APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations

arXiv:2607.09945 · physics.med-ph · Submitted 2026-07-10 · Read on arXiv

Listen

Radio episode about this paper

Transcript

Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.

Kai: We're looking at the paper: "APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations".

Mira: Here's the gist of the paper. Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research.

Kai: This study aims to develop a scheme called APHABAMAS to rigorously assess the accuracy of existing high-resolution 3D MRI motion-artifact simulations by providing a ground-truth reference free from sampling-induced error.

Mira: Necessity for Ground Truth The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error.

Kai: Existing simulation algorithms introduce inevitable errors because they rely on motion-free MR scans as a proxy for the true spin density, which are discrete and band-limited approximations of the underlying object's spin density.

Paper summary: Kai: So, looking at the title APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution three dee MRI motion-artifact simulations, it really just sounds like a structured way to validate simulation tools rather than just another motion correction technique itself <ref:2607.09945#pg1,APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high>.

Mira: Right. The authors are Tianqi Wu and Hui Zhang from University College London, and they're tackling the problem of having no existing framework to evaluate the accuracy of these various simulation methods. Their core contribution is building this scheme using an analytical phantom that bypasses sampling errors in the ground truth reference generation process.

Lev: It’s important to remember that their finding highlights that the Type two NUFFT based algorithm is superior in terms of consistency, especially when you look at the median SSIM and RMSD values across different motion types. This suggests a specific mathematical approach is more robust for this task than the Type one NUFFT method they tested.

Kai: So what does this mean for us in terms of practical application? It means that if you're developing a pipeline to correct MRI scans, you can use APHABAMAS to test different simulation models and pick the one that gives you the best fidelity when dealing with motion corruption.

Mira: Precisely. The analytical phantom approach is flexible, which they mention, meaning they could extend this scheme to incorporate different k-space sampling patterns or more complex acquisition models in future work without completely rebuilding the entire reference generation system. It sets up a framework for continuous improvement in simulation assessment tools.

Conclusion: Kai: So we're wrapping up on APHABAMAS, this paper by Wu and Zhang at UCL, which is about checking how good these high-resolution three dee MRI motion simulations actually are.

Mira: Exactly. They put forward this new system using an analytical phantom to create a perfect truth reference that doesn't have those sampling errors you get from real scans.

Lev: From my side, I’m thinking about what this means for the actual hardware we build; if the simulation tool is inaccurate, it means our error correction routines are probably misinformed.

Kai: So they found that using a specific type of non-uniform Fourier transform really gave them the best match to that perfect reference across different motion types and severities.

Mira: And they found that this Type two non-uniform transform method was the most consistent, achieving the lowest error measures in most cases, even with slow drift motion.

Lev: But we have to remember that this is just a test on a phantom; it doesn't tell us how much overhead that specific transform takes on real MRI scanners.

Kai: Right, so the implication is they give researchers a way to rigorously compare the three existing simulation algorithms and pick the one that’s actually most physically faithful before they deploy it.

Mira: It lets you select your correction method based on accuracy metrics like median signal-to-noise ratio improvement or root mean square deviation.

Lev: For real hardware, if we're going to use this, we need to make sure the computational cost of that analytical setup doesn't outweigh the benefit of picking a better simulation model.

Kai: That’s the trade-off, right? Rigor versus practicality. Next up, we look at how this framework might let us test things like different k-space sampling patterns in future work.

Hawkes Institute, University College London · Department of Medical Physics and Biomedical Engineering, University College London · Department of Computer Science, University College London

physics.med-ph

Submitted: 2026-07-10

Updated: 2026-11-14

Code: https://github.com/zcahtwu/APHABAMAS

License: http://creativecommons.org/licenses/by-nc-nd/4.0/

Importance score: 76/100

The gist: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research.

Key concepts

Ground Truth Reference
This is a perfect, error-free image used to check the accuracy of simulations. In this study, it's created using an analytical digital phantom that allows for exact calculations of the true underlying spin density and its Fourier transform, overcoming sampling errors found in real MR scans.
Sampling-Induced Error
This is an unavoidable inaccuracy introduced when simulating motion artifacts because simulations rely on discrete, band-limited approximations of the true underlying object's spin density. This error varies depending on the specific algorithm used to create the simulation, making it hard to compare them fairly.
APHABAMAS Scheme
This is a new assessment method designed to evaluate different motion simulation algorithms. It works by comparing their outputs against the perfect ground truth reference generated by an analytical phantom. This allows researchers to objectively determine which algorithm is most faithful to reality for synthesizing motion-corrupted data.

Terminology

Summary

Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. p. 1

Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

This study aims to develop a scheme called APHABAMAS to rigorously assess the accuracy of existing high-resolution 3D MRI motion-artifact simulations by providing a ground-truth reference free from sampling-induced error. p. 1
Where the paper says it

This study aims to develop such a scheme. Methods: The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error.

Necessity for Ground Truth

The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error. Existing simulation algorithms introduce inevitable errors because they rely on motion-free MR scans as a proxy for the true spin density, which are discrete and band-limited approximations of the underlying object's spin density. This sampling-induced error varies across algorithms, necessitating an assessment scheme to determine their relative strengths and weaknesses. p. 3

Where the paper says it

To the best of our knowledge, however, no assessment scheme currently exists for evaluating the simulation accuracy of these algorithms. Consequently, their fidelity has not been systematically assessed, leaving the relative strengths and weaknesses of the algorithms unclear and preventing prospective users from making informed choices.

Ground Truth Generation

The proposed APHABAMAS scheme leverages a digital phantom whose image- and Fourier-domain representations can be expressed analytically under arbitrary rigid-body transformations. This analytical phantom enables computing the ground-truth simulation by satisfying both the requirement for true underlying spin density and the need for closed-form expressions in its Fourier transform. p. 3

Where the paper says it

The key challenge in establishing such a scheme is producing a ground-truth simulation free from sampling-induced error. To address this, we propose representing the underlying object using an analytical phantom—a digital phantom whose imageand Fourier-domain representations have closed-form expressions.

The study uses a 3D Shepp–Logan phantom as the choice because it meets these requirements and has a manageable computational cost. p. 7
Where the paper says it

Because the latter’s computational cost is prohibitive, approximately 0. 4s to simulate a k-space sample, and extreme anatomical realism is unnecessary for our purpose, this study uses the former choice.

Algorithm Assessment Results

The assessment demonstrated that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1) NUFFT-based algorithm deteriorates substantially with increasing motion severity. p. 1

Where the paper says it

APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1) NUFFT-based algorithm, which does not mimic the MR acquisition process, deteriorates substantially with increasing motion severity.

Specifically, the Type-2 NUFFT-based algorithm showed the most consistent agreement with the ground truth, achieving the highest median SSIM and lowest median RMSD in all conditions except for mild slow-drift motion. p. 1
Where the paper says it

APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1) NUFFT-based algorithm, which does not mimic the MR acquisition process, deteriorates substantially with increasing motion severity.

Conclusion and Utility

APHABAMAS provides a rigorous tool for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations. The Type-2 NUFFT-based algorithm is identified as the most consistently accurate and physically faithful among the evaluated algorithms. p. 1

Where the paper says it

APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. The non-uniform-to-uniform (Type-1) NUFFT-based algorithm, which does not mimic the MR acquisition process, deteriorates substantially with increasing motion severity.

This scheme enables accuracy-based assessment of three existing algorithms, thereby facilitating informed selection of the most suitable one for synthesizing motion-corrupted data. p. 1
Where the paper says it

Conclusions: APHABAMAS provides a rigorous tool for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations. It enables accuracy-based assessment of three existing algorithms, thereby facilitating informed selection of the most suitable one for synthesizing motion-corrupted data.

The analytical phantom's flexibility allows for straightforward extensions to additional k-space sampling patterns and more complex acquisition models in future work.

References

[1] Mugler III John P, Brookeman James R. Three-dimensional magnetizationprepared rapid gradient-echo imaging (3D MP RAGE). p. 2

Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

Magnetic resonance in medicine. p. 5
Where the paper says it

This obviates modeling intra-readout motion. Following the above definitions and assumptions, the simulated signal can be expressed as: S( ⃗k(t)) = Z ρt(⃗r) e −i2π⃗k(t)·(⃗r−⃗riso,t) d⃗r , (2) where ⃗riso,t represents the isocenter location at time t, corresponding to the physical position where the main magnetic field strength remains invariant regardless of gradient activity.

p. 5
Where the paper says it

This obviates modeling intra-readout motion. Following the above definitions and assumptions, the simulated signal can be expressed as: S( ⃗k(t)) = Z ρt(⃗r) e −i2π⃗k(t)·(⃗r−⃗riso,t) d⃗r , (2) where ⃗riso,t represents the isocenter location at time t, corresponding to the physical position where the main magnetic field strength remains invariant regardless of gradient activity.

p. 5
Where the paper says it

This obviates modeling intra-readout motion. Following the above definitions and assumptions, the simulated signal can be expressed as: S( ⃗k(t)) = Z ρt(⃗r) e −i2π⃗k(t)·(⃗r−⃗riso,t) d⃗r , (2) where ⃗riso,t represents the isocenter location at time t, corresponding to the physical position where the main magnetic field strength remains invariant regardless of gradient activity.

1990;15(1):152–157 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg29>

[2] Mugler III John P, Bao Sumi, Mulkern Robert V, et al. Optimized single-slab threedimensional spin-echo MR imaging of the brain. p. 2

Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

Radiology. 2000;216(3):891–899 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[3] Jack Jr Clifford R, Bernstein Matt A, Fox Nick C, et al. The Alzheimer’s disease neuroimaging initiative (ADNI): MRI methods. p. 2

Where the paper says it

Consequently, these techniques have become routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK Biobank [6]. Despite these advantages, however, high-resolution 3D MRI are prone to motion artifacts [7], as their acquisition times often exceed the period during which children, elderly individuals, and patients with neurological disorders can remain still.

Journal of Magnetic Resonance Imaging: An Official Journal of the International Society for Magnetic Resonance in Medicine. p. 5
Where the paper says it

This obviates modeling intra-readout motion. Following the above definitions and assumptions, the simulated signal can be expressed as: S( ⃗k(t)) = Z ρt(⃗r) e −i2π⃗k(t)·(⃗r−⃗riso,t) d⃗r , (2) where ⃗riso,t represents the isocenter location at time t, corresponding to the physical position where the main magnetic field strength remains invariant regardless of gradient activity.

p. 5
Where the paper says it

This obviates modeling intra-readout motion. Following the above definitions and assumptions, the simulated signal can be expressed as: S( ⃗k(t)) = Z ρt(⃗r) e −i2π⃗k(t)·(⃗r−⃗riso,t) d⃗r , (2) where ⃗riso,t represents the isocenter location at time t, corresponding to the physical position where the main magnetic field strength remains invariant regardless of gradient activity.

2008;27(4):685–691 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[4] Gunter J, Thostenson K, Borowski B, et al. ADNI-3 MRI PROTOCOL. p. 0

Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

Alzheimer’s Dementia. p. 2
Where the paper says it

They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Consequently, these techniques have become routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK Biobank [6].

2017;13(7):P104–P105 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[5] Van Essen David C, Smith Stephen M, Barch Deanna M, et al. The WU-Minn human connectome project: an overview. p. 2

Where the paper says it

They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Consequently, these techniques have become routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK Biobank [6].

Neuroimage. 2013;80:62–79 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[6] Littlejohns Thomas J, Holliday Jo, Gibson Lorna M, et al. The UK Biobank imaging enhancement of 100,000 participants: rationale, data collection, management and future directions. p. 2

Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

Nature communications. 2020;11(1):2624 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[7] Zaitsev Maxim, Maclaren Julian, Herbst Michael. Motion artifacts in MRI: A complex problem with many partial solutions. p. 2

Where the paper says it

For motion-artifact detection, labels describing the presence or severity of motion artifacts generally require labor-intensive manual rating. For motion correction, many approaches [14, 15] require paired motion-free and motion-corrupted scans, which are rarely obtainable.

Journal of Magnetic Resonance Imaging. p. 2
Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

2015;42(4):887–901 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[8] Reuter Martin, Tisdall M Dylan, Qureshi Abid, Buckner Randy L, Kouwe Andre JW, Fischl Bruce. Head motion during MRI acquisition reduces gray matter volume and thickness estimates. p. 2

Where the paper says it

Such artifacts degrade image quality and bias downstream analyses, leading to inaccurate biomarker estimates, e. , gray matter volume and cortical thickness [8, 9].

Neuroimage. 2015;107:107–115 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[9] Alexander-Bloch Aaron, Clasen Liv, Stockman Michael, et al. Subtle in-scanner motion biases automated measurement of brain anatomy from in vivo MRI. p. 2

Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

Human brain mapping. p. 2
Where the paper says it

They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology. Consequently, these techniques have become routine choices for large neuroimaging studies, including the Alzheimer’s Disease Neuroimaging Initiative [3, 4], the WU-Minn Human Connectome Project [5], and the UK Biobank [6].

2016;37(7):2385–2397 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[10] Lee Seul, Jung Soozy, Jung Kyu-Jin, Kim Dong-Hyun. Deep learning in MR motion correction: a brief review and a new motion simulation tool (view2Dmotion). p. 1

Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

Investigative Magnetic Resonance Imaging. p. 2
Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

2020;24(4):196–206 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. p. 1

Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

IEEE Transactions on Medical Imaging. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

2023;43(2):846–859 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[12] Shaw Richard, Sudre Carole H, Varsavsky Thomas, Ourselin Sebastien, Cardoso M Jorge. A k-space model of movement artefacts: application to segmentation augmentation and artefact removal. p. 2

Where the paper says it

Shaw et al. [12] demonstrated that training with artifact-augmented data improves robustness of brain segmentation to motion.

IEEE transactions on medical imaging. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

2020;39(9):2881–2892 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[13] Fantini Irene, Yasuda Clarissa, Bento Mariana, Rittner Leticia, Cendes Fernando, Lotufo Roberto. Automatic MR image quality evaluation using a Deep CNN: A reference-free method to rate motion artifacts in neuroimaging. p. 1

Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

Computerized Medical Imaging and Graphics. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

2021;90:101897 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[14] Johnson Patricia M, Drangova Maria. Conditional generative adversarial network for 3D rigid-body motion correction in MRI. p. 1

Where the paper says it

To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourierdomain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing algorithms under synthetic and tracking-data-derived motion trajectories at different motion severity levels.

Magnetic resonance in medicine. 2019;82(3):901–910 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[15] Duffy Ben A, Zhao Lu, Sepehrband Farshid, et al. Retrospective motion artifact correction of structural MRI images using deep learning improves the quality of cortical surface reconstructions. p. 2

Where the paper says it

Therefore, identifying and either discarding or correcting these low-quality scans is essential. Recent advances in deep-learning (DL), particularly in computer vision, have driven a growing interest in DL-based approaches for mitigating motion artifacts in 3D MR [10, 11].

Neuroimage. 2021;230:117756 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[16] Al-Masni Mohammed A, Lee Seul, Yi Jaeuk, et al. Stacked U-Nets with selfassisted priors towards robust correction of rigid motion artifact in brain MRI. p. 2

Where the paper says it

Shaw et al. [12] demonstrated that training with artifact-augmented data improves robustness of brain segmentation to motion.

NeuroImage. 2022;259:119411 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[17] Nghiem Brian, Wu Zhe, Kashyap Sriranga, Kasper Lars, Uludag K ˘ amil. A network-assisted joint image and motion estimation approach for robust 3D MRI motion correction across severity levels. p. 2

Where the paper says it

More recently, Nghiem et al. [17] combined DL with physics-informed joint estimation for 3D MRI motion correction, and Singh et al.

Magnetic Resonance in Medicine. 2026;95(1):363–381 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[18] Singh Munendra, Mahmud Sultan Z, Yedavalli Vivek, et al. Learning-based motion artifact correction in the Z-spectral domain for chemical exchange saturation transfer MRI. p. 2

Where the paper says it

[17] combined DL with physics-informed joint estimation for 3D MRI motion correction, and Singh et al. [18] developed a physicsdriven DL framework for motion correction in the Z-spectral domain of CEST MRI.

Magnetic resonance in medicine. 2025;94(1):331–345 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[19] Narai ´ Ad´ am, Hermann Petra, Auer Tibor, et al. Movement-related artefacts (MR- ART) dataset of matched motion-corrupted and clean structural MRI brain scans. p. 2

Where the paper says it

1 Introduction Modern 3D MRI sequences, particularly MPRAGE [1] and SPACE [2], enable highresolution 3D brain imaging with various contrasts within a feasible time frame. They generate essential data, like T1- and T2-weighted scans, allowing detailed examination of brain anatomy and pathology.

Scientific data. p. 1
Where the paper says it

Abstract Purpose: Motion compromises the utility of high-resolution 3D MRI, an established tool in quantitative neuroimaging research. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data.

2022;9(1):630 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[20] Perez-Garc ´ ´ıa Fernando, Sparks Rachel, Ourselin Sebastien. TorchIO: a Python library for efficient loading, preprocessing, augmentation and patch-based sampling of medical images in deep learning. p. 3

Where the paper says it

For example, the TorchIO [20] Python library offers a simulator based on the simulation algorithm by Shaw et al. Duffy et al.

Computer Methods and Programs in Biomedicine. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

2021;208:106236 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[21] Reguig Ghiles, Lapert Marc, Lehericy Stephane, Valabregue Romian. Global displacement induced by rigid motion simulation during MRI acquisition. p. 1

Where the paper says it

To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourierdomain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing algorithms under synthetic and tracking-data-derived motion trajectories at different motion severity levels.

p. 1
Where the paper says it

To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourierdomain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing algorithms under synthetic and tracking-data-derived motion trajectories at different motion severity levels.

arXiv preprint arXiv:2204. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

03522. 2022 <ref:2607. p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

p. 0
Where the paper says it

APHABAMAS: An analytical phantom-based scheme for assessing the accuracy of high-resolution 3D MRI motion-artifact simulations Tianqi Wu1,2 and Hui Zhang1,3 1Hawkes Institute, University College London, London, United Kingdom 2Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom 3Department of Computer Science, University College London, London, United Kingdom Correspondence: Tianqi Wu (tianqi. uk) Keywords: motion artifacts, motion-artifact simulation, analytical phantom Abstract word count: 239 Word count (body text only): 4991 1 arXiv:2607.

09945#pg4>

[21] Reguig Ghiles, Lapert Marc, Lehericy Stephane, Valabregue Romian. Global displacement induced by rigid motion simulation during MRI acquisition. arXiv preprint arXiv:2204.03522. 2022 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging. 2023;43(2):846–859 <ref:2607.09945#pg4>

[11] Spieker Veronika, Eichhorn Hannah, Hammernik Kerstin, et al. Deep learning for retrospective motion correction in MRI: a comprehensive review. IEEE Transactions on Medical Imaging.

Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.

Improvements for AI systems

  1. APHABAMAS provides a rigorous tool for assessing simulation accuracy by comparing algorithm outputs against a ground-truth simulation free from sampling-induced error, enabling an accuracy-based assessment of three existing algorithms.

  2. The improved AI system can systematically rank and select the most suitable motion artifact simulation tool by quantifying the consistency with the ground truth across different motion types and severity levels.

  3. The system can distinguish between simulation errors caused by fundamental theoretical inconsistencies, such as the error in translational motion modeling in Type-1 NUFFT algorithms, from errors arising from approximation techniques like interpolation or internal gridding steps of Type-2 NUFFT.

Abstract

Purpose: Motion compromises the utility of high-resolution 3D MRI. Deep learning-based methods have shown promise for mitigating motion-induced artifacts, but their development typically requires simulated motion-corrupted data. Several open-source tools exist for this task. They share a common formulation but implement this formulation with different algorithms. It is unclear which algorithm implements the formulation more accurately. This study aims to develop a scheme to answer this question. Methods: The essential ingredient of the desired scheme is a ground-truth reference that does not suffer from sampling-induced error. To meet this requirement, the proposed scheme, APHABAMAS, leverages a digital phantom whose image- and Fourier-domain representations can be expressed analytically under arbitrary rigid-body transformations. APHABAMAS is used to quantify sampling-induced errors of three existing algorithms under synthetic and tracking-data-derived motion trajectories at different motion severity levels. Results: All three algorithms produce distinct but visually plausible artifacts, demonstrating the need for a ground-truth reference. APHABAMAS shows that the algorithm utilizing the uniform-to-non-uniform (Type-2) Non-Uniform Fast Fourier Transform (NUFFT) provides the most consistent agreement with the ground-truth reference across motion types and severity levels. For the non-uniform-to-uniform (Type-1) NUFFT-based algorithm, agreement with the ground-truth reference deteriorates substantially with increasing motion severity. Conclusion: APHABAMAS provides a rigorous tool for assessing how accurately algorithms implement the common formulation for simulating motion artifacts in high-resolution 3D MRI. It enables accuracy-based ranking of the three assessed algorithms, thereby facilitating informed selection of the most suitable one for synthesizing motion-corrupted data.

Sources

Related papers