APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations
Listen
Radio episode about this paper
Transcript
Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.
Kai: We're looking at the paper: "APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations".
Mira: Here's the gist of the paper. Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research.
Kai: This study aims to develop a scheme called APHABAMAS to rigorously assess the accuracy of existing high-resolution 3D MRI motion-artifact simulations by providing a ground-truth reference free from sampling-induced error.
Mira: Necessity for Ground Truth The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error.
Kai: Existing simulation algorithms introduce inevitable errors because they rely on motion-free MR scans as a proxy for the true spin density, which are discrete and band-limited approximations of the underlying object's spin density.
Paper summary: Kai: So, looking at the title APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution three dee MRI motion-artifact simulations, it really just sounds like a structured way to validate simulation tools rather than just another motion correction technique itself <ref:2607.09945#pg1,APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high>.
Mira: Right. The authors are Tianqi Wu and Hui Zhang from University College London, and they're tackling the problem of having no existing framework to evaluate the accuracy of these various simulation methods. Their core contribution is building this scheme using an analytical phantom that bypasses sampling errors in the ground truth reference generation process.
Lev: It’s important to remember that their finding highlights that the Type two NUFFT based algorithm is superior in terms of consistency, especially when you look at the median SSIM and RMSD values across different motion types. This suggests a specific mathematical approach is more robust for this task than the Type one NUFFT method they tested.
Kai: So what does this mean for us in terms of practical application? It means that if you're developing a pipeline to correct MRI scans, you can use APHABAMAS to test different simulation models and pick the one that gives you the best fidelity when dealing with motion corruption.
Mira: Precisely. The analytical phantom approach is flexible, which they mention, meaning they could extend this scheme to incorporate different k-space sampling patterns or more complex acquisition models in future work without completely rebuilding the entire reference generation system. It sets up a framework for continuous improvement in simulation assessment tools.
Conclusion: Kai: So we're wrapping up on APHABAMAS, this paper by Wu and Zhang at UCL, which is about checking how good these high-resolution three dee MRI motion simulations actually are.
Mira: Exactly. They put forward this new system using an analytical phantom to create a perfect truth reference that doesn't have those sampling errors you get from real scans.
Lev: From my side, I’m thinking about what this means for the actual hardware we build; if the simulation tool is inaccurate, it means our error correction routines are probably misinformed.
Kai: So they found that using a specific type of non-uniform Fourier transform really gave them the best match to that perfect reference across different motion types and severities.
Mira: And they found that this Type two non-uniform transform method was the most consistent, achieving the lowest error measures in most cases, even with slow drift motion.
Lev: But we have to remember that this is just a test on a phantom; it doesn't tell us how much overhead that specific transform takes on real MRI scanners.
Kai: Right, so the implication is they give researchers a way to rigorously compare the three existing simulation algorithms and pick the one that’s actually most physically faithful before they deploy it.
Mira: It lets you select your correction method based on accuracy metrics like median signal-to-noise ratio improvement or root mean square deviation.
Lev: For real hardware, if we're going to use this, we need to make sure the computational cost of that analytical setup doesn't outweigh the benefit of picking a better simulation model.
Kai: That’s the trade-off, right? Rigor versus practicality. Next up, we look at how this framework might let us test things like different k-space sampling patterns in future work.
Hawkes Institute, University College London · Department of Medical Physics and Biomedical Engineering, University College London · Department of Computer Science, University College London
physics.med-ph
Submitted: 2026-07-10
Updated: 2026-11-14
Code: https://github.com/zcahtwu/APHABAMAS
License: http://creativecommons.org/licenses/by-nc-nd/4.0/
Importance score: 76/100
The gist: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research.
Key concepts
- Ground Truth Reference
- This is a perfect, error-free image used to check the accuracy of simulations. In this study, it's created using an analytical digital phantom that allows for exact calculations of the true underlying spin density and its Fourier transform, overcoming sampling errors found in real MR scans.
- Sampling-Induced Error
- This is an unavoidable inaccuracy introduced when simulating motion artifacts because simulations rely on discrete, band-limited approximations of the true underlying object's spin density. This error varies depending on the specific algorithm used to create the simulation, making it hard to compare them fairly.
- APHABAMAS Scheme
- This is a new assessment method designed to evaluate different motion simulation algorithms. It works by comparing their outputs against the perfect ground truth reference generated by an analytical phantom. This allows researchers to objectively determine which algorithm is most faithful to reality for synthesizing motion-corrupted data.
Terminology
Summary
Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. p. 1 Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. This study aims to develop such a scheme. Methods: The essential ingredient of the desired scheme is a ground-truth reference
that does not suffer from sampling-induced error.Where the paper says it
Where the paper says it
Necessity for Ground Truth
The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error. Existing simulation algorithms introduce inevitable errors because they rely on motion-free MR scans as a proxy for the true spin density, which are discrete and band-limited approximations of the underlying object's spin density. This sampling-induced error varies across algorithms, necessitating an assessment scheme to determine their relative strengths and weaknesses. p. 3 To the best of our knowledge, however, no assessment scheme currently exists for evaluating the simulation
accuracy of these algorithms. Consequently, their fidelity has not been systematically
assessed, leaving the relative strengths and weaknesses of the algorithms unclear
and preventing prospective users from making informed choices.Where the paper says it
Ground Truth Generation
The proposed APHABAMAS scheme leverages a digital phantom whose image- and Fourier-domain representations can be expressed analytically under arbitrary rigid-body transformations. This analytical phantom enables computing the ground-truth simulation by satisfying both the requirement for true underlying spin density and the need for closed-form expressions in its Fourier transform. p. 3 The key challenge in establishing such a scheme is producing a ground-truth simulation free from sampling-induced error. To address this, we propose representing
the underlying object using an analytical phantom—a digital phantom whose imageand Fourier-domain representations have closed-form expressions. Because the latter’s computational cost is prohibitive, approximately 0. 4s to simulate a k-space sample, and extreme anatomical realism is
unnecessary for our purpose, this study uses the former choice.Where the paper says it
Where the paper says it
Algorithm Assessment Results
The assessment demonstrated that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1) NUFFT-based algorithm deteriorates substantially with increasing motion severity. p. 1 APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1)
NUFFT-based algorithm, which does not mimic the MR acquisition process, deteriorates substantially with increasing motion severity. APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1)
NUFFT-based algorithm, which does not mimic the MR acquisition process, deteriorates substantially with increasing motion severity.Where the paper says it
Where the paper says it
Conclusion and Utility
APHABAMAS provides a rigorous tool for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations. The Type-2 NUFFT-based algorithm is identified as the most consistently accurate and physically faithful among the evaluated algorithms. p. 1 APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1)
NUFFT-based algorithm, which does not mimic the MR acquisition process, deteriorates substantially with increasing motion severity. Conclusions: APHABAMAS provides a rigorous tool for assessing the accuracy of
high-resolution 3D MRI motion-artifact simulations. It enables accuracy-based assessment of three existing algorithms, thereby facilitating informed selection of the
most suitable one for synthesizing motion-corrupted data.Where the paper says it
Where the paper says it
References
[1] Mugler III John P, Brookeman James R. Three-dimensional magnetizationprepared rapid gradient-echo imaging (3D MP RAGE). p. 2 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. This obviates modeling intra-readout
motion. Following the above definitions and assumptions, the simulated signal can be expressed as:
S(
⃗k(t)) = Z
ρt(⃗r) e
−i2π⃗k(t)·(⃗r−⃗riso,t)
d⃗r , (2)
where ⃗riso,t represents the isocenter location at time t, corresponding to the physical
position where the main magnetic field strength remains invariant regardless of gradient activity. This obviates modeling intra-readout
motion. Following the above definitions and assumptions, the simulated signal can be expressed as:
S(
⃗k(t)) = Z
ρt(⃗r) e
−i2π⃗k(t)·(⃗r−⃗riso,t)
d⃗r , (2)
where ⃗riso,t represents the isocenter location at time t, corresponding to the physical
position where the main magnetic field strength remains invariant regardless of gradient activity. This obviates modeling intra-readout
motion. Following the above definitions and assumptions, the simulated signal can be expressed as:
S(
⃗k(t)) = Z
ρt(⃗r) e
−i2π⃗k(t)·(⃗r−⃗riso,t)
d⃗r , (2)
where ⃗riso,t represents the isocenter location at time t, corresponding to the physical
position where the main magnetic field strength remains invariant regardless of gradient activity. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
[2] Mugler III John P, Bao Sumi, Mulkern Robert V, et al. Optimized single-slab threedimensional spin-echo MR imaging of the brain. p. 2 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[3] Jack Jr Clifford R, Bernstein Matt A, Fox Nick C, et al. The Alzheimer’s disease neuroimaging initiative (ADNI): MRI methods. p. 2 Consequently, these techniques have become
routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK
Biobank [6]. Despite these advantages, however, high-resolution 3D MRI are prone to motion
artifacts [7], as their acquisition times often exceed the period during which children,
elderly individuals, and patients with neurological disorders can remain still. This obviates modeling intra-readout
motion. Following the above definitions and assumptions, the simulated signal can be expressed as:
S(
⃗k(t)) = Z
ρt(⃗r) e
−i2π⃗k(t)·(⃗r−⃗riso,t)
d⃗r , (2)
where ⃗riso,t represents the isocenter location at time t, corresponding to the physical
position where the main magnetic field strength remains invariant regardless of gradient activity. This obviates modeling intra-readout
motion. Following the above definitions and assumptions, the simulated signal can be expressed as:
S(
⃗k(t)) = Z
ρt(⃗r) e
−i2π⃗k(t)·(⃗r−⃗riso,t)
d⃗r , (2)
where ⃗riso,t represents the isocenter location at time t, corresponding to the physical
position where the main magnetic field strength remains invariant regardless of gradient activity. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
[4] Gunter J, Thostenson K, Borowski B, et al. ADNI-3 MRI PROTOCOL. p. 0 APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Consequently, these techniques have become
routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK
Biobank [6]. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[5] Van Essen David C, Smith Stephen M, Barch Deanna M, et al. The WU-Minn human connectome project: an overview. p. 2 They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Consequently, these techniques have become
routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK
Biobank [6]. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[6] Littlejohns Thomas J, Holliday Jo, Gibson Lorna M, et al. The UK Biobank imaging enhancement of 100,000 participants: rationale, data collection, management and future directions. p. 2 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[7] Zaitsev Maxim, Maclaren Julian, Herbst Michael. Motion artifacts in MRI: A complex problem with many partial solutions. p. 2 For motion-artifact detection, labels describing the
presence or severity of motion artifacts generally require labor-intensive manual rating. For motion correction, many approaches [14, 15] require paired motion-free and
motion-corrupted scans, which are rarely obtainable. 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[8] Reuter Martin, Tisdall M Dylan, Qureshi Abid, Buckner Randy L, Kouwe Andre JW, Fischl Bruce. Head motion during MRI acquisition reduces gray matter volume and thickness estimates. p. 2 Such
artifacts degrade image quality and bias downstream analyses, leading to inaccurate
biomarker estimates, e. , gray matter volume and cortical thickness [8, 9]. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[9] Alexander-Bloch Aaron, Clasen Liv, Stockman Michael, et al. Subtle in-scanner motion biases automated measurement of brain anatomy from in vivo MRI. p. 2 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Consequently, these techniques have become
routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK
Biobank [6]. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[10] Lee Seul, Jung Soozy, Jung Kyu-Jin, Kim Dong-Hyun. Deep learning in MR motion correction: a brief review and a new motion simulation tool (view2Dmotion). p. 1 Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. p. 1 Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
[12] Shaw Richard, Sudre Carole H, Varsavsky Thomas, Ourselin Sebastien, Cardoso M Jorge. A k-space model of movement artefacts: application to segmentation augmentation and artefact removal. p. 2 Shaw et al. [12] demonstrated that training with artifact-augmented data improves
robustness of brain segmentation to motion. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[13] Fantini Irene, Yasuda Clarissa, Bento Mariana, Rittner Leticia, Cendes Fernando, Lotufo Roberto. Automatic MR image quality evaluation using a Deep CNN: A reference-free method to rate motion artifacts in neuroimaging. p. 1 Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[14] Johnson Patricia M, Drangova Maria. Conditional generative adversarial network for 3D rigid-body motion correction in MRI. p. 1 To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourierdomain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing
algorithms under synthetic and tracking-data-derived motion trajectories at different
motion severity levels. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[15] Duffy Ben A, Zhao Lu, Sepehrband Farshid, et al. Retrospective motion artifact correction of structural MRI images using deep learning improves the quality of cortical surface reconstructions. p. 2 Therefore,
identifying and either discarding or correcting these low-quality scans is essential. Recent advances in deep-learning (DL), particularly in computer vision, have driven
a growing interest in DL-based approaches for mitigating motion artifacts in 3D MR [10,
11]. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[16] Al-Masni Mohammed A, Lee Seul, Yi Jaeuk, et al. Stacked U-Nets with selfassisted priors towards robust correction of rigid motion artifact in brain MRI. p. 2 Shaw et al. [12] demonstrated that training with artifact-augmented data improves
robustness of brain segmentation to motion. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[17] Nghiem Brian, Wu Zhe, Kashyap Sriranga, Kasper Lars, Uludag K ˘ amil. A network-assisted joint image and motion estimation approach for robust 3D MRI motion correction across severity levels. p. 2 More recently, Nghiem et al. [17] combined DL with physics-informed joint
estimation for 3D MRI motion correction, and Singh et al. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[18] Singh Munendra, Mahmud Sultan Z, Yedavalli Vivek, et al. Learning-based motion artifact correction in the Z-spectral domain for chemical exchange saturation transfer MRI. p. 2 [17] combined DL with physics-informed joint
estimation for 3D MRI motion correction, and Singh et al. [18] developed a physicsdriven DL framework for motion correction in the Z-spectral domain of CEST MRI. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
[19] Narai ´ Ad´ am, Hermann Petra, Auer Tibor, et al. Movement-related artefacts (MR- ART) dataset of matched motion-corrupted and clean structural MRI brain scans. p. 2 1 Introduction
Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They
generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI, an established
tool in quantitative neuroimaging research. Deep learning-based methods have shown
promise for mitigating motion-induced artifacts, but their development typically requires
simulated motion-corrupted data. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[20] Perez-Garc ´ ´ıa Fernando, Sparks Rachel, Ourselin Sebastien. TorchIO: a Python library for efficient loading, preprocessing, augmentation and patch-based sampling of medical images in deep learning. p. 3 For example, the TorchIO [20] Python library offers a simulator based on the simulation
algorithm by Shaw et al. Duffy et al. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
[21] Reguig Ghiles, Lapert Marc, Lehericy Stephane, Valabregue Romian. Global displacement induced by rigid motion simulation during MRI acquisition. p. 1 To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourierdomain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing
algorithms under synthetic and tracking-data-derived motion trajectories at different
motion severity levels. To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourierdomain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing
algorithms under synthetic and tracking-data-derived motion trajectories at different
motion severity levels. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607. APHABAMAS: An analytical phantom-based
scheme for assessing the accuracy of
high-resolution 3D MRI motion-artifact
simulations
Tianqi Wu1,2 and Hui Zhang1,3
1Hawkes Institute, University College London, London, United Kingdom
2Department of Medical Physics and Biomedical Engineering, University College
London, London, United Kingdom
3Department of Computer Science, University College London, London, United
Kingdom
Correspondence: Tianqi Wu (tianqi. uk)
Keywords: motion artifacts, motion-artifact simulation, analytical phantom
Abstract word count: 239
Word count (body text only): 4991
1
arXiv:2607.Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
Where the paper says it
[21] Reguig Ghiles, Lapert Marc, Lehericy Stephane, Valabregue Romian. Global displacement induced by rigid motion simulation during MRI acquisition. arXiv preprint arXiv:2204.03522. 2022 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>
[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging.
Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.
Improvements for AI systems
-
APHABAMAS provides a rigorous tool for assessing simulation accuracy by comparing algorithm outputs against a ground-truth simulation free from sampling-induced error, enabling an
accuracy-based assessment of three existing algorithms.
-
The improved AI system can systematically rank and select the most suitable motion artifact simulation tool by quantifying the
consistency with the ground truth
across different motion types and severity levels. -
The system can distinguish between simulation errors caused by fundamental theoretical inconsistencies, such as the error in translational motion modeling in Type-1 NUFFT algorithms, from errors arising from approximation techniques like interpolation or internal gridding steps of Type-2 NUFFT.
Abstract
Purpose: Motion compromises the utility of high-resolution 3D MRI. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data. Several open-source tools exist for this task. They share a common formulation but implement this formulation with different algorithms. It is unclear which algorithm implements the formulation more accurately. This study aims to develop a scheme to answer this question. Methods: The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error. To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourier-domain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing algorithms under synthetic and tracking-data-derived motion trajectories at different motion severity levels. Results: All three algorithms produce distinct but visually plausible artifacts, demonstrating the need for a ground-truth reference. APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. For the non-uniform-to-uniform (Type-1) NUFFT-based algorithm, agreement with the ground-truth reference deteriorates substantially with increasing motion severity. Conclusion: APHABAMAS provides a rigorous tool for assessing how accurately algorithms implement the common formulation for simulating motion artifacts in high-resolution 3D MRI. It enables accuracy-based ranking of the three assessed algorithms, thereby facilitating informed selection of the most suitable one for synthesizing motion-corrupted data.
Sources
Related papers
- BART Online Open-Source Sequence Toolbox for Computational MRI
- Constrained estimation of rotational invariants of the cumulant expansion (RICE) for rapid tensor-valued diffusion MRI
- Improved Myelin Water Fraction Estimation and Uncertainty Mapping in Multicomponent T2 Relaxometry via NNLS Bootstrap Aggregation
- Radiation damage to normal mammalian tissue in vivo with laser-driven protons at ultra-high instantaneous dose rate
- BEAM3R: Beam's-eye-view architecture with Mamba-3 for implicit dose reconstruction
- MORSE-PI -- Flexible and artefact-free image reconstruction for structural and functional QSM and other phase-critical imaging applications