a reduction of 0. p. 1 The spinal cord and gallbladder were least impacted by DLIR. Across all
organs, the relative dose reduction increased, the NM-to-NI ratio decreased, and the noise
reduction increased across NI. Recently, Kyoto Kagaku (Kyoto, Japan) has developed
anthropomorphic phantoms which can be imaged using CT scanners that provides realistic
models of organs and their associated expected CT numbers10
. While scanner and organ agnostic studies have been explored to study the optimal tradeoffs of
image noise and dose11, limited work has employed this methodology to study the behavior
specifically on the GE HealthCare CT scanners using an anthropomorphic phantom which can
realistically simulate noise behavior across organs.Where the paper says it
Where the paper says it
Relationship Between Reconstruction Techniques
The relationship between the noise reduction ratio and the prescribed NI is shown in Figure 8 All relationships were positive, meaning that increasing the prescribed NI increases the noise reduction capabilities of the DLIR reconstruction compared to the ASiR-V reconstruction The steepest slope was seen in the liver at 1.02 while the shallowest was seen in the gallbladder at 0. p. 4 The average tube current across all slices
was reported due to the use of SmartmA. A summary of the final acquisition parameters can be
seen in Table 1 while a visualization of the varying tube current can be seen in Supplementary
Figure S1.Where the paper says it
Limitations and Future Directions
There were several limitations to this study, including scanning only the inferior portion of the chest, entire abdomen, and superior portion of the pelvis in the phantom The study also only investigated a single slice thickness and a pitch of 1.375 Future work should investigate how these results fit into the broader scope of the diagnostic imaging practice and more in-depth analysis into the noise profiles of each organ using metrics like the noise power spectrum
Conclusion
This study explored the effects of varying acquired CT dose levels, in terms of the NI, on the reconstructed image noise, in terms of the NM, in an anthropomorphic phantom with realistic organ volumes and respective CT numbers using the 50% ASiR-V and DLIR reconstruction algorithms Several observations were observed: (1) sharper edges were seen in the noise image at the lowest NI setting in both the ASiR-V and DLIR images, (2) CTDIvol related to NI by a power law with an exponent of-2.23, (3) the ratio of the maximum to minimum mA falls monotonically from 1. p. 4 1), the only setting that was changed across
acquisitions was the prescribed NI. Setting
Tube Voltage (kVp) 120
NI 5 7 8 9 10 13 15 17 19 21 30
Minimum Tube Current (mA) 546 264 197 153 121 69 50 39 30 24 11
Average Tube Current (mA) 690 328 243 187 147 82 59 45 35 28 12
Maximum Tube Current (mA) 837 397 291 223 173 94 67 51 39 31 13
Exposure Time (ms) 500 However, it is unclear how these
reconstruction algorithms reduce noise as NI changes, particularly across multiple organs. Purpose: The goal of this study was to utilize an anthropomorphic phantom to: (1) evaluate
image noise against the prescribed NI across different organs, (2) determine if DLIR reduces
image noise compared to ASiR-V, and (3) use this relationship to determine the maximum
volume CT dose index (CTDIvol) reduction at equivalent noise levels.Where the paper says it
Where the paper says it
References
-
Kopka L, Funke M, Breiter N, Hermann KP, Vosshenrich R, Grabbe E. [An anatomically adapted variation of the tube current in CT. p. 6
Studies on radiation dosage reduction and image quality]. p. 3Where the paper says it
Statistical Analysis All computational analysis was completed using Python 3. To assess the repeatability of the consecutive acquisitions, the coefficient of variation was used for the tube current values and averaged across each slice and compared across prescribed NI values.
RofoWhere the paper says it
Introduction Computed tomography (CT) has become a mainstay in current clinical practice due to its fast acquisition times, particularly in the pediatric setting. These fast acquisition times are also driven by CT’s use of ionizing radiation which should be minimized while achieving the required image quality, especially in the pediatric setting.
-
Lauritsch G, Haerer WH. Theoretical framework for filtered back projection in tomosynthesis
-
Stiller W. Basics of iterative reconstruction methods in computed tomography: A vendorindependent overview
-
Prakash P, Kalra MK, Kambadakone AK, et al. Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique
-
Thibault J, Sauer KD, Bouman CA, Hsieh J. A three‐dimensional statistical approach to improved image quality for multislice helical CT
-
Kwon H, Cho J, Oh J, et al. The adaptive statistical iterative reconstruction-V technique for radiation dose reduction in abdominal CT: comparison with the adaptive statistical iterative reconstruction technique
-
De Marco P, Origgi D. New adaptive statistical iterative reconstruction ASiR‐V: Assessment of noise performance in comparison to ASiR
-
Koetzier LR, Mastrodicasa D, Szczykutowicz TP, et al. Deep Learning Image Reconstruction for CT: Technical Principles and Clinical Prospects
-
Dixon R, Anderson J, Bakalyar D, et al. Comprehensive Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography
-
Ono K, Asada Y. Analysis of material composition and attenuation characteristics of anthropomorphic torso phantoms for dosimetry using dual energy CT technology
-
Zarei M, Abadi E, Segars WP, Samei E. p. 0
Coronary stenosis quantification in cardiac computed tomography angiography: multi-factorial optimization of image quality and radiation doseWhere the paper says it
1 Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom Lucas McCullum (0000-0001-9788-7987) 1,2 , Mojtaba Zarei (0000-0002-0997-7411) 3 , Andy Tsai (0000-0002-0089-6463) 1,2 , Michael J. Callahan (0000-0002-7833-4524) 1,2 , Peng Sun (0000-0002-9697-4656) 1,2‡ , Da Zhang (0000-0001-5092-5421) 1,2‡* 1Department of Radiology, Boston Children’s Hospital, 300 Longwood Avenue, Boston, MA, 02215, USA 2Harvard Medical School, 25 Shattuck Street, Boston, MA, 02115, USA 3GE HealthCare, Research and Scientific Affairs, Waukesha, WI, 53188, USA ‡ Shared last author *Corresponding Author: Da Zhang <Da.
-
Wasserthal J, Breit HC, Meyer MT, et al. TotalSegmentator: Robust Segmentation of 104 Anatomic Structures in CT Images
-
Isensee F, Jaeger PF, Kohl SAA, Petersen J, Maier-Hein KH. nnU-Net: a self-configuring method for deep learning-based biomedical image segmentation
-
Fedorov A, Beichel R, Kalpathy-Cramer J, et al. 3D Slicer as an image computing platform for the Quantitative Imaging Network
-
Tian X, Samei E. Accurate assessment and prediction of noise in clinical CT images
-
Steel RGD, Torrie JH. Principles and Procedures of Statistics: With Special Reference to the Biological Sciences
-
Scott AW, Alancherry ASP, Lee C, Eastman E, Zhou Y. Computed tomography dose index determination in dose modulation prospectively involving the third‐generation iterative reconstruction and noise index
-
Greffier J, Si-Mohamed S, Frandon J, et al. Impact of an artificial intelligence deep-learning reconstruction algorithm for CT on image quality and potential dose reduction: A phantom study
-
Yoo YJ, Choi IY, Yeom SK, et al. Evaluation of Abdominal CT Obtained Using a Deep Learning-Based Image Reconstruction Engine Compared with CT Using Adaptive Statistical Iterative Reconstruction
-
Wang S, Medrano MJ, Imran AAZ, et al. Automated estimation of individualized organ-specific dose and noise from clinical CT scans
-
Tzanis E, Damilakis J. A machine learning-based pipeline for multi-organ/tissue patient‐specific radiation dosimetry in CT
-
Myronakis M, Stratakis J, Damilakis J. Rapid estimation of patient‐specific organ doses using a deep learning network
-
Schmidt TG, Wang AS, Coradi T, Haas B, Star-Lack J. p. 5
Accuracy of patient-specific organ dose estimates obtained using an automated image segmentation algorithmWhere the paper says it
The remaining segmentations were not used either due to only partially being inside the imaging volume (i. , lung) or lack of confidence with accurate CT numbers (i.
-
Pace E, Caruana CJ, Bosmans H, Cortis K, D’Anastasi M, Valentino G. An inventory of patient-image based risk/dose, image quality and body habitus/size metrics for adult abdomino-pelvic CT protocol optimisation
-
Gandhi D, Crotty DJ, Stevens GM, Schmidt TG. Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ‐based tube current modulation technique
-
Christianson O, Winslow J, Frush DP, Samei E. Automated Technique to Measure Noise in Clinical CT Examinations
-
Ahmad M, Jacobsen MC, Thomas MA, Chen HS, Layman RR, Jones AK. A Benchmark for automatic noise measurement in clinical computed tomography
-
S J Riederer, N J Pelc, D A Chesler. The noise power spectrum in computed X-ray tomography
-
Lee C, Liu J, Griffin K, Folio L, Summers RM. Adult patient-specific CT organ dose estimations using automated segmentations and Monte Carlo simulations
-
Lambin P, Rios-Velazquez E, Leijenaar R, et al. Radiomics: Extracting more information from medical images using advanced feature analysis
-
Mackin D, Fave X, Zhang L, et al. p. 0
Measuring Computed Tomography Scanner Variability of Radiomics Features: Investigative RadiologyWhere the paper says it
1 Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom Lucas McCullum (0000-0001-9788-7987) 1,2 , Mojtaba Zarei (0000-0002-0997-7411) 3 , Andy Tsai (0000-0002-0089-6463) 1,2 , Michael J. Callahan (0000-0002-7833-4524) 1,2 , Peng Sun (0000-0002-9697-4656) 1,2‡ , Da Zhang (0000-0001-5092-5421) 1,2‡* 1Department of Radiology, Boston Children’s Hospital, 300 Longwood Avenue, Boston, MA, 02215, USA 2Harvard Medical School, 25 Shattuck Street, Boston, MA, 02115, USA 3GE HealthCare, Research and Scientific Affairs, Waukesha, WI, 53188, USA ‡ Shared last author *Corresponding Author: Da Zhang <Da.
-
Hsieh SS, Leng S, Yu L, Huber NR, McCollough CH.
Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.
Improvements for AI systems
-
Adaptive noise reduction for iterative reconstruction can be improved by using a power law of exponent under 1 for DLIR, which is noted as
smaller exponents for DLIR
in the results section. This allows AI systems to better model how noise suppression scales with prescribed NI across different organs compared to ASiR-V. -
Dose reduction prediction can be made more organ specific by incorporating the derived relationships between the NM-to-NI ratio and prescribed NI, specifically noting that
the equivalent prescribed NI value for the DLIR reconstruction that would result in the same computed NM as the ASiR-V reconstruction is given in Figure 6.
-
Protocol optimization AI can be enhanced to suggest specific NI values for dose reduction based on organ type, as seen in Table 2, which shows
the optimal NI for the gallbladder is 12.7 and 15.0 for the spinal cord.
-
Noise assessment AI can differentiate between noise sources by analyzing the relationship between computed NM and prescribed NI, noting that "a consistent negative association was seen across each anatomical region as shown in Figure 9 with detailed numerical results in Supplementary Table S2 where the exponent is subtracted by 1 since the NM fit is divided by NI."
-
AI systems can be trained to recognize when DLIR benefits are maximized for specific organs, as demonstrated by
the benefit of using the DLIR reconstruction increases as the prescribed NI for the ASiR-V reconstruction increases
withThe benefit is seen the most in the liver while the detrimental effect is seen the most in the spinal cord.
Abstract
Computed tomography (CT) is used ubiquitously for diagnostic imaging, with novel iterative reconstructions such as hybrid iterative reconstruction with partial model-based component (ASiR-V) and deep learning image reconstruction (DLIR) increasingly utilized. In conjunction, CT vendors have developed a noise index (NI) to reduce radiation dose during protocol development. However, it is unclear how these reconstruction algorithms reduce noise as NI changes, particularly across multiple organs. The goal of this study was to utilize an anthropomorphic phantom to evaluate image noise against the prescribed NI across different organs, determine if DLIR reduces image noise compared to ASiR-V, and use this relationship to determine the maximum volume CT dose index (CTDIvol) reduction at equivalent noise levels. Repeat images were acquired on a GE HealthCare Revolution Apex CT scanner across NI values in an anthropomorphic adult-sized phantom from Kyoto Kagaku with ASiR-V and DLIR. Organs-of-interest were contoured using TotalSegmentator in 3D Slicer. The correlation between the CTDIvol and NI was determined. The standard deviation of the difference in the repeat images was used as a noise metric (NM), correlated with NI, used to determine the NI at equivalent NM between ASiR-V and DLIR, then applied to determine the dose reduction and difference in NM. The relationship between NI and the resulting CTDIvol was found to be proportional to a power law of exponent-2.23. The maximum potential absolute dose reduction from using DLIR was found in the liver at 1.3 mGy with an NI of 7.9. Across all organs, the relative dose reduction increased, the NM-to-NI ratio decreased, and the noise reduction increased across NI. The relationship between image noise and the prescribed NI differed between ASiR-V and DLIR and varied by organ, with model-based dose reductions of up to 1.3 mGy in the liver.
Related papers
- BART Online Open-Source Sequence Toolbox for Computational MRI
- Constrained estimation of rotational invariants of the cumulant expansion (RICE) for rapid tensor-valued diffusion MRI
- Improved Myelin Water Fraction Estimation and Uncertainty Mapping in Multicomponent T2 Relaxometry via NNLS Bootstrap Aggregation
- Radiation damage to normal mammalian tissue in vivo with laser-driven protons at ultra-high instantaneous dose rate
- Fast ungated five-dimensional cardiac MRI on a 1.5 T MR-linac for MRI-guided radiotherapy
- BEAM3R: Beam's-eye-view architecture with Mamba-3 for implicit dose reconstruction