Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors
eess.IV
Submitted: 2026-11-02
Updated: 2026-11-02
Comments: This has been presented at a conference, proceedings are pending
License: http://creativecommons.org/licenses/by-nc-nd/4.0/
The gist: Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk.
Terminology
Abstract
Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk. Conventional measurements at high energy using line-pair gauges or knife-edge phantoms require expensive, carefully fabricated objects and precise alignment, and despite these, scatter from the phantom, detector, and imaging room can further confound the measurement. We present a method for estimating the detector point spread function (PSF) from the mean noise power spectrum (NPS) calculated from repeated radiographs in air. The approach is mathematically equivalent to an autocovariance-based method but is more practical for long-tailed PSFs. We validate the method using simulated air data and Monte Carlo N-particle (MCNP6) calculations of energy deposition in a flat-panel detector. We show that deconvolving the experimental PSF increases the accuracy of transmission radiographs of step wedges. The method requires no phantom or alignments and is straightforward to reproduce.
Related papers
- HDNet: High-resolution Dual-domain Learning for Spectral Compressive Imaging
- Mask-guided Spectral-wise Transformer for Efficient Hyperspectral Image Reconstruction
- FDIR: Fidelity-Anchored Decompressed Image Restoration for Human and Machine Perception
- A Disease-Centric Vision-Language Foundation Model for Precision Oncology in Kidney Cancer
- CADS: A Comprehensive Anatomical Dataset and Segmentation for Whole-Body Anatomy in Computed Tomography
- Task-Aware Bitrate Thresholds for Codec-Robust Ultrasound Segmentation