Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination

arXiv:2606.24384 · physics.med-ph, physics.ins-det · Submitted 2026-06-23 · Read on arXiv

Listen

Radio episode about this paper

Transcript

Introduction to the show: ident: Physics. The day in general physics: instruments, methods and results across the field.

Kai: We're looking at the paper: "Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates".

Mira: Here's the gist of the paper. Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions Modeling Coupled Effects The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions The implementation was validated against analytical and numerical models from the literature The framework extends the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations Physical Mechanisms of Charge Transport The dominant mechanism responsible for the formation of negative ions in air is the attachment of electrons to oxygen molecules Two different attachment mechanisms exist: two-body and three-body attachment Electron transport is performed using AvalancheMicroscopic, which tracks electrons collision-by-collision at the microscopic level, while ion transport is handled by AvalancheMC, which uses Monte Carlo techniques to simulate ion drift lines Implementation of Recombination and Space Charge In ion–ion recombination processes, a distinction is usually made between initial and volume recombination Initial recombination occurs between ions of opposite charge created within the same ionisation track of an incident particle Volume recombination corresponds to recombination between ions originating from different particle tracks In contrast, volume recombination corresponds to recombination between ions originating from different particle tracks Impact of UHDR on Free Electron Fraction The reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment The FEF p is given by p = 1/η d (1 − exp(−η d)) (21) As the DPP increases, the FEF progressively decreases This behaviour can be explained by the strong electric field distortion induced by space charge The results suggest that this behaviour originates from the strong dependence of the electron attachment coefficient on the local electric field The reduction of the CCE under UHDR conditions cannot be explained by a longer coexistence time of positive and negative ions within the chamber This observation suggests that the UHDR dosimetry problem in ionisation chambers may be largely reduced to the accurate description of the FEF evolution The implementation of these physical mechanisms within Garfield++ provides a complete, flexible, and self-consistent Monte Carlo framework capable of accurately reproducing UHDR effects in ionisation chambers Code Availability The two main physical processes implemented in this work, namely ion–ion recombination and space-charge electric-field calculations based on spatial grid discretisation, have been integrated into the open-source Garfield++ software (https://garfieldpp.docs.cern.ch) The complete simulation framework developed in this work, named GarFLIC (Garfield++ for FLASH Ionisation Chamber), is openly available at https://gitlab.com/blond-bay/garflic/tree/1.0.0 Acknowledgements This work and the associated doctoral research of Pierre Gérard Ortega were funded by the Walloon Region through the PIT ProtherWal programme under grant agreement No.

Kai: 7289 Computational resources have been provided by the Consortium des Équipements de Calcul Intensif (CÉCI), funded by the Fonds de la Recherche Scientifique de Belgique (F.R.S.-FNRS) under Grant No.

Mira: 2.5020.11 and by the Walloon Region The authors gratefully acknowledge José Paz Martín for his valuable support and insightful discussions References

1: V.

Kai: Favaudon et al.

Paper summary: Mira: So, looking at the Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination, we see they are putting together a self consistent Monte Carlo model to handle both ion recombination and space charge effects.

Lev: It’s about getting a picture of what happens when you have those intense dose rates and all that ion activity happening inside the chamber simultaneously <ref:2606.24384#pg1>.

Kai: It sets a new standard for modeling these challenging phenomena in ion chambers.

Conclusion: Kai: So, we’re wrapping up this talk on "Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination."

Mira: It boils down to taking a standard simulation and adding the complex stuff—how ions recombine and how the electric field gets messed up by all those ions.

Lev: What this means for us is that when you’re pushing these devices to their limits with high dose rates, you can't just use simple old models anymore because they miss that dynamic distortion.

Kai: I mean if we’re building experimental setups, we need to know exactly how the charge actually gets collected at those high speeds, and this model is trying to capture that self-consistent reality.

Mira: And the authors say their method is more accurate than previous approaches because it accounts for how the electron attachment coefficient changes based on that local electric field.

Lev: That dependence on the electric field is a huge thing; it suggests we need to focus our error analysis on that specific relationship when designing real hardware.

Kai: So if you look at the title, "Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination," it really tells you exactly what physics they are trying to handle.

Mira: They are tackling the coupling between ion recombination and space charge effects, which is usually treated separately in simpler models.

Lev: The core research here is building a self-consistent Monte Carlo framework that handles these things together for the first time in this context.

Kai: It’s about getting a picture of what happens when you have those intense dose rates and all that ion activity happening inside the chamber simultaneously.

Mira: And they show that this approach reveals how much the free electron fraction changes depending on where you are in the chamber, which is critical for accurate measurement.

Lev: It points toward a new way to think about ultra high dose rate dosimetry problems, moving beyond just static measurements to understanding the dynamic evolution of charge collection.

Kai: This work gives us a very detailed tool to simulate these complex physical interactions under extreme conditions.

Mira: And since they made the framework open source, researchers can use it to test their own assumptions about how recombination and space charge interact in these chambers.

Lev: So this paper is providing the necessary physics for anyone trying to design or measure detectors in FLASH environments.

Kai: It sets a new standard for modeling these challenging phenomena in ion chambers.

Mira: This opens up a whole new level of testing how those physical mechanisms actually play out when you crank up the dose rate.

Lev: We need to see if this framework holds up when we move from simulation to building actual hardware components that need to survive those conditions.

Service de Physique des particules elementaires (IIHE) University libre de Bruxelles · Ion Beam Applications SA

physics.med-ph, physics.ins-det

Submitted: 2026-06-23

Updated: 2026-09-17

Comments: 18 pages, 11 figures

DOI: 10.1088/1361-6560/aeb99e

License: http://creativecommons.org/licenses/by/4.0/

Importance score: 71/100

The gist: Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to…

Key concepts

Charge Collection Efficiency (CCE)
This measures how effectively an ionization chamber collects the charge produced by incident radiation. In this context, it is reduced under UHDR because strong electric fields distort the space, affecting electron movement and collection.
Ion-Ion Recombination
This process describes when positive and negative ions created in the same track recombine. The model distinguishes between initial recombination (ions from the same track) and volume recombination (ions from different tracks), both of which affect charge transport.
Space-Charge Electric Field Distortions
When high densities of ions are present, they create a strong electric field distortion. This distortion is a key factor in reducing CCE under UHDR because the electron attachment coefficient becomes dependent on this local electric field strength.

Terminology

Summary

Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions

Modeling Coupled Effects

The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions The implementation was validated against analytical and numerical models from the literature The framework extends the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations

Physical Mechanisms of Charge Transport

The dominant mechanism responsible for the formation of negative ions in air is the attachment of electrons to oxygen molecules Two different attachment mechanisms exist: two-body and three-body attachment Electron transport is performed using AvalancheMicroscopic, which tracks electrons collision-by-collision at the microscopic level, while ion transport is handled by AvalancheMC, which uses Monte Carlo techniques to simulate ion drift lines

Implementation of Recombination and Space Charge

In ion–ion recombination processes, a distinction is usually made between initial and volume recombination Initial recombination occurs between ions of opposite charge created within the same ionisation track of an incident particle Volume recombination corresponds to recombination between ions originating from different particle tracks In contrast, volume recombination corresponds to recombination between ions originating from different particle tracks

Impact of UHDR on Free Electron Fraction

The reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment The FEF p is given by p = 1/η d (1 − exp(−η d)) (21) As the DPP increases, the FEF progressively decreases This behaviour can be explained by the strong electric field distortion induced by space charge

Conclusion and Future Directions

The results suggest that this behaviour originates from the strong dependence of the electron attachment coefficient on the local electric field The reduction of the CCE under UHDR conditions cannot be explained by a longer coexistence time of positive and negative ions within the chamber This observation suggests that the UHDR dosimetry problem in ionisation chambers may be largely reduced to the accurate description of the FEF evolution The implementation of these physical mechanisms within Garfield++ provides a complete, flexible, and self-consistent Monte Carlo framework capable of accurately reproducing UHDR effects in ionisation chambers

Code Availability

The two main physical processes implemented in this work, namely ion–ion recombination and space-charge electric-field calculations based on spatial grid discretisation, have been integrated into the open-source Garfield++ software (https://garfieldpp.docs.cern.ch) The complete simulation framework developed in this work, named GarFLIC (Garfield++ for FLASH Ionisation Chamber), is openly available at https://gitlab. p. 0

Where the paper says it

This finding indicates that the complex problem of recombination under UHDR conditions may be largely governed by the evolution of the FEF, opening promising perspectives for the development of improved analytical models and real-time correction methods for ionisation chamber dosimetry under UHDR irradiation conditions. This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy.

com/blond-bay/garflic/tree/1.0.0

Acknowledgements

This work and the associated doctoral research of Pierre Gérard Ortega were funded by the Walloon Region through the PIT ProtherWal programme under grant agreement No. 7289 Computational resources have been provided by the Consortium des Équipements de Calcul Intensif (CÉCI), funded by the Fonds de la Recherche Scientifique de Belgique (F.R.S.-FNRS) under Grant No. 2.5020.11 and by the Walloon Region The authors gratefully acknowledge José Paz Martín for his valuable support and insightful discussions

References

[1] V. Favaudon et al. p. 0

Where the paper says it

This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy. Keywords: plane-parallel ionisation chambers, UHDR, charge collection efficiency, recombination, space charge, Monte Carlo simulation 1 Introduction Since the discovery of what is now referred to as the FLASH effect by Favaudon et al.

“Ultrahigh Dose-Rate FLASH Irradiation Increases the Differential Response between Normal and Tumor Tissue in Mice”. p. 0
Where the paper says it

Subsequent studies have confirmed normal-tissue sparing in a variety of preclinical models and tissues. In particular, FLASH irradiation was shown to improve survival following abdominal irradiation in mice [2], preserve cognitive function after whole-brain irradiation [3], and reduce brain toxicity using both electron and X-ray beams [4, 5].

In: Science Translational Medicine 6.245 (July 16, 2014). p. 0
Where the paper says it

Keywords: plane-parallel ionisation chambers, UHDR, charge collection efficiency, recombination, space charge, Monte Carlo simulation 1 Introduction Since the discovery of what is now referred to as the FLASH effect by Favaudon et al. [1] in 2014, research on ultra-high dose rate (UHDR) radiation (≥ 40 Gy s−1 ) has grown rapidly.

issn: 1946-6234, 1946-6242. doi: 10.1126 / scitranslmed.3008973. url: https: //www. science. org / doi / 10.1126 / scitranslmed.3008973 >

[2] B. W. Loo et al. “(P003) Delivery of Ultra-Rapid Flash Radiation Therapy and Demonstration of Normal Tissue Sparing After Abdominal Irradiation of Mice”. p. 0

Where the paper says it

Subsequent studies have confirmed normal-tissue sparing in a variety of preclinical models and tissues. In particular, FLASH irradiation was shown to improve survival following abdominal irradiation in mice [2], preserve cognitive function after whole-brain irradiation [3], and reduce brain toxicity using both electron and X-ray beams [4, 5].

In: International Journal of Radiation Oncology, Biology, Physics 98. p. 0
Where the paper says it

be widen the therapeutic window of radiotherapy. Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [7] for the determination of absorbed dose to water.

2 (2017), E16. doi: 10.1016/j.ijrobp.2017.02.101 >

[3] P. Montay-Gruel et al. “Irradiation in a flash: Unique sparing of memory in mice after whole brain irradiation with dose rates above 100Gy/s”. p. 0

Where the paper says it

Subsequent studies have confirmed normal-tissue sparing in a variety of preclinical models and tissues. In particular, FLASH irradiation was shown to improve survival following abdominal irradiation in mice [2], preserve cognitive function after whole-brain irradiation [3], and reduce brain toxicity using both electron and X-ray beams [4, 5].

In: Radiotherapy and Oncology 124. p. 0
Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

3 (2017). 15th International Wolfsberg Meeting 2017, pp. p. 0
Where the paper says it

be widen the therapeutic window of radiotherapy. Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [7] for the determination of absorbed dose to water.

365–369. issn: 0167-8140. doi: https://doi.org/10.1016/j.radonc.2017.05. 003 >

[4] P. Montay-Gruel et al. “X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in mice”. p. 0

Where the paper says it

In particular, FLASH irradiation was shown to improve survival following abdominal irradiation in mice [2], preserve cognitive function after whole-brain irradiation [3], and reduce brain toxicity using both electron and X-ray beams [4, 5]. The FLASH effect has also been demonstrated in larger animals and veterinary cancer patients [6], further highlighting its potential to ∗Corresponding author: pierre.

In: Radiotherapy and Oncology 129. p. 0
Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

3 (2018), pp. 582–588. doi: https: //doi.org / 10.1016 / j.radonc. 2018. 08. 016 >

[5] P. Montay-Gruel et al. “Long-term neurocognitive benefits of FLASH radiotherapy driven by reduced reactive oxygen species”. In: Proceedings of the National Academy of Sciences 116.22 (2019), pp. 10943–10951. doi: 10.1073/pnas. 1901777 >

[6] M.-C. Vozenin et al. “The Advantage of FLASH Radiotherapy Confirmed in Mini-pig and Catcancer Patients”. p. 0

Where the paper says it

Subsequent studies have confirmed normal-tissue sparing in a variety of preclinical models and tissues. In particular, FLASH irradiation was shown to improve survival following abdominal irradiation in mice [2], preserve cognitive function after whole-brain irradiation [3], and reduce brain toxicity using both electron and X-ray beams [4, 5].

In: Clinical Cancer Research 25. p. 0
Where the paper says it

Keywords: plane-parallel ionisation chambers, UHDR, charge collection efficiency, recombination, space charge, Monte Carlo simulation 1 Introduction Since the discovery of what is now referred to as the FLASH effect by Favaudon et al. [1] in 2014, research on ultra-high dose rate (UHDR) radiation (≥ 40 Gy s−1 ) has grown rapidly.

1 (Jan. 2019), pp. 35–42. issn: 1078-0432. doi: 10.1158 / 1078 - 0432. CCR - 17 - 3375 >

[7] INTERNATIONAL ATOMIC ENERGY AGENCY. Absorbed Dose Determination in External Beam Radiotherapy. p. 0

Where the paper says it

Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [7] for the determination of absorbed dose to water. In particular, plane-parallel ionisation chambers (PPICs) are widely used in radiotherapy both for reference dosimetry and for beam monitoring during patient treatment.

Rev. 1. Technical Reports Series. INTERNATIONAL ATOMIC ENERGY AGENCY, 2024. p. 0
Where the paper says it

be widen the therapeutic window of radiotherapy. Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [7] for the determination of absorbed dose to water.

isbn: 978-92-0-146022-6. doi: 10.61092/iaea. p. 0
Where the paper says it

be widen the therapeutic window of radiotherapy. Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [7] for the determination of absorbed dose to water.

ve7q-y94k >

[8] “Metrology for advanced radiotherapy using particle beams with ultra-high dose rates”. In: Physics in Medicine & Biology 69. p. 0

Where the paper says it

1 arXiv:2606. 24384v2 [physics.

p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

14 (July 2024), 14TR01. p. 3
Where the paper says it

10 1 10 0 10 1 10 2 10 24 10 25 Magboltz Davies and Chantry (1985) de Urquijo (2024) Reduced electric field E/N [Td] Ele ctro n m o bility e N [V 1 m 1 s 1 ] (a) Electron mobility in dry air. 10 1 10 0 10 1 10 2 10 48 10 47 Magboltz Davies and Chantry (1985) de Urquijo (2024) Reduced electric field E/N [Td] Atta c h m e nt c o e f ficie nt /N 2 [m5 ] (b) Electron attachment coefficient in dry air.

doi: 10.1088 / 1361 - 6560 / ad539d >

[9] A. Bourgouin et al. “Charge collection efficiency of commercially available parallel-plate ionisation chambers in ultra-high dose-per-pulse electron beams”. p. 0

Where the paper says it

In particular, plane-parallel ionisation chambers (PPICs) are widely used in radiotherapy both for reference dosimetry and for beam monitoring during patient treatment. However, dosimetry under UHDR conditions remains challenging [8], and experimental studies of commercially available PPICs have shown a strong dependence of their charge collection efficiency on dose per pulse, while also highlighting the need for improved ion recombination models for accurate dosimetry under UHDR conditions [9].

In: Physics in Medicine & Biology 68. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

23 (Nov. 2023), p. 235002. doi: 10.1088 / 1361 - 6560/ad0a58 >

[10] J. W. Boag. “Ionization Measurements at Very High Intensities. p. 0

Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

Pulsed Radiation Beams. p. 1
Where the paper says it

Several studies [17, 18, 19] have shown that these distortions can locally exceed the nominal applied electric field by several times. To better understand recombination effects in PPICs under UHDR pulsed beams, we extended the Garfield++ [20] framework and developed a selfconsistent Monte Carlo simulation coupling particle transport, electron attachment, ion–ion recombination, and space-charge effects.

” In: The British journal of radiology 23.274 (Oct. 1950), pp. 601–611. issn: 0007-1285. doi: 10.1259 / 0007 - 1285 - 23 - 274 - 601 >

[11] J. W. Boag, E. Hochhäuser, and O. p. 1

Where the paper says it

Under conventional radiotherapy conditions, ion recombination corrections are commonly performed using the two-voltage method, which is based on the model originally developed by Boag [10] in 1950. This model was later extended by Boag, Hochhäuser, and Balk [11] in 1996 to account for the fraction of electrons escaping attachment to electronegative molecules, referred to as the free electron fraction (FEF).

A. Balk. “The Effect of Free-Electron Collection on the Recombination Correction to Ionization Measurements of Pulsed Radiation”. p. 0
Where the paper says it

Keywords: plane-parallel ionisation chambers, UHDR, charge collection efficiency, recombination, space charge, Monte Carlo simulation 1 Introduction Since the discovery of what is now referred to as the FLASH effect by Favaudon et al. [1] in 2014, research on ultra-high dose rate (UHDR) radiation (≥ 40 Gy s−1 ) has grown rapidly.

In: Physics in Medicine and Biology 41. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

5 (May 1, 1996), pp. p. 1
Where the paper says it

Under conventional radiotherapy conditions, ion recombination corrections are commonly performed using the two-voltage method, which is based on the model originally developed by Boag [10] in 1950. This model was later extended by Boag, Hochhäuser, and Balk [11] in 1996 to account for the fraction of electrons escaping attachment to electronegative molecules, referred to as the free electron fraction (FEF).

885–897. issn: 0031-9155, 1361-6560. doi: 10.1088/0031-9155/41/5/005 >

[12] P. R. Almond et al. “AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams”. p. 0

Where the paper says it

Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [7] for the determination of absorbed dose to water. In particular, plane-parallel ionisation chambers (PPICs) are widely used in radiotherapy both for reference dosimetry and for beam monitoring during patient treatment.

In: Medical Physics 26. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

9 (1999), pp. 1847–1870. doi: https://doi.org/10.1118/ 1.598691 >

[13] R. F. Laitano et al. “Charge Collection Efficiency in Ionization Chambers Exposed to Electron Beams with High Dose per Pulse”. p. 0

Where the paper says it

In particular, plane-parallel ionisation chambers (PPICs) are widely used in radiotherapy both for reference dosimetry and for beam monitoring during patient treatment. However, dosimetry under UHDR conditions remains challenging [8], and experimental studies of commercially available PPICs have shown a strong dependence of their charge collection efficiency on dose per pulse, while also highlighting the need for improved ion recombination models for accurate dosimetry under UHDR conditions [9].

In: Physics in Medicine and Biology 51. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

24 (Dec. 21, 2006), pp. 6419–6436. issn: 0031-9155, 1361-6560. doi: 10.1088/0031-9155/51/24/009 >

[14] J. D. Fenwick and S. p. 6

Where the paper says it

In all charge-transport simulations presented here, electron–ion pairs were generated directly and uniformly within the irradiated volume. A square temporal pulse profile was used, except for the comparison with the Boag and Fenwick models in Section 4.

Kumar. “Collection Efficiencies of Ionization Chambers in Pulsed Radiation Beams: An Exact Solution of an Ion Recombination Model Including Free Electron Effects”. p. 0
Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Physics in Medicine & Biology 68.1 (Jan. 7, 2023), p. p. 6
Where the paper says it

4 Results and discussion 4. 1 Charge collection efficiency without space charge In 2023, Fenwick and Kumar [14] revisited Boag’s model proposed in 1996 and highlighted one of its main limitations: the oversimplified assumption regarding the spatial distribution of anions.

015016. issn: 0031-9155, 1361-6560. doi: 10.1088/1361-6560/aca74e >

[15] M. Orts et al. “Monte Carlo Simulation of Initial and Volume Ion Recombination Correction Factor in Plane Parallel Ionization Chambers Exposed to Ion Beams”. p. 0

Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Physics in Medicine & Biology 69.16 (Aug. 21, 2024), p. p. 3
Where the paper says it

10 1 10 0 10 1 10 2 10 24 10 25 Magboltz Davies and Chantry (1985) de Urquijo (2024) Reduced electric field E/N [Td] Ele ctro n m o bility e N [V 1 m 1 s 1 ] (a) Electron mobility in dry air. 10 1 10 0 10 1 10 2 10 48 10 47 Magboltz Davies and Chantry (1985) de Urquijo (2024) Reduced electric field E/N [Td] Atta c h m e nt c o e f ficie nt /N 2 [m5 ] (b) Electron attachment coefficient in dry air.

165012. issn: 0031-9155, 1361-6560. doi: 10.1088 / 1361 - 6560/ad6730 >

[17] R. Kranzer et al. p. 7

Where the paper says it

In particular, the electric field can locally exceed four times the nominal applied field, while dropping to nearly zero in other regions of the chamber. Moreover, very good agreement is observed with the results reported by Kranzer et al.

“Ion Collection Efficiency of Ionization Chambers in Ultra‐high Dose‐per‐pulse Electron Beams”. p. 0
Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Medical Physics 48. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

2 (Feb. 2021), pp. 819–830. issn: 0094-2405, 2473-4209. doi: 10.1002/mp.14620 >

[18] L. Baack et al. “Reduction of Recombination Effects in Large Plane Parallel Beam Monitors for FLASH Radiotherapy with Scanned Ion Beams”. p. 0

Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Physica medica: PM: an international journal devoted to the applications of physics to medicine and biology: official journal of the Italian Association of Biomedical Physics (AIFB) 104 (Dec. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

2022), pp. 136–144. issn: 1724-191X. doi: 10.1016/j.ejmp.2022.

[35] R. Brun and F. Rademakers. “ROOT: An Object Oriented Data Analysis Framework”. p. 2

Where the paper says it

To achieve this, several existing Garfield++ classes and methods were extended to account for space-charge effects and ion–ion recombination. 1 Garfield++ simulation toolkit Garfield++ [20] is an object-oriented simulation toolkit dedicated to the study of charged particle transport and interactions in gases and semiconductors.

In: Nucl. Instrum. Meth. A 389 (1997). Ed. by M. Werlen and D. Perret-Gallix, pp.: 81–86. doi: 10.1016/S0168-9002(97)00048-X >

[36] I. B. Smirnov. “Modeling of Ionization Produced by Fast Charged Particles in Gases”. p. 2

Where the paper says it

Originally developed by Rob Veenhof and now actively maintained at CERN, Garfield++ combines Monte Carlo transport techniques with electric and magnetic field maps to simulate charged particle transport in detector systems. The framework interfaces with several external packages, including ROOT [35] for visualisation, Magboltz [24] for the calculation of electron transport properties in gas mixtures, and Heed [36] for the simulation of primary ionisation produced by charged particles.

In: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 554.1 (Dec. 1, 2005), pp.: 474–493. issn: 0168-9002. doi: 10.1016/j.nima.2005.

[37] H. Schindler. “Microscopic Simulation of Particle Detectors”. p. 1

Where the paper says it

1 Electron swarm parameters and attachment processes The dominant mechanism responsible for the formation of negative ions in air is the attachment of electrons to oxygen molecules. Unlike noble gases, often used in gaseous detectors to avoid such phenomena, oxygen requires only a relatively low collision energy to capture free electrons.

PhD thesis. Technische Universität Wien, Oct. 2012

[38] J. M. Dawson. “Particle Simulation of Plasmas”. p. 1

Where the paper says it

Several studies [17, 18, 19] have shown that these distortions can locally exceed the nominal applied electric field by several times. To better understand recombination effects in PPICs under UHDR pulsed beams, we extended the Garfield++ [20] framework and developed a selfconsistent Monte Carlo simulation coupling particle transport, electron attachment, ion–ion recombination, and space-charge effects.

In: Reviews of Modern Physics 55. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

2 (Apr. 1, 1983), pp.: 403–447. doi: 10.1103/RevModPhys.55.403 >

[39] J. Villasenor and O. Buneman. “Rigorous Charge Conservation for Local Electromagnetic Field Solvers”. p. 0

Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Computer Physics Communications 69. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

2 (Mar. 1, 1992), pp.: 306–316. issn: 0010-4655 >

[40] C. K. Birdsall, A. B. Langdon, and A. B. Langdon. Plasma Physics via Computer Simulation. p. 0

Where the paper says it

1 arXiv:2606. 24384v2 [physics.

Boca Raton: CRC Press, Oct. 8, 2018

[41] R. W. Hockney and J. W. Eastwood. Computer Simulation Using Particles. p. 2

Where the paper says it

The TrackHeed class simulates the ionisation pattern produced by relativistic charged particles. Electron transport is performed using AvalancheMicroscopic, which tracks electrons collision-by-collision at the microscopic level, while ion transport is handled by AvalancheMC, which uses Monte Carlo techniques to simulate ion drift lines.

Boca Raton: CRC Press, Mar.: 978-0-367-80693-4 >

[42] L. Seidel. “Über ein Verfahren, die Gleichungen, auf welche die Methode der kleinsten Quadrate führt, sowie lineare Gleichungen überhaupt, durch successive Annäherung aufzulösen.” In: Abhandlungen der Mathematisch-Physikalischen Klasse der Königlich Bayerischen Akademie der Wissenschaften 11.3 (1874), pp.: 81–108. url: https: //www. biodiversitylibrary. org / item/110049>

[43] D. Young. “Iterative Methods for Solving Partial Difference Equations of Elliptic Type”. p. 5

Where the paper says it

A detailed description of the algorithm is given in Appendix B. Solving the Poisson equation with an iterative method is computationally advantageous in comparison, for instance, with the particle-to-particle or nodeto-node methods.

In: Transactions of the American Mathematical Society 76.1 (1954), pp.: 92–111. issn: 0002-9947, 1088-6850. doi: 10.1090/S0002-9947- 1954-

[44] W. Shockley. “Currents to Conductors Induced by a Moving Point Charge”. p. 5

Where the paper says it

The total electric field is the sum of the electric fields of each component. This class also allows the calculation of currents induced by drifting charged particles using the Shockley–Ramo theorem [44, 45].

In: Journal of Applied Physics 9. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

10 (Oct. 1938), pp.: 635–636. issn: 0021-8979. doi: 10.1063 / 1. 1710367 >

[45] S. Ramo. “Currents Induced by Electron Motion”. p. 5

Where the paper says it

This class also allows the calculation of currents induced by drifting charged particles using the Shockley–Ramo theorem [44, 45]. The pulse structure can be recreated in two different ways: either by assuming that ionisation is homogeneous and isotropic in the chamber and distributing the electron–ion pairs accordingly, or by using the

In: Proceedings of the IRE 27.9 (1939), pp.: 584–585. doi: 10.1109 / JRPROC. 1939.228757 >

[46] D. C. Meeker. “Improvised Asymptotic Boundary Conditions for Electrostatic Finite Elements”. p. 7

Where the paper says it

This configuration provides a known analytical solution for the electric field while remaining representative of charge distributions encountered in PPICs. To approximate open boundary conditions within a finite computational domain, mixed boundary conditions were applied.

In: IEEE Transactions on Magnetics 50.6 (2014), pp. p. 0
Where the paper says it

Keywords: plane-parallel ionisation chambers, UHDR, charge collection efficiency, recombination, space charge, Monte Carlo simulation 1 Introduction Since the discovery of what is now referred to as the FLASH effect by Favaudon et al. [1] in 2014, research on ultra-high dose rate (UHDR) radiation (≥ 40 Gy s−1 ) has grown rapidly.

: 1–9. doi: 10.1109 / TMAG.2014.2300196 >

[47] J. Jin. The Finite Element Method in Electromagnetics. p. 7

Where the paper says it

This configuration provides a known analytical solution for the electric field while remaining representative of charge distributions encountered in PPICs. To approximate open boundary conditions within a finite computational domain, mixed boundary conditions were applied.

en. 3rd ed. Wiley - IEEE. Nashville, TN: John Wiley & Sons, Mar.: 2014 >

[48] J. Paz-Martín et al. p. 8

Where the paper says it

Such perturbations strongly affect electron and ion transport properties, including drift velocity, mobility, and electron attachment coefficient, thereby modifying the free electron fraction. 3 Free electron fraction Following the work of Paz-Martín et al.

, “Numerical Modeling of AirVented Parallel Plate Ionization Chambers for Ultra-High Dose Rate Applications”. p. 0
Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Physica Medica 103 (Nov. 2022), pp.: 147–156. issn: 11201797. doi: 10.1016/j.ejmp.2022.

[33] S. McGowan. “Ion-Ion Recombination in Laboratory Air”. p. 0

Where the paper says it

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Pierre Gérard Ortega1,∗ , Gilles De Lentdecker1 , Quentin Flandroy2 , Jarrick Nys2 1 Service de Physique des particules élémentaires (IIHE), Université libre de Bruxelles (ULB), Brussels, Belgium 2 Ion Beam Applications SA, Louvain-la-Neuve, Belgium September 18, 2026 Abstract Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion–ion recombination and self-consistent space-charge electric field calculations.

In: Physics in Medicine & Biology 10. p. 0
Where the paper says it

1 arXiv:2606. 24384v2 [physics.

1 (Jan. 1, 1965), pp.: 25–40. issn: 0031-9155, 1361-6560. doi: https: //iopscience.iop.

Green — the paper says it, and the page is printed. Red — no passage carries it, so treat that sentence as unverified. The same check runs on your own text: check a summary against its source, $5.

Improvements for AI systems

  1. Improved dosimetry models can accurately predict charge collection efficiency under ultra-high dose rate irradiation by incorporating self-consistent space-charge electric field calculations and ion recombination processes, as demonstrated by extending Garfield++ to couple particle transport, electron attachment, and ion ion recombination.

  2. The improved system can calculate the free electron fraction (FEF) under ultra-high dose rate conditions by accounting for electric field dependent electron attachment, which is shown to be the primary driver of CCE reduction.

  3. The AI system can simulate the temporal evolution of electric field distortions inside ionization chambers, reproducing results like the electric field can locally exceed four times the nominal applied field or drop to nearly zero in some regions of the chamber after a pulse.

  4. The framework enables the development of real-time correction methods for ionisation chamber dosimetry under ultra-high dose rate irradiation by providing a flexible and self-consistent Monte Carlo simulation environment.

  5. The system can accurately reproduce charge collection efficiency as a function of dose per pulse (DPP) by comparing simulation results with experimental data, validating the model's ability to handle space-charge effects and recombination under UHDR conditions.

Abstract

Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion--ion recombination and self-consistent space-charge electric field calculations. The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions. The implementation was validated against analytical and numerical models from the literature. Excellent agreement was obtained for the free electron fraction (FEF), CCE, induced current, and electric field evolution under UHDR conditions. The simulations show that space charge can locally increase the electric field by more than a factor of four or reduce it to nearly zero in some regions of the chamber. The results further suggest that the reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment. This finding indicates that the complex problem of recombination under UHDR conditions may be largely governed by the evolution of the FEF, opening promising perspectives for the development of improved analytical models and real-time correction methods for ionisation chamber dosimetry under UHDR irradiation conditions. This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy.