Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging
Listen
Radio episode about this paper
Transcript
Introduction to the show: ident: Electrical Engineering. The day in electrical engineering: signals, systems and the hardware that carries them.
Rosa: We're looking at the paper: "Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging".
Dev: Here's the gist of the paper. Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging proposes a complementary-aperture pulse sequence (CAPS) that recovers the linearly scaled lowpressure signal by coherently summing RF acquisitions from interleaved transmit subapertures whereas a full-aperture transmission provides the high-pressure signal maintaining the same excitation voltage configured in the ultrasound system.
Rosa: Using this method, CAPS demonstrates that it preserves depletion-based B/A estimation without using the pressure scaling factor as a reconstruction input, supporting experimental and in vivo implementation without pressure-ratio calibration.
Dev: The Problem with Current Methods The depletion method (DM) enables the estimation of the acoustic nonlinearity parameter, B/A, in pulse-echo ultrasound using fundamental-band signals from low- and highpressure acquisitions but requires accurate knowledge of the acoustic-pressure scaling factor.
Rosa: This requirement limits experimental translation because of the slightly nonlinear scaling between excitation voltage set on the ultrasound system and pressure at the probe.
Title and authors: Rosa: So we’re talking about this paper now called Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging. It looks like they are trying to get a better way to figure out that acoustic nonlinearity parameter, B/A, without needing that tricky pressure scaling factor.
Dev: Yeah, exactly. The core idea is using a complementary-aperture pulse sequence called CAPS to recover the signal from low-pressure acquisitions in a linearly scaled way while using the full aperture transmission for the high-pressure signal.
Taro: I wonder how this actually helps us when we move this out of a controlled lab setting and into something more real, like in an animal or even outside?
Rosa: That’s what I'm thinking. Because the depletion method usually demands you know that pressure scaling factor precisely, which is hard to get right because the voltage you set on your ultrasound machine doesn't perfectly translate to the actual pressure at the probe.
Dev: Right, and that scaling factor uncertainty is a big issue for experimental translation, so this method tries to bypass it by using coherent summation of those interleaved subapertures.
Rosa: So the paper claims CAPS preserves the depletion-based B/A estimation without needing that pressure scaling factor as an input for reconstruction, which sounds really useful for making things practical.
Dev: It does, and they show it works across several aperture partitions, achieving mean B/A values of eight point nine when they aimed for a nine with biases under one point three percent.
Taro: That stability across different subaperture partitions is what interests me—it suggests the reconstruction isn't overly sensitive to how you partition the array into those subapertures.
Rosa: It sounds like the method is robust because it handles that scaling issue internally by summing up those RF acquisitions before envelope detection, approximating that linear pressure scaling factor.
Dev: If we look at their theoretical basis, they’re starting with the fundamental-band pressure equation for low-level source pressure P0 and then comparing it to a high-level source pressure PH which has that scaling factor nu built into it.
Taro: So they’re essentially using the difference between those two pressures, that depletion D equals nuP'L minus P'H, to estimate B/A without needing the scaling factor itself.
Title and authors: Rosa: That’s a nice way to look at it because the true depletion is defined by that difference between the low and high pressure acquisitions, even when you don't know how those pressures scale exactly.
Dev: The mechanism they propose involves partitioning an N-element linear array into m complementary interleaved subapertures, where m divides N, and using a full aperture transmission alongside those subapertures to get that nominal scaling factor nu approximately equal to m.
Taro: But what happens when the world misbehaves? How does this sequence handle noise or unexpected scatterers in the medium?
Rosa: The simulation section shows they used a nonlinear acoustic solver and introduced random spatial density fluctuations with a standard deviation of two percent of the mean, which tests how it handles those kinds of things <ref:2609.23155#pg1>.
Dev: They tested partitions m equals two, four, eight, and sixteen for the nominal scaling factor nu equaling m to see how stable the estimation was.
Taro: And what did they find about that stability when you looked at those different partition sizes?
Rosa: The performance analysis showed that the pressure scaling factor uncertainty is actually identified as a critical failure mode in standard depletion-based B/A imaging, which CAPS seems to solve by not using it for reconstruction.
Dev: They concluded that this method can provide B/A maps equivalent to those from a correctly calibrated depletion method and maintain stable estimates across several aperture partitions.
Taro: So if you’re looking at real-world failure modes, the paper suggests that relying on that external pressure ratio calibration is a major weakness for any deployment.
Rosa: Exactly. It shifts the focus from calibrating the scaling factor to using the sequence structure itself to manage it, which makes it much more practical for in vivo work or experimental trials where you don't have perfect lab conditions.
Dev: So, essentially, they developed this Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging as a way to get reliable B/A estimates without the need for explicit pressure ratio calibration.
Taro: It seems like a solid approach because it builds in this inherent compensation through coherent summation of those subapertures, which is a structural solution rather than just an external correction.
Title and authors: Rosa: It’s interesting how they managed to recover the linearly scaled low-pressure signal through that specific coherent summing process, which allows for this robust estimation.
Dev: The results they got, mean B/A values of eight point nine for a target of nine with biases under one point three percent across all partitions, really validates their approach against what we know from the ideally calibrated depletion method.
Taro: If you’re building a system, you’d have to watch that pressure scaling factor uncertainty closely because they flagged it as the main vulnerability in traditional methods.
Rosa: It sounds like this paper offers a way forward for experimental and in vivo implementations by removing that calibration dependency entirely, which is a big step for moving nonlinear imaging into the clinic.
Dev: So, to wrap up on this paper called Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging, it provides B/A maps that are stable across different array partitions and don't require pressure ratio calibration.
Taro: For me, the key is that the system handles the scaling error internally via that coherent summation, which means you get consistent estimates even when things aren't perfectly calibrated externally.
Rosa: It’s a strong result because it shows how to maintain reliable depletion-based estimation without needing that external pressure scaling factor as a reconstruction input.
Dev: And for us in terms of real-time loops, the idea of using an interleaved transmit subaperture structure to recover the low-pressure signal linearly is something we need to keep in mind for latency and loop rate considerations.
Taro: I’m still thinking about how this sequence would perform when there are significant random scatterers in a complex medium, because that's where the real test of any method lies.
Rosa: Well, it shows promising performance against simulated environments with density fluctuations, suggesting it has the stability we need for future trials outside the lab.
Dev: Overall, this work on Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging offers a path to more practical B/A estimation by eliminating the dependency on pressure ratio calibration.
Taro: It’s a solid piece of research because it addresses that fundamental limitation head-on by using array geometry in a clever way to compensate for nonlinear scaling issues.
The paper's summary: Rosa: So, to recap where we are, this paper proposes a way to get that acoustic nonlinearity parameter B/A without needing that pressure scaling factor calibration you have to do for depletion methods.
Dev: Right, it uses a complementary-aperture pulse sequence, CAPS. It essentially gets the linearly scaled low-pressure signal back by coherently summing up the RF signals from those interleaved transmit subapertures.
Taro: So if I’m trying to make this practical, like for an in vivo study, does this mean I can just run the experiment and not spend hours calibrating that pressure ratio?
Rosa: Exactly. That’s the main experimental win they are pointing toward. They show that you can use this method to estimate B/A without needing that pressure scaling factor as a reconstruction input at all.
Dev: The theoretical basis here is looking at how sound waves propagate in fluid-like media, specifically how the true depletion D relates to the high-level source pressure PH and the low-level source pressure P0.
Taro: That sounds like they’re using a math model to define what a real depletion looks like, even when we don't know exactly how our ultrasound machine is scaling that pressure up or down.
Rosa: That’s right. The mechanism relies on partitioning the array into m complementary interleaved subapertures and then using a full aperture transmission alongside those subapertures to get a nominal pressure scaling factor, which they say is roughly equal to m.
Dev: And the key is that you have to coherently sum those subaperture RF signals before you detect the envelope so that it approximates that linearly scaled low-pressure signal.
Taro: So what’s the practical result? The paper mentions they got mean B/A values of eight point nine when they aimed for a nine, with biases under one point three percent across all those different partitions.
Rosa: That stability across different subaperture partitions is significant because it suggests the reconstruction stays accurate even if you use different ways to slice up that physical array.
Dev: They explicitly identify the pressure scaling factor uncertainty as a critical failure mode in standard depletion-based B/A imaging, which CAPS seems to avoid by not using that input for reconstruction.
Taro: It sounds like the system is robust because it handles that scaling error internally through the summation process, meaning you get consistent estimates even when things aren't perfectly calibrated externally.
Rosa: So what this means for us in terms of real-world failure modes, it’s that you don't need to trust an external calibration factor to get a reliable B/A map.
Dev: It shifts the focus from calibrating the scaling factor to using the sequence structure itself to manage it, which makes it much more practical for in vivo work or experimental trials where you don't have perfect lab conditions.
Taro: It’s a structural solution rather than just an external correction, which is what I need when testing how gracefully something recovers from noise or scatterers in a complex medium.
Rosa: Well, they did test it against simulated environments with density fluctuations and showed promising performance there, suggesting it has the stability needed for future trials outside the lab.
Dev: So this work on Complementary-Aperture Pulse Sequencing provides a path to more practical B/A estimation by eliminating that dependency on pressure ratio calibration.
Taro: It’s a solid piece of research because it addresses that fundamental limitation head-on by using array geometry in a clever way to compensate for nonlinear scaling issues.
The paper's improvements: Rosa: So, we've talked about how CAPS works to get B/A without needing that pressure scaling factor for reconstruction, but now let's look at what they suggest improving about it.
Dev: They are suggesting an AI system that can estimate the acoustic nonlinearity parameter B/A using those fundamental-band signals from both low and high pressure acquisitions without requiring you to know the acoustic-pressure scaling factor accurately.
Taro: That means we don't have to rely on that manual calibration anymore, right? We can just let the AI figure it out.
Rosa: Exactly. The core improvement is building an AI system capable of doing that estimation on its own, which opens the door for experimental and in vivo use without having to perform a pressure ratio calibration beforehand.
Dev: This means the system can automatically compensate for those pressure scaling errors by utilizing the complementary-aperture pulse sequence structure itself. It’s like the sequence structure does some of the heavy lifting for you.
Taro: So if I'm testing real-world failure modes, does this mean my system can handle a lot more noise or unexpected scatterers gracefully?
Rosa: Yes, that's what they're aiming for. The second improvement is implementing a system that automatically compensates for the pressure scaling errors during B/A estimation just by using the complementary-aperture pulse sequence structure.
Dev: This means the system can estimate B/A using only those summed subaperture signals, because CAPS preserves the depletion-based B/A estimation without using that pressure scaling factor as a reconstruction input. That's a big thing for real-time loops.
Taro: And what about stability when I use different partitions of the array? Does this improvement help with that?
Rosa: The third improvement is enhancing tissue characterization AI by allowing it to maintain accurate B/A estimations across different aperture partitions of an array. That stability comes from the fact that coherent summation of all complementary acquisitions reconstructs the full-aperture linear response and suppresses their combined contribution.
Dev: So, if you use a sparse subaperture, the reconstruction still works because the full aperture signal is implicitly handled by that summing process.
Taro: That sounds like it could really make our autonomy research more reliable when dealing with complex geometries where we can't always get a perfect array configuration.
Rosa: It really suggests that even if you only have partial data from the array, the structure of the sequence itself helps maintain consistency in your B/A maps.
Dev: Overall, this pushes us toward a system that’s less dependent on external calibration inputs and more self-contained in its estimation process.
Taro: It's interesting how they solved the scaling issue not just by adding another measurement, but by using the inherent structure of the pulse sequence itself to handle the physics.
Conclusion: Rosa: So we've covered how CAPS recovers that low-pressure signal linearly by coherently summing subapertures, and how they show this method can handle scaling errors without needing a pressure ratio calibration.
Dev: Right, it basically shows that you can get a reliable B/A map using only the fundamental-band signals if you use this specific pulse sequence structure.
Taro: It sounds like we could finally deploy these kinds of imaging techniques in situations where we don't have perfect lab conditions, which is huge for field robotics.
Rosa: That’s the implication—it moves B/A estimation away from requiring precise external calibration to being more self-contained within the measurement sequence itself.
Dev: The paper, "Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging," demonstrates that this method yields maps equivalent to those from a correctly calibrated depletion method while maintaining stability across different array partitions.
Taro: Stability is key for autonomy; I need to know that the system doesn't start spitting out wildly different B/A values just because the sensor arrangement changes slightly.
Rosa: It does, but CAPS manages that by ensuring coherent summation reconstructs the full response and suppresses those combined contributions from sparse subapertures.
Dev: So this gives us a much stronger safety case for real-time loops because we’re relying on internal signal processing instead of an external scaling factor being perfectly known at all times.
Taro: I just want to know how resilient this is when the world misbehaves, like when there are significant random scatterers in a complex medium or something.
Rosa: Well, they did test it against simulated environments with density fluctuations and showed promising performance there, suggesting it has the stability needed for future trials outside the lab.
Dev: It’s a solid piece of research because it addresses that fundamental limitation head-on by using array geometry in a clever way to compensate for nonlinear scaling issues.
Taro: I think the main thing here is that we can start thinking about how to build these imaging systems on top of this structural compensation, rather than just adding more complex calibration steps.
Rosa: Exactly. We're moving toward systems where the sequence design itself handles a lot of the physical unknowns.
Dev: It’s a big step for making nonlinear imaging practical outside of highly controlled laboratory settings where you can guarantee every parameter is exact and known.
Taro: And once we have these stable maps, it opens up new avenues for real-time decision-making in complex environments like autonomous navigation or remote sensing.
Esteban Aviles, Michael Oelze, Roberto Lavarello, Andres Coila
Laboratorio de Imagenes Medicas, Departamento de Ingenieria, Pontificia Universidad Catolica del Peru · Beckman Institute for Advanced Science and Technology, Department of Electrical and Computer Engineering, University of Illinois Urbana-Champaign
eess.SP, physics.med-ph
Submitted: 2026-09-19
Updated: 2026-11-14
Comments: 4 pages, 3 figures, accepted to 22nd International Symposium on Medical Information Processing and Analysis (SIPAIM)
License: http://creativecommons.org/licenses/by/4.0/
Importance score: 84/100
The gist: Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging proposes a complementary-aperture pulse sequence (CAPS) that recovers the linearly scaled lowpressure…
Key concepts
- Depletion Method (DM)
- This is a technique used to estimate B/A by comparing fundamental-band signals from low- and high-pressure acquisitions. However, it requires knowing the exact acoustic-pressure scaling factor between these two measurements, which is often unknown in practice.
- Complementary-Aperture Pulse Sequence (CAPS)
- CAPS partitions an array into interleaved subapertures. It uses a full aperture transmission alongside these subapertures. By coherently summing the RF signals from the subapertures, it effectively recovers the linearly scaled low-pressure signal without needing external pressure ratio calibration.
- Acoustic Nonlinearity Parameter (B/A)
- B/A is a parameter derived from pulse-echo ultrasound that quantifies how much tissue nonlinearity influences the acoustic wave propagation. Estimating this parameter accurately is crucial for characterizing biological tissues, and CAPS provides a robust way to do so.
- Coherent Summation
- This involves combining the RF signals acquired from multiple complementary subapertures in phase before performing envelope detection. This specific summation process allows the system to approximate the fundamental-band envelope corresponding to the linearly scaled low-pressure signal.
Terminology
Summary
Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging proposes a complementary-aperture pulse sequence (CAPS) that recovers the linearly scaled lowpressure signal by coherently summing RF acquisitions from interleaved transmit subapertures whereas a full-aperture transmission provides the high-pressure signal maintaining the same excitation voltage configured in the ultrasound system. p. 0 This requirement limits experimental translation because of the slightly nonlinear scaling
between excitation voltage set on the ultrasound system and
pressure at the probe. We propose a complementary-aperture
pulse sequence (CAPS) that recovers the linearly scaled lowpressure signal by coherently summing RF acquisitions from
interleaved transmit subapertures whereas a full-aperture transmission provides the high-pressure signal maintaining the same
excitation voltage configured in the ultrasound system. 3% across all partitions, matching an ideally
calibrated DM. These results demonstrate that CAPS preserves
depletion-based B/A estimation without using the pressure scaling
factor as a reconstruction input, supporting experimental and in
vivo implementation without pressure-ratio calibration.Where the paper says it
Where the paper says it
The Problem with Current Methods
The depletion method (DM) enables the estimation of the acoustic nonlinearity parameter, B/A, in pulse-echo ultrasound using fundamental-band signals from low- and highpressure acquisitions but requires accurate knowledge of the acoustic-pressure scaling factor. This requirement limits experimental translation because of the slightly nonlinear scaling between excitation voltage set on the ultrasound system and pressure at the probe. p. 0 pe
Abstract—The depletion method (DM) enables the estimation
of the acoustic nonlinearity parameter, B/A, in pulse-echo ultrasound using fundamental-band signals from low- and highpressure acquisitions, but requires accurate knowledge of the
acoustic-pressure scaling factor. This requirement limits experimental translation because of the slightly nonlinear scaling
between excitation voltage set on the ultrasound system and
pressure at the probe. It is based on the fundamentalband depletion predicted by weakly nonlinear propagation
theory for plane waves in attenuating fluids [4]. Despite operating in a pulse-echo, fundamental-band framework, the depletion method (DM) requires accurate knowledge
of the source-pressure scaling factor between the low- and
high-pressure acquisitions.Where the paper says it
Where the paper says it
Theoretical Basis of Depletion Estimation
The DM is derived from the theory of weakly nonlinear plane-wave propagation in attenuating fluid-like media. The fundamental-band pressure at depth z is given by equation (1). p. 1 2
a low-level source pressure P0, the fundamental-band pressure
at depth z is
PL = P0e
−αz
1 −
1
32
Γ
2
1 − e
−2αz2
, (1)
where α is the attenuation coefficient and Γ is the Gol’dberg
number [10], given by
Γ = 2πf0βP0
ρ0c
3
0α
. (2)
Here, f0 is the fundamental frequency, β = 1 + 1
2
(B/A)
is the Beyer nonlinearity coefficient, ρ0 is the equilibrium
density, and c0 is the speed of sound. 2
a low-level source pressure P0, the fundamental-band pressure
at depth z is
PL = P0e
−αz
1 −
1
32
Γ
2
1 − e
−2αz2
, (1)
where α is the attenuation coefficient and Γ is the Gol’dberg
number [10], given by
Γ = 2πf0βP0
ρ0c
3
0α
. (2)
Here, f0 is the fundamental frequency, β = 1 + 1
2
(B/A)
is the Beyer nonlinearity coefficient, ρ0 is the equilibrium
density, and c0 is the speed of sound. Since depletion
is the difference between two nearly equal signals, errors in ν
can be strongly amplified. Let D = νP′
L −P
′
H denote the true
depletion and νb = ν + ∆ν the measured scaling factor.Where the paper says it
Where the paper says it
Where the paper says it
Mechanism of CAPS
The proposed CAPS partitions an N-element linear array into m complementary interleaved subapertures where m divides N. Each subaperture contains N/m elements selected every m-th element while an additional transmission uses the full aperture. p. 1 Complementary-Aperture Pulse Sequence
The proposed CAPS partitions an N-element linear array
into m complementary interleaved subapertures, where m
divides N. Each subaperture contains N/m elements, selected
every m-th element, while an additional transmission uses
the full aperture. Since each subaperture contains N/m elements, its axial pressure is approximately 1/m of the full-aperture pressure. The sequence
therefore provides a nominal pressure-scaling factor of ν ≈ m. The sequence
therefore provides a nominal pressure-scaling factor of ν ≈ m. The subaperture RF signals must be coherently summed
before envelope detection so that their fundamental-band envelope approximates νP′
L
.Where the paper says it
Where the paper says it
Where the paper says it
Performance and Conclusion
CAPS yielded mean B/A values of 8.9 for a prescribed value of 9 with biases of less than 1. p. 0 CAPS yielded
mean B/A values of 8. 9 for a prescribed value of 9, with biases
of less than 1. 9 for a prescribed value of 9, with biases
of less than 1. 3% across all partitions, matching an ideally
calibrated DM. DISCUSSION AND CONCLUSION
This work introduced CAPS for fundamental-band B/A
imaging without providing the pressure-scaling factor to the
estimator. The sensitivity analysis identifies the pressure scaling factor uncertainty as a critical failure mode of depletionbased B/A imaging. CAPS could
also be incorporated into depletion-based spatial compounding
[12], [13] and joint attenuation and B/A estimation [14]. In conclusion, CAPS can provide B/A maps equivalent to
those of the correctly calibrated depletion method and stable
estimates across several aperture partitions.Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
ACKNOWLEDGMENT
This research was funded by the Consejo Nacional de Ciencia, Tecnolog´ıa e Innovacion Tecnol ´ ogica (CONCYTEC) and ´the Programa Nacional de Investigacion Cient ´ıfica y Estudios Avanzados (PROCIENCIA) under the contest E073-2025-01 “Tesis de Pregrado y Postgrado en Ciencia, Tecnolog´ıa e Innovacion Tecnol ´ ogica” (award number PE501099673-2025).
REFERENCES
[1] A. Panfilova, R. p. 3 ´
REFERENCES
[1] A. Panfilova, R. Panfilova, R. van Sloun, H. van Sloun, H. Wijkstra, O. Wijkstra, O. Sapozhnikov, and
M. Sapozhnikov, and
M. Mischi, “A review on B/A measurement methods with a clinical
perspective,” J.Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
[2] D. Zhang, X.-F. Gong, and X. p. 3 2200–2237, Apr. Gong, and X. Gong, and X. Chen, “Experimental imaging of the
acoustic nonlinearity parameter B/A for biological tissues via a parametric array,” Ultrasound Med.Where the paper says it
Where the paper says it
[3] R. L. Errabolu, C. p. 3 1359–1365,
Oct. Errabolu, C. Errabolu, C. Bahn, and J. Bahn, and J. Greenleaf,
“Measurement of ultrasonic nonlinear parameter in excised fat tissues,”
Ultrasound Med.Where the paper says it
Where the paper says it
Where the paper says it
[4] M. F. Hamilton and D. p. 3 137–146, Mar. Hamilton and D. Hamilton and D. Blackstock, Nonlinear Acoustics.Where the paper says it
Where the paper says it
[5] X. Gong, D. Zhang, J. Liu, H. Wang, Y.-Yan, and X. p. 3 San Diego:
Academic Press, 1998. Yan, and X. Yan, and X. Xu, “Study of
acoustic nonlinearity parameter imaging methods in reflection mode for
biological tissues,” J.Where the paper says it
Where the paper says it
[6] R. J. G. van Sloun, L. p. 3 Panfilova, R. van Sloun, H. van Sloun, L. Shan, and M.Where the paper says it
Where the paper says it
[7] M. Nikoonahad and D. p. 3 1331–1341, Jul. Nikoonahad and D.Where the paper says it
[8] M. Fatemi and J. p. 3 127–134, May 1990. Fatemi and J. Fatemi and J. Greenleaf, “Real-time assessment of the parameter
of nonlinearity in tissue using ‘nonlinear shadowing’,” Ultrasound Med.Where the paper says it
Where the paper says it
[9] A. Coila, A. Romero, M. L. Oelze, and R. p. 3 1215–1228, Nov. Oelze, and R. Oelze, and R. Lavarello, “Nonlinearity
parameter estimation method from fundamental band signal depletion
in pulse-echo using a dual-energy model,” J.Where the paper says it
Where the paper says it
[10] L. E. Kinsler, A. p. 3 1969–1980, Mar. Kinsler, A. Kinsler, A. Coppens, and J. Coppens, and J. Sanders, Fundamentals of Acoustics.Where the paper says it
Where the paper says it
Where the paper says it
[11] B. E. Treeby, J.-Jaros, A. P Rendell, and B T Cox, “Modeling nonlinear ultrasound propagation in heterogeneous media with power law absorption using a k-space pseudospectral method,” J. p. 3 Rendell, and B. Cox, “Modeling
nonlinear ultrasound propagation in heterogeneous media with power
law absorption using a k-space pseudospectral method,” J.Where the paper says it
[12] E. Aviles, R. Lavarello, and A. p. 3 Oelze, and R. Lavarello, “Nonlinearity
parameter estimation method from fundamental band signal depletion
in pulse-echo using a dual-energy model,” J. Oelze, and R. Lavarello, “Nonlinearity
parameter estimation method from fundamental band signal depletion
in pulse-echo using a dual-energy model,” J. Oelze, and R. Lavarello, “Nonlinearity
parameter estimation method from fundamental band signal depletion
in pulse-echo using a dual-energy model,” J.Where the paper says it
Where the paper says it
Where the paper says it
[13] E. Aviles, R. Lavarello, and A. Coila, “Nonlinearity parameter imaging ´ using a multi-view joint inverse problem formulation,” in Proc IEEE Int Ultrason Symp (IUS), Sep 2025, pp 1–4
[14] S. Merino, A. Romero, R. Lavarello, and A. Coila, “Regularized joint estimator of the nonlinearity parameter and attenuation coefficient using nonlinear least-squares algorithm,” Ultrason Imag. p. 3 Lavarello, and A. Coila, “Regularized joint
estimator of the nonlinearity parameter and attenuation coefficient using
nonlinear least-squares algorithm,” Ultrason.Where the paper says it
Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.
Improvements for AI systems
-
Develop an AI system capable of estimating acoustic nonlinearity parameter B/A in pulse-echo ultrasound using fundamental-band signals from both low- and high-pressure acquisitions without requiring accurate knowledge of the acoustic-pressure scaling factor. This allows for experimental and in vivo implementation without pressure ratio calibration, as CAPS
recovers the linearly scaled lowpressure signal by coherently summing RF acquisitions from interleaved transmit subapertures.
-
Implement a system that automatically compensates for pressure scaling errors during B/A estimation by utilizing the complementary-aperture pulse sequence structure. This means the system can estimate B/A using only the summed subaperture signals, as
CAPS preserves depletion-based B/A estimation without using the pressure scaling factor as a reconstruction input.
-
Enhance tissue characterization AI by allowing it to maintain accurate B/A estimations across different aperture partitions of an array. The system can achieve this stability because
coherent summation of all complementary acquisitions reconstructs the full-aperture linear response and suppresses their combined contribution,
leading to stable estimates even for sparse subapertures.
Abstract
The depletion method (DM) enables the estimation of the acoustic nonlinearity parameter, B/A, in pulse-echo ultrasound using fundamental-band signals from low- and high-pressure acquisitions, but requires accurate knowledge of the acoustic-pressure scaling factor. This requirement limits experimental translation because of the slightly nonlinear scaling between excitation voltage set on the ultrasound system and pressure at the probe. We propose a complementary-aperture pulse sequence (CAPS) that recovers the linearly scaled low-pressure signal by coherently summing radio-frequency acquisitions from interleaved transmit subapertures, whereas a full-aperture transmission provides the high-pressure signal maintaining the same excitation voltage configured in the ultrasound system. In silico phantoms of a homogeneous sample and a reference evaluated four aperture partitions of CAPS and compared them with the DM under relative scaling factor errors from 0.01% to 10%. Using the DM, errors of 0.1% and 1% produced absolute B/A biases of up to 27.3% and 58.3%, respectively. CAPS yielded mean B/A values of 8.9 for a prescribed value of 9, with biases of less than 1.3% across all partitions, matching an ideally calibrated DM. These results demonstrate that CAPS preserves depletion-based B/A estimation without using the pressure scaling factor as a reconstruction input, supporting experimental and in vivo implementation without pressure-ratio calibration.
Related papers
- Open-H-RF: A Large-Scale Ultrasound Raw Channel Dataset Initiative
- Scalable Attention for 5G NR Channel Estimation
- Low-Altitude Wireless Networks: A Comprehensive Survey
- Sensing-Assisted OFDM with Index Modulation: A Waveform for Optical Wireless Integrated Communication and Localisation
- Iteration-Free Candidate Pruning and an Index-Domain VLSI Architecture for Sort-Free MIMO Box Decoding up to 1024-QAM
- A Foundation Model for Large-Scale Wireless Network Planning, Operation and Optimization