Absolute calibration of scintillators for laser-driven ion spectroscopy using in situ CR-39 sampling detectors

arXiv:2610.38063 · physics.plasm-ph, physics.ins-det · Submitted 2026-10-27 · Read on arXiv

physics.plasm-ph, physics.ins-det

Submitted: 2026-10-27

Updated: 2026-10-27

Comments: 5 pages, 4 figures

License: http://creativecommons.org/licenses/by-nc-nd/4.0/

The gist: Quantitative studies of laser-driven ion acceleration and laser-induced nuclear fusion require accurate measurements of ion yields and energy spectra.

Terminology

Abstract

Quantitative studies of laser-driven ion acceleration and laser-induced nuclear fusion require accurate measurements of ion yields and energy spectra. Although scintillator screens integrated with Thomson parabola spectrometers provide real-time diagnostics, their nonlinear scintillation response requires absolute calibration. We present an in situ, single-shot cross-calibration method using a 2-mm-period CR-39 sampling mask to simultaneously record particle counts and scintillation signals from laser-accelerated proton beams. The method is applied to calibrate a 400- μ m-thick CsI:Tl crystal and a 6- μ m-thick P43 phosphor screen over a broad proton energy range. The measured scintillation responses are compared with a theoretical model combining Birks' law and simulated proton stopping powers. The calibration quantitatively reveals the distinct response characteristics of the two scintillators, including thickness-dependent punch-through effects and constraints on ionization quenching, and demonstrates a practical approach for the absolute calibration of scintillator screens used in laser-driven ion spectroscopy.